This document is a product specification sheet for the Illumina NextSeq 1000 and NextSeq 2000 Sequencing Systems. The NextSeq 1000 and NextSeq 2000 are benchtop next-generation sequencing platforms that offer a range of features and capabilities, including high-performance sequencing by synthesis (SBS) chemistry, streamlined workflows, and integrated secondary analysis with the DRAGEN bioinformatics pipeline. Key details include:
The systems offer outputs from 10 Gb to 540 Gb, with read lengths from 1x50 bp to 2x300 bp, supporting a variety of applications such as small whole-genome sequencing, whole-exome sequencing, total RNA-seq, and single-cell RNA-seq. The NextSeq 2000 P4 flow cell can deliver up to 1.8 billion single-end reads per run. The systems provide flexible analysis options, including onboard DRAGEN secondary analysis and cloud-based analysis through BaseSpace Sequence Hub. The document also highlights the systems' modular design, Illumina Proactive remote support service, and comprehensive training and support offerings.
The related federal contract opportunity is for the purchase and installation of a NextSeq 1000 system (model #20038898) at Camp Zama in Zama-shi, Kanagawa, Japan. The contractor shall provide the system, deliver it to the specified location, and provide on-site training in both English and Japanese.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 3 Atch2 Class_Deviation_2018-O0019.pdf | ||
| 4 BNJ 24Q063A.pdf | ||
| 1 Solicitation - FA520924Q0071.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
| 1
NextSeq™ 1000 & NextSeq 2000 Sequencing Systems Expansive application breadth and operational simplicity with proven performance
SPECIFICATION SHEET
• Broad platform utility with the flexibility and scalability to expand benchtop applications
• Streamlined workflow powered by XLEAP-SBSTM chemistry and onboard DRAGENTM secondary analysis
• Robust performance and high-quality data from a proven global leader
M-NA-00008 v6.0For Research Use Only. Not for use in diagnostic procedures.
RFQ No. FA520924Q0071 Attachment 1 Date: 15 Aug 2024
For Research Use Only. Not for use in diagnostic procedures.2 | M-NA-00008 v6.0
NExTSEq 1000 AND NExTSEq 2000 SEquENCING SySTEMS
Introduction
Next-generation sequencing (NGS) has revolutionized biological research, expanding the range and depth of sequencing applications that labs can perform. The research community can ask, and answer, increasingly complex scientific questions. Across the spectrum of oncology and microbiome research, and other emerging applications, more scientists are looking to harness the power of NGS technology.
Illumina is committed to upholding a proven track record of providing solutions that empower scientists to advance and accelerate their sequencing goals. For decades, Illumina has led the charge to improve sequencing capabilities by expanding benchtop applications, streamlining workflows, and maximizing accuracy.
With the NextSeq 1000 and NextSeq 2000 Sequencing Systems, Illumina has continued this commitment to innovation (Figure 1, Table 1). These market-leading benchtop platforms are powered by high-performance Illumina sequencing by synthesis (SBS) chemistry and onboard integrated informatics. Users can obtain the throughput, data quality, and cost required to meet their needs, from small to large batch sizes and simple to data-intensive applications, all on a benchtop system. Add in unwavering expert support and labs are set for today’s workload and future emerging applications.
Robust performance and streamlined workflow
High-performance SBS chemistry
The NextSeq 1000 and NextSeq 2000 Systems are powered by XLEAP-SBS chemistry, a faster, higher quality, and more robust SBS chemistry built on the proven foundation of standard Illumina SBS chemistry.
XLEAP-SBS nucleotides use state-of-the-art dyes and novel linkers and blocks that are more resistant to heat, show a 50× reduction in hydrolysis, and 2.5× faster block cleavage to reduce phasing and prephasing.
The XLEAP-SBS polymerase is engineered to incorporate nucleotides faster and with higher fidelity than ever before.
Faster turnaround times with XLEAP-SBS chemistry can result in full-day time savings* compared to run times with standard SBS.
