Power Meters MSA Questions-Responses 11-10-2026.pdf
PDF 175 KB Posted
- Solicitation number
- FAA-DME-PWRMTR-MSA2026
About this file
This document is a Questions and Responses file for Market Survey Announcement FAA-DME-PWRMTR-MSA2026 regarding commercially available Distance Measuring Equipment (DME) Power Meters. The FAA is conducting market research to identify single-channel, high-performance RF power meters currently available in the commercial market. OEM-supported reseller responses are acceptable provided the reseller has received authorization from the OEM. Respondents should attribute deployment history and production capacity to the OEM when applicable.
Key technical requirements clarified through the Q&A include: accuracy specifications apply to the complete meter-and-sensor system rather than the meter alone; the pulse spectrum requirement is greater than or equal to 1 MHz; the display operating temperature requirement can be relaxed from 55°C to 40°C due to expected use in controlled environments; adapters alone (without additional cable) are acceptable for Type N female connections, though computer-based displays may require additional FAA IT coordination for installation on technician computers; and the FAA does not currently require dual-channel power meters or equipment to supply DME signals to ground stations. Respondents may provide photographs of equipment instead of detailed technical drawings and may submit estimates of currently deployed power meters in lieu of exact sales data. This is a market survey announcement and not a Request for Proposal, so vendors should refer to Section 1 of the MSA for guidance on quantities, contract terms, delivery locations, and shipment schedules.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| DME Power Meters MSA Final.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Acquisition Sensitive – This document may contain information considered sensitive unclassified information (SUI) and would merit protection from unauthorized disclosure under 5 U.S.C, Section 552(b)(4). No information shall be released except as authorized by the Contracting Officer.
1 | P a g e
Questions/Responses
Market Survey Announcement FAA-DME-PWRMTR-MSA2026 Commercially Available Distance Measuring Equipment (DME)
Power Meters
1. Are OEM-supported reseller responses acceptable, with deployment history and production capacity attributed to the OEM?
Response: Yes, provided that the reseller has received authorization from the OEM to submit a response.
2. What quantities, contract terms, option periods and ordering assumptions should vendors use for the ROM?
Response: This notice is not being issued for purpose of issuing a Request for Proposal.
Please refer to Section 1 (Notice) of the MSA.
3. What delivery locations, lead times and shipment schedule should be assumed?
Response: See response to question number 2.
4. Do the accuracy limits apply to the meter alone or the complete meter-and-sensor system?
Response: Accuracy applies to meter plus sensor system.
5. Must the 65 dB range support peak measurements at ≥1.5 MHz bandwidth and must the display also operate through 55°C?
Response: The pulse spectrum is within 1 MHz; the requirement would be greater than or equal to 1 MHz. The 55°C requirement can be relaxed to a value of 40°C given the power meter is expected to be used in a controlled environment.
6. Is an adapter acceptable for the Type N female connection, and are computer-based displays acceptable?
Response: If it is an adapter alone (no additional cable) it would be acceptable.
Regarding computer-based displays, it appears to be the current trend for power meters. However, the FAA IT would have to load it to installers computers, which may require additional coordination/requirements
Acquisition Sensitive – This document may contain information considered sensitive unclassified information (SUI) and would merit protection from unauthorized disclosure under 5 U.S.C, Section 552(b)(4). No information shall be released except as authorized by the Contracting Officer.
2 | P a g e
7. Is a one or two channel power meter needed for this requirement?
Response: At the present time, the FAA does not have a requirement for a dual power meter. For the response to this MSA, the single-channel, high-performance RF power meter is the minimum requirement.
8. Does the FAA have a need to supply a DME signal to the DME ground station?
Response: At this time, the FAA does not have a requirement for supporting future DME installations/maintenance
9. What kind of drawings or illustrations of the equipment are needed? Would photographs of the equipment be sufficient?
Response: Photographs of the equipment is acceptable.
10. In lieu of providing information on the number of systems sold, is an estimate of the number of power meters that are currently deployed acceptable?
Response: Yes. Estimates are acceptable.
| Questions/Responses |
| Market Survey Announcement FAA-DME-PWRMTR-MSA2026 |
File details come from the government source that posted it. Updated .