P24-017 Rubber Room Stability Chamber Specification Sheet.1696871921779 (2).pdf

PDF 363 KB Posted

Attached to
RUBBER ROOM STABILITY CHAMBER Federal contract opportunity
Solicitation number
FA813224Q0011
Issued by
Department of the Air Force Materiel Command Air Force Sustainment Center

About this file

This document contains a technical specification sheet for Climacell climate test chambers. The specification sheet provides details on various Climacell chamber models including their interior dimensions, temperature and humidity ranges and controls, construction materials, electrical specifications, optional equipment, and controller features.

The related federal contract opportunity is a solicitation seeking offers for a rubber room stability chamber to be used by the Department of the Air Force Materiel Command Air Force Sustainment Center. Offerors should review the solicitation documentation including the contract statement of services and work for full requirement details. The specification sheet provides relevant technical information on Climacell chamber features that may inform proposals.

View the file

Other files for this federal contract opportunity

Other files attached to RUBBER ROOM STABILITY CHAMBER, newest first.
File Type Posted
Updated Spec Sheet for CSZ Model STH-52-0.50-H AC StableClimate II Temperature Humidity Stability Chamber.pdf PDF
P24-017 Rubber Room Stability Chamber SOW_REDACTED.pdf PDF
Spec_521125.pdf PDF
CSS RUBBER ROOM STABILITY CHAMBER.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Climacell Eco + Evo Materials Testing Chambers

Patented Forced Air Convection

Materials testing in both automotive and aerospace industries

Electronics testing including use in metrology labs

Quality and durability testing of materials, components, and packaging

CLIMACELL® climate chambers are designed to create exact reproducible simulations of a wide range of environmental conditions through precise control of temperature, humidity and optional lighting.

CLIMACELL® test chambers applications include degradation testing, QC/QA testing of products and packaging, and material quality and durability testing. CLIMACELL® EVO models can be equipped with programmable LED or fluorescent shelf and door lighting for a variety of VIS or UV light testing studies.

Temperature Range:

0°C up to 100°C (10°C up to 95°C with humidity)

• -10°C min. temperature option available on ECO units

• -20°C min. temperature option available on EVO units

Humidity Control:

ECO: 10% up to 95% – controlled in 1% increments starting @ 10°C EVO: 10% up to 98% – controlled in 1% increments starting @ 10°C

Temperature Control:

Short recovery times with precise temperature regulation and uniformity

Chamber Volumes:

111 (4 ft³) • 222 (8 ft³) • 404 (14.3 ft³) • 707 (25 ft³) • 1212 (43 ft³)

Chamber Construction:

• AISI 304 stainless steel chamber (AISI 316 SS option available)

• Double wall, seamless main chamber & removeable inner chamber

Electrical Data:

230V 50/60Hz, UL & CE

5.7” LCD touch display

Fuzzy Logic algorithm constantly monitors chamber conditions and continuously optimizes parameters.

(100) programs with (100) segments each for varying loads and parameters

Audible & visual alarms – temperature, time & humidity

Warmcomm Software for data capture and device control (P version)

USB device, RS232 & optional Ethernet port

Integrated SD card 30-day data logger & multi-level secure user authentication

50mm (2”) port

3” LCD display

Fuzzy Logic algorithm constantly monitors chamber conditions and continuously optimizes parameters.

(9) programs with (8) segments each for varying loads and parameters

Audible & visual alarms – temperature, time & humidity

Warmcomm Software for data capture and device control (P version)

USB flash & device, RS232 & optional Ethernet port

Integrated USB 30-day data logger for temperature measurement & recording

50mm (2”) port

Optional Equipment:

• AISI 316 stainless steel chamber

• 160°C sterilization cycle (EVO only)

• Access ports 25 (1”), 50 (2”), 100 (4”) mm

• Ethernet communication port

• Rolling carts for 111 & 222 models

• Door sensor and alarm

• Automatic and key door lock

• Warmcomm data acquisition software:

✓ 4.0B – Receive data

✓ 4.0P – Receive data and control the device

✓ 4.0F – FDA 21 CFR part 111 compliant

• BMS – Building monitoring alarm contact

• 4-20mA contacts for temperature & humidity

• Flexible PT 100 sensor

• Stainless steel exterior

• Heavy load chamber

• IQ / OQ protocols with 9pt temperature mapping

Optional Programmable Lighting:

• VIS & UV fluorescent lighting

• VIS LED shelf and door lighting

• Light intensity controlled in 1% increments

• LED Door lighting (pictured)

ECO Controller EVO Controller

Climacell ECO/EVO Technical Data Model 111 222 404 707 1212

Interior Dimensions

Chamber:

AISI 304 stainless steel (AISI 316 stainless steel option available)