Innovations enabling productivity gains
The NextSeq 1000 and NextSeq 2000 Systems leverage advances in optics, instrument design, and reagent chemistry to miniaturize the volume of the sequencing reaction while increasing output and reducing the cost per run. This miniaturization provides scalability for various output quantities while maintaining the same high standards of data quality as experienced with other Illumina platforms. Additional innovations include high cluster–density patterned flow cells and a super-resolution optics system that yields highly accurate imaging data with greater resolution and higher sensitivity than traditional benchtop systems.
The NextSeq 1000 and NextSeq 2000 Systems offer a simplified workflow with load-and-go ease (Figure 2).
Included onboard DRAGEN hardware powers fast secondary analysis and data compression. The instrument includes onboard cluster generation and prepares FASTq files as part of the sequencing run. No extra equipment or steps are required.
* For example, times for a P3 flow cell 2 × 150 bp run are eight hours faster with XLEAP-SBS reagents vs standard SBS.
Figure 1: The NextSeq 2000 Sequencing System—The NextSeq 2000 System offers innovative design features, advanced chemistry, simplified bioinformatics, and an intuitive workflow that enable the widest range of applications and flexibility of scale on a benchtop sequencing system.
RFQ No. FA520924Q0071
For Research Use Only. Not for use in diagnostic procedures. | 3M-NA-00008 v6.0
Table 1: Performance parameters for the NextSeq 1000 and NextSeq 2000 Systems
Read length NextSeq 1000/2000 P1 XLEAP-SBS Reagentsa
NextSeq 1000/2000 P2 XLEAP-SBS Reagentsa
NextSeq 2000 P3 XLEAP-SBS Reagentsa,b
NextSeq 2000 P4 XLEAP-SBS Reagentsb
Output per flow cellc
Single-end reads 100M 400M 1.2B 1.8B
1 × 50 bp – – – 90 Gb
2 × 50 bp 10 Gb 40 Gb 120 Gb 180 Gb
2 × 100 bp – 80 Gb 240 Gb 360 Gb
2 × 150 bp 30 Gb 120 Gb 360 Gb 540 Gb
2 × 300 bp 60 Gb 240 Gb – – quality scoresd
1 × 50 bp, 2 × 50 bp, 2 × 100 bp, 2 × 150 bp ≥ 90% of bases higher than Q30
2 × 300 bp ≥ 85% of bases higher than Q30
Run time
1 × 50 bp – – – 12 hr
2 × 50 bp 8 hr 12 hr 18 hr 20 hr
2 × 100 bp – 19 hr 31 hr 34 hr
2 × 150 bp 17 hr 22 hr 40 hr 44 hr
2 × 300 bp 34 hr 42 hr – –
a. XLEAP-SBS reagents for P1, P2, and P3 flow cells available Q2 2024.
b. P3 and P4 flow cells only available on the NextSeq 2000 System.
c. Output specifications based on a single flow cell using an Illumina PhiX control library at supported cluster densities.
d. Quality scores are based on an Illumina PhiX control library; performance may vary based on library type and quality, insert size, loading concentration, and other experimental factors.
Manage Prepare Sequence Analyze Monitor
Track samples and manage runs easily and efficiently with cloud-based and manual options
Prepare libraries with a wide ecosystem of compatible library preparation kits
Load prepared libraries into the sequencing cartridge and insert into the NextSeq 1000 or NextSeq 2000 System
Analyze using onboard DRAGEN secondary analysis or stream to the cloud for additional analysis flexibility
Increase lab efficiency with Illumina Proactive instrument performance service
Figure 2: Intuitive library-to-analysis workflow—The NextSeq 1000 and NextSeq 2000 Systems provide a comprehensive workflow that includes user-friendly run setup, a wide ecosystem of compatible library prep kits, load-and-go operation, and integrated, onboard secondary analysis.
