Technical Specification Sheet - ETF Hygrometers.pdf

PDF 161 KB Posted

Attached to
ETF Hygrometers
Solicitation number
FA910126QB073
Issued by

About this file

This is a Technical Specifications Sheet for the procurement of nine (9) dew point hygrometers, which are commercially available equipment required to meet specific performance and operational standards.

The equipment must feature a primary multi-stage chilled mirror sensor technology with drift-free, continuous dew/frost point temperature measurement capability. The hygrometers shall be configured as 19-inch rack/panel mountable consoles (approximately 19" W × 7" H) with remote-mounted sensor capability supporting distances up to 250 feet. Critical performance requirements include a minimum 95°C cooling depression capability (3-stage thermoelectric cooling equivalent), measurement accuracy of ±0.2°C across a dew/frost point range of −65°C to +75°C, a chromium/chrome mirror surface, and an Automatic Balance Cycle (ABC) or equivalent optical auto-standardization feature for contamination compensation. The equipment must support both Programmable Balance Cycles (PABC) and Manual Balance Cycles (MABC), include multiple user-configurable analog outputs (4–20 mA and 0–5 Vdc ranges), a dedicated bi-directional RS-232 serial interface for data acquisition and remote programming, and minimum two programmable form C SPDT dry-contact relay alarms. Additional requirements include a universal power supply (90–230 VAC, ±10%, 50/60 Hz), a NIST-traceable calibration certificate from an ISO/IEC 17025 accredited laboratory, and a minimum one-year warranty on all parts, sensors, and labor. No auxiliary probes for ambient air temperature or active inline pressure sensors are required. The specification references Edgetech Instruments part number DM-PMDM-X3SF-STD-RC(250FT.)-AFSP as a reference standard or equivalent "or equal" commercially available products.

View the file

Other files

File Type Posted
Cover Page ETF Hygrometers CO.pdf PDF
Cover Page ETF Hygrometers.pdf PDF
Solicitation - FA910126QB073.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Technical Specifications Sheet

The Government requires commercially available equipment meeting the specifications provided below.

Item – Dew Point Hygrometer Quantity: 9

Specification Required Value / Performance

Measurement Technology Primary method, multi-stage chilled mirror sensor providing drift-free, continuous dew/frost point temperature measurement.

Chassis / Enclosure Type 19–inch rack/panel mountable console configuration, standard rack dimensions (approx. 19” W × 7” H).

Sensor Installation Remote-mounted configuration supporting sensor-to-console distances up to 250 feet (76 meters).

Mirror Surface Material Standard chromium/chrome mirror surface.

Cooling Depression Capability

Minimum 95 °C cooling depression capability (equivalent to 3-stage thermoelectric cooling performance).

Dew/Frost Point Range Able to cover a dew/frost point range of −65 °C to +75 °C (at nominal ambient temperature).

Measurement Accuracy ±0.2 °C (±0.36 °F) or better dew/frost point accuracy across the full range.

Contamination Compensation

Must include an Automatic Balance Cycle (ABC) or equivalent optical auto-standardization feature to correct for mirror contaminants.

Operational Modes Supports Programmable Balance Cycles (PABC) and Manual Balance Cycles (MABC).

Calibration Standards Ships with a NIST-traceable certificate of calibration performed by an ISO/IEC 17025 accredited laboratory.

Analog Outputs Multiple user-configurable/scalable analog outputs, including 4–20 mA and 0–5 Vdc ranges.

Specification Required Value / Performance

Digital Communication Dedicated bi-directional RS-232 serial interface for data acquisition and remote programming.

Alarm Relays Minimum two (2) programmable, form C SPDT dry-contact relays rated for process alarms.

Power Requirements Universal power supply accepting 90 to 230 VAC (±10%), 50/60 Hz.

Auxiliary Probes Required None (No ambient Air Temperature or active inline pressure sensors required).

Warranty Support Minimum one (1) year warranty on all parts, sensors, and labor.

NOTE: The requirement can be met by manufacturer Edgetech Instruments, part number DM-PMDM-X3SF-STD-RC(250FT.)-AFSP, or an “or equal” commercially available product.

Technical Specifications Sheet

File details come from the government source that posted it. Updated .