Attachment 2 Evaluation Criteria.xlsx

XLSX spreadsheet 12 KB Posted

Attached to
Automated Chemistry Testing Platform Federal contract opportunity
Solicitation number
FA8601-21-Q-0093
Issued by
Department of the Air Force Materiel Command Lifecycle Management Center Wright Patterson Air Force Base

View the file

Other files for this federal contract opportunity

Other files attached to Automated Chemistry Testing Platform, newest first.
File Type Posted
Core Lab Layout_Projected (002).pptx PPTX presentation
Questions and Answers.docx DOCX document
Attachment 2 Evaluation Criteria-UPDATED.xlsx XLSX spreadsheet
Attachment 4-SOW-July 2021.pdf PDF
Attachment 1 DRAFT BPA Document.pdf PDF
Solicitation.pdf PDF
Attachment 3 Pricing Sheet.xlsx XLSX spreadsheet

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sheet1

EVALUATION CRITERIA

*REFERENCE THE SOW TO SEE SUB REQUIREMENTS

OBJECTIVESMinimum RequirementMinimum Requirement RatingJUSTIFICATION FOR RATING: (Vendor must provide narrative to get credit - can attach narrative not to exceed one page per objective)SUPPORTING INFORMATION LOCATION: Indicate where in the vendors offer the minimum requirement or objective is supported
1.1.6. The contractor shall provide a fully integrated testing platform solution consisting of 2 identical analyzers capable of performing all tests listed in Section 1.1.1. (No rating assigned)N/AMeets
1.1.7. The contractor shall provide a comprehensive solution that addresses the reduction, mitigation, or complete elimination of carryover between samples. Please provide mitigation plan in proposal.

Evaluation: Excellent, Satisfactory or Unsatisfactory Definitions.

Excellent: Solution is expected to eliminate carryover between samples.

Satisfactory: Solution is expected to reduce or mitigate some carryover between samples.

Unsatisfactory: Solution was not provided or is not reasonably expected to reduce or mitigate carryover between samples.Does Not Meet
1.1.11. Analyzer must have minimal maintenance requirements. Provide a detailed schedule of daily maintenance performed by the user with hands-on times listed.

Evaluation: Excellent, Satisfactory or Unsatisfactory Definitions.

Excellent: <10 minutes of hands-on time per day.

Satisfactory: 10-30 minutes of hands-on time per day.

Unsatisfactory: >30 minutes of hands-on time per day, or list was not provided.Excellent
1.1.25. Analyzer should have Authorization to Operate (ATO) on Defense Health Agency (DHA) networks. Provide eMASS number.

Evaluation: Excellent, Satisfactory or Unsatisfactory Definitions.

Excellent: Contractor has approved ATO.

Satisfactory: Started process to get an ATO.

Unsatisfactory: ATO process not started.

Satisfactory
1.2.6 Reagents must be ready for use without reconstitution. Provide a list of any reagents that require preparation.Unsatisfactory
1.2.14. Must have sufficient onboard reagent stability. Provide a list of any reagents with less than 14 days onboard stability.

Evaluation: Excellent, Satisfactory or Unsatisfactory Definitions.

Excellent: 3 or less reagents have onboard stability <14 days.

Satisfactory: 10 or less reagents have onboard stability <14 days.

Unsatisfactory: More than 10 reagents have onboard stability < 14 days, or list was not provided.

1.2.15. Assays must have sufficient calibration stability. Provide a list of calibration stabilities for each assay.

Evaluation: Excellent, Satisfactory or Unsatisfactory Definitions.

Excellent: >90% of assays have >90 day stability Satisfactory: 50-90% of assays have >90 day stability Unsatisfactory: <50% of assays have >90 day stability, or list was not provided

Completed Excel Pricing Sheet (No rating assigned) N/A