Attachment 2 Evaluation Criteria-UPDATED.xlsx

XLSX spreadsheet 12 KB Posted

Attached to
Automated Chemistry Testing Platform
Solicitation number
FA8601-21-Q-0093
Issued by

View the file

Other files

File Type Posted
Core Lab Layout_Projected (002).pptx PPTX presentation
Questions and Answers.docx DOCX document
Solicitation.pdf PDF
Attachment 2 Evaluation Criteria.xlsx XLSX spreadsheet
Attachment 4-SOW-July 2021.pdf PDF
Attachment 1 DRAFT BPA Document.pdf PDF
Attachment 3 Pricing Sheet.xlsx XLSX spreadsheet

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sheet1

EVALUATION CRITERIA

*REFERENCE THE SOW TO SEE SUB REQUIREMENTS

OBJECTIVESMinimum RequirementMinimum Requirement RatingJUSTIFICATION FOR RATING: (Vendor must provide narrative to get credit - can attach narrative not to exceed one page per objective)SUPPORTING INFORMATION LOCATION: Indicate where in the vendors offer the minimum requirement or objective is supported
1.1.6. The contractor shall provide a fully integrated testing platform solution consisting of 2 identical analyzers capable of performing all tests listed in Section 1.1.1. (No rating assigned)N/AMeets
1.1.7. The contractor shall provide a comprehensive solution that addresses the reduction, mitigation, or complete elimination of carryover between samples. Please provide mitigation plan in proposal.

Evaluation: Excellent, Satisfactory or Unsatisfactory Definitions.

Excellent: Solution is expected to eliminate carryover between samples.

Satisfactory: Solution is expected to reduce or mitigate some carryover between samples.

Unsatisfactory: Solution was not provided or is not reasonably expected to reduce or mitigate carryover between samples.Does Not Meet
1.1.11. Analyzer must have minimal maintenance requirements. Provide a detailed schedule of daily maintenance performed by the user with hands-on times listed.

Evaluation: Excellent, Satisfactory or Unsatisfactory Definitions.

Excellent: <10 minutes of hands-on time per day.

Satisfactory: 10-30 minutes of hands-on time per day.

Unsatisfactory: >30 minutes of hands-on time per day, or list was not provided.Excellent
1.1.25. Analyzer should have Authorization to Operate (ATO) on Defense Health Agency (DHA) networks. Provide eMASS number.

Evaluation: Excellent, Satisfactory or Unsatisfactory Definitions.

Excellent: Contractor has approved ATO.

Satisfactory: Started process to get an ATO.

Unsatisfactory: ATO process not started.

Satisfactory
1.2.6 Reagents must be ready for use without reconstitution. Provide a list of any reagents that require preparation.

Evaluation: Excellent, Satisfactory or Unsatisfactory Definitions.

Excellent: 100% of reagents ready for use without reconstitution.

Satisfactory: >95% of reagents ready for use without reconstitution.

Unsatisfactory: Less than or equal to 95% of reagents ready for use without reconstitution, or list not provided.Unsatisfactory
1.2.14. Must have sufficient onboard reagent stability. Provide a list of any reagents with less than 14 days onboard stability.

Evaluation: Excellent, Satisfactory or Unsatisfactory Definitions.

Excellent: 3 or less reagents have onboard stability <14 days.

Satisfactory: 10 or less reagents have onboard stability <14 days.

Unsatisfactory: More than 10 reagents have onboard stability < 14 days, or list was not provided.

1.2.15. Assays must have sufficient calibration stability. Provide a list of calibration stabilities for each assay.

Evaluation: Excellent, Satisfactory or Unsatisfactory Definitions.

Excellent: >90% of assays have >90 day stability Satisfactory: 50-90% of assays have >90 day stability Unsatisfactory: <50% of assays have >90 day stability, or list was not provided

Completed Excel Pricing Sheet (No rating assigned) N/A