RFQ No. FA520924Q0071
For Research Use Only. Not for use in diagnostic procedures.4 | M-NA-00008 v6.0
Easy-to-use integrated platform
The NextSeq 1000 and NextSeq 2000 Systems take advantage of an integrated reagent cartridge that includes fluidics and the waste holder, simplifying library loading and instrument use (Figure 3). Thaw and prepare the reagent cartridge, load the flow cell and prepared libraries into the cartridge, and then insert the assembled cartridge into the instrument. Denaturation and dilution steps occur onboard automatically.
In addition to ease of use, the fully integrated cartridge design improves efficiency throughout the sequencing run. By miniaturizing many of the sequencing reactions, the unique design:
• Lowers operating costs
• Improves recyclability
• Minimizes waste volume
Reagents never leave the cartridge, resulting in a dry instrument design that does not require washing, enables streamlined instrument maintenance, and optimizes instrument efficiency.
Flexibility to do more, scalability to grow more The NextSeq 1000 and NextSeq 2000 Systems offer broad platform utility with the flexibility to expand applications and scale efficiently. With fast, accurate results, outputs from 10 Gb up to 540 Gb, and integrated informatics options, the NextSeq 1000 and NextSeq 2000 Systems are ideal for a wide range of applications (Table 2, Table 3) across metagenomics, spatial transcriptomics, single-cell studies, and more.
Breadth of applications
The NextSeq 2000 System offers four flow cell types (P1, P2, P3, and P4) and 14 kit configurations, while the NextSeq 1000 System† offers two flow cell types (P1 and P2) and seven kit configurations (Table 1). Read lengths from 1 × 50 bp to 2 × 300 bp support a wide variety of applications and study sizes (Table 2):
• 600-cycle kits for read length–dependent applications such as de novo assembly of small whole genomes, targeted metagenomics, shotgun metagenomics, meta-transcriptomics, and immune repertoire profiling
• 300-cycle or 200-cycle kits for sequencing of targeted panels, exomes and large panels, transcriptomes, circulating tumor DNA, and large whole genomes
• 100-cycle or 50-cycle kits for counting-focused applications such as gene expression, single-cell, and spatial profiling
The NextSeq 2000 P4 flow cell delivers up to
1.8 billion single-end reads per run, maximizing batching and powering data-rich projects like exome sequencing, multiomics, single-cell, and spatial analysis.
† To ensure future scalability, customers who purchase a NextSeq 1000 System can easily upgrade to the NextSeq 2000 System.
Figure 3: NextSeq 1000 and NextSeq 2000 reagent cartridge— The integrated cartridge includes reagents, fluidics, and the waste holder. Simply thaw and prepare the reagent cartridge, insert the flow cell, load the library, and place into the instrument.
RFQ No. FA520924Q0071
For Research Use Only. Not for use in diagnostic procedures. | 5M-NA-00008 v6.0
Streamlined analysis options
The NextSeq 1000 and NextSeq 2000 Systems offer access to onboard, local, and cloud-based analysis software, giving users the flexibility to analyze data in a manner that meets their needs (Figure 4).
Runs can be set up locally or in the cloud. For local setup, users can create their own sample sheet or take advantage of a convenient, preset Illumina template. Cloud-based setup uses the Run Planner app in BaseSpace™ Sequence Hub. Once the run setup information is ready, it is imported into the NextSeq 1000 and NextSeq 2000 Systems. Users then select and start the run of interest.
Optimized instrument software provides a cleaner interface and improved visualization of instrument and run status compared to earlier benchtop systems.
