Attachment 1 - Thermocouple Specificaitons.pdf

PDF 104 KB Posted

Attached to
Coaxial Thermocouple Requirement
Solicitation number
FA9101-25CoaxialThermocoupleNOI
Issued by

About this file

This document is a detailed technical specification for coaxial thermocouples used in the AEDC Hypervelocity Wind Tunnel 9 facility. The specifications outline requirements for Type-E, Chromel-Constantan thermocouples with precise technical parameters, including outer sensor cylinder diameters of 0.061" and 0.031", sensor length options of 0.375" and 0.5" or longer, and the ability to be trimmed and contoured to test article outer mold lines. The thermocouples are designed for heat transfer measurements, with fast response times typically on the order of microseconds, and are intended to be installed in stainless steel substrates with wall thicknesses between 0.25"-0.5".

Key installation specifications include using adhesives like Loctite #271 or #620, creating a thermocouple junction at the outer surface through abrasion, and using Type-E thermocouple lead wires no larger than 30 AWG Teflon insulated solid wires. The document emphasizes the thermocouples' proven durability, having been successfully utilized for over 30 years in the unique and stressing Tunnel 9 hypervelocity test environment, with the ability to derive heat transfer data by using the thermocouple temperature as a boundary condition in heat conduction calculations.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

AEDC Tunnel 9 Coaxial Thermocouple Requirements

Background The majority of the heat transfer measurements made in the AEDC Hypervelocity Wind Tunnel 9 facility have been collected using coaxial thermocouples mounted directly into a stainless steel substrate (15-5PH or 17-4PH stainless steel) test article. Typical wall thickness of the steel substrate generally ranges between 0.25”- 0.5”, and the interior volume to route lead wires can be limited. These installation limitations require the sensor length to be similar to the wall thickness, and the transition from the sensor shell to the lead wires must occur close to the base of the sensor. Finally, a thin layer of temperature sensitive paint is often applied directly on the wind tunnel model after the coaxial thermocouples have been installed and contoured to the test article.

Type-E, Chromel-Constantan, coaxial thermocouples have been successfully utilized for over 30 years in the unique and stressing Tunnel 9 hypervelocity test environment due to their high durability, fast-response times, and their ability to be contoured to the outer mold lines of the test article. Surface temperature response is acquired during the test run using a coaxial thermocouple and the heat transfer is derived using the coaxial thermocouple temperature data as a boundary condition to a finite-differencing solution of the transient heat conduction equation for a finite thickness wall.

Technical Specifications

1. NIST-traceable Type-E, Chromel-Constantan thermocouple materials (outer sheath Chromel and inner pin Constantan)

2. Outer sensor cylinder diameters of 0.061” and 0.031”

3. Sensor length options of 0.375” and 0.5” or longer

4. The sensor length may be trimmed down for installation in thinner walls or to fit into limited internal volume of the model without adverse effects to sensor operation

5. Sensor to be installed into test article using adhesives such as Loctite #271 or #620

6. Sensor can be contoured to outer-mold-lines of a test article after adhesive fully cured

7. Thermocouple junction at the outer surface of the model created by abrasion (file and/or sand paper).

8. Fast response time to metal wall surface temperature changes, typically on the order of microseconds

9. Transition from coaxial thermocouple body to lead wires to occur close to body, approximately 0.15 or less”

10. Any strain relief applied between the coaxial thermocouple body and lead wires cannot be larger than the cylindrical diameter of the coaxial thermocouple body

11. Lead wires from coaxial thermocouple assembly to be provided as Type-E thermocouple wire. Lead wires should be no larger than 30 AWG Teflon insulated solid wires.

AEDC Tunnel 9 Coaxial Thermocouple Requirements
Background
Technical Specifications

File details come from the government source that posted it. Updated .