FA8601-17-T-0056 Questions and Answers Document Last updated: 14 April 2017 at 10:18 AM Eastern Time
1. Question: Are the researchers asking for dedicated calibration lamps for in house use or just that the system is calibrated and checked.
Answer: We require dedicated calibration lamps for in house use.
2. Question: Lines 2 and 3 state; "Double grating Czerny-Turner monochromator in..."
Does this mean;
1. a double or triple grating turret in a single monochromator, or 2. a double monochromater, each with its own grating turrets.
Answer: We require a double monochromator - not a double grating turret.