Volume ft3 3.9 7.8 14.3 25 43 liters 111 222 404 707 1212

Width inches 21.25 21.25 21.25 37 3 x 21.25 mm 540 540 540 940 3 x 540

Depth inches 14.6/15 20.5/20.9 20.5/20.9 20.5/20.9 20.5/20.9 mm 370/380 520/530 520/530 520/530 520/530

Height inches 20.9/21 29.9/30.1 55.7 55.7 55.7 mm 530/535 760/765 1415 1415 1415

Exterior Dimensions (ECO/EVO)

Width inches 29.9/30.7 29.9/30.7 41.7/43.1 57.5/59.2 103.5/103.5 mm 760/780 760/780 1060/1100 1460/1500 2360/2360

Depth inches 29.5/29.7 35.4/34.8 33.9/34.8 33.9/34.8 34.2/35.4 mm 750/755 895/885 860/885 860/885 870/898

Height (Legs L, Caster C) inches 44.9L/47.8C 53.9L/57.1C 75L/74.4C 75L/74.4C 75.2L/75.6C mm 1140L/1215C 1370L/1450C 1905L/1890C 1905L/1890C 1905L/1921C

Shelves: Stainless Steel Capacity: # of shelf guides in chamber side walls

Maximum # 7 10 19 19 3 x 19

Standard # 2 2 2 2 6

Shelf Distance Min. distance between trays Inches 2.8 2.8 2.8 2.8 2.8 mm 70 70 70 70 70

Useable Shelf Area Width x Depth Inches 20.5x13.2 20.5x19.1 20.5x19.1 36.3x19.1 20.5x19.1x3 mm 520x335 520x485 520x485 920x485 520x485x3

Maximum Shelf Load

One Shelf lbs 44 66.1 66.1 110.2 66.1 kg 20 30 30 50 30

Total Per Unit lbs 110.2 154.3 220.5 286.6 661 kg 50 70 100 130 300

# Outer Metal Doors 1 1 1 2 3

# Inner Glass Doors 1 1 1 2 3

Volume of Steam Space ft3 5.8/5.9 10.6/10.8 18.5/18.7 30.9/31 61.9 liters 163/167 299/305 524/530 876/878 1753

Operation Temperature From 0°C (-10°C /-20°C ) Up to °C 100 100 100 100 100

Temperature Accuracy

Distribution @ 10°C ± °C <0.5 <0.5 <1 <1 <0.9

Distribution @ 37°C ± °C <0.5 <0.5 <1 <1 <0.5

Uniformity ± °C <0.2 <0.2 <0.3 <0.4 <0.2

Heating Time to 37°C From the Ambient Temperature Minutes <11 <11 <13 <13 <30

Cooling Down Time From 22°C to 10°C Minutes <21 <17 <19 <21 <21

Recovery Time After Door Opened for 30 s according to DIN 12880

@ 37°C Minutes <4 <3 <3 <6 <10

@ 50°C Minutes <5 <6 <7 <6 <10

Relative Humidity (RH) Range % 10% - 98% 10% - 98% 10% - 98% 10% - 98% 10% - 98%

Accuracy RH (TCHAMBER ≥ 21°C) In Time % <2 <2 <2 <2 <2

Heat Emission @ 37°C W 70 63 123 148 200

CO2 Concentration (EVO Only) % 0.1 – 20 0.1 – 20 0.1 – 20 - -

CO2 Required Pressure (EVO Only) Bar/PSI 0.3-0.7/5—10 0.3-0.7/5--10 0.3-0.7/5--10 - -

Noise Level of Complete Device dB 46 50 56 58 60

Electrical Data: ECO & EVO 0°C Max Consumption 50/60Hz

W 2050 2200 3150 3400 3500

A 8.9 9.6 13.7 14.8 15.2

V 230 230 230 230 230

Electrical Data: EVO -20°C Protective system: 16A minimum

Max Consumption 50/60Hz W 1630 1780 2115 2640 3215

V 230 230 230 230 230

IP Code IP20 IP20 IP20 IP20 IP20

ECO/EVO Weight

Net lbs 229.3/242.5 297/324.1 520.3/529.1 599.6/617.3 1193/1250 kg 104/110 135/147 236/240 272/280 541/567

Gross lbs 288.8/390 372.6/390 696.2/617 696.7/719 1898 kg 131/140 169/177 270/280 316/326 861

EVO Weight -20°C

Net lbs 264.5 331 551 639 750 kg 120 150 250 290 340

Gross lbs 331 412 639 741 871 kg 150 187 290 336 395

BMT USA, LLC • 360.863.2252 • sales@bmtusa.com • www.bmtusa.com mailto:sales@bmtusa.com http://www.bmtusa.com/

File details come from the government source that posted it. Updated .