Table 2: Some of the broader applications available on the NextSeq 1000 and NextSeq 2000 Systems
Applicationa
P1 XLEAP-SBSb P2 XLEAP-SBSb P3 XLEAP-SBSb P4 XLEAP-SBS
No.
samples Time No.
samples Time No.
samples Time No.
samples Time
Small whole-genome sequencing (300 cycles) 130 Mb genome, > 30× coverage 7 17 hr 30 22 hr 92 40 hr 138 44 hr
Whole-exome sequencing (200 cycles) ~8 Gb per exome, 100× mean coverage ~2c 17 hrc 10 19 hr 30 31 hr 45 34 hr
Total RNA-Seq (200 cycles) 50M read pairs per sample 2c,d 17 hrc 8 19 hr 24 31 hr 36 34 hr mRNA-Seq (200 cycles) 25M read pairs per sample 4c,d 17 hrc 16 19 hr 48 31 hr 72 34 hr
Single-cell RNA-Seq (100 cycles) 5K cells, 20K reads per cell 1e 8 hr 4 12 hr 12 18 hr 18 20 hr miRNA-Seq or small RNA analysis (50 cycles) 11M reads per sample 9f 8 hrf 36f 12 hrf 108f 18 hrf 163 12 hr
16S RNA sequencing (600 cycles) 384g 34 hr 384g 42 hr – – – –
a. Recommended sequencing depth will largely depend on sample type and experimental objective and will need to be optimized for each study.
b. XLEAP-SBS reagents for P1, P2, and P3 flow cells available Q2 2024.
c. 200-cycle kits not available for P1 flow cells. Use P1 300-cycle kit.
d. Recommended read lengths are 2 × 75 bp for Illumina Stranded Total RNA Prep and Illumina Stranded mRNA Prep and 2 × 100 bp for Illumina RNA Prep with Enrichment.
e. P1 reagents are a good option for single-cell quality control experiments.
f. 50-cycle kits not available for P1, P2, or P3 XLEAP-SBS flow cells. Use 100-cycle kits.
g. A maximum of 384 unique dual indexes are available.
The NextSeq 1000 and NextSeq 2000 Systems output industry-standard file formats used by various laboratory information management systems (LIMS) for secure, automated sample tracking and information management.
Secondary analysis can be configured as part of run setup, reducing the number of user touchpoints required.
Accurate and efficient analysis with onboard DRAGEN secondary analysis
Onboard DRAGEN (Dynamic Read Analysis for GENomics) secondary analysis offers an accurate, efficient solution for variant calling. The DRAGEN platform uses optimized, hardware-accelerated algorithms for a wide variety of genomic analysis solutions, including base call (BCL) file conversion, compression, mapping, alignment, sorting, duplicate marking, and variant calling.
RFQ No. FA520924Q0071
For Research Use Only. Not for use in diagnostic procedures.6 | M-NA-00008 v6.0
Table 3: DRAGEN informatics pipelines integrated into the NextSeq 1000 and NextSeq 2000 Systems
Pipelinea Applications
DRAGEN Enrichment • Whole-exome sequencing
• Targeted resequencing
DRAGEN RNA • Whole-transcriptome gene expression
• Gene fusion detection
DRAGEN Single-Cell
RNA
• Single-cell whole-transcriptome sequencing
DRAGEN Germline • Whole-genome sequencing
DRAGEN Ampliconb • DNA amplicon panels
• Targeted resequencing
DRAGEN ORAc Compression • Compression of FASTQ files
a. Additional DRAGEN informatics pipelines are available in the cloud; visit illumina.com/DRAGEN for a complete list.
b. Supported for DNA samples only.
c. ORA, original read archive; DRAGEN ORA Compression can be enabled with any of the DRAGEN pipelines.
The onboard solution provides access to select DRAGEN informatics pipelines (Table 3), enabling users to generate results in as little as two hours. DRAGEN informatics use best-in-class pipeline algorithms to help users overcome bottlenecks in data analysis and reduce reliance on external informatics experts. Onboard DRAGEN analysis is included in the instrument cost and does not require the purchase of an additional license.
Automatic integration with the cloud
For users who prefer a cloud-based analysis solution, the NextSeq 1000 and NextSeq 2000 Systems are integrated with BaseSpace Sequence Hub and Illumina Connected Analytics, our genomics cloud-computing platforms. DRAGEN analysis on BaseSpace Sequence Hub and Illumina Connected Analytics features advanced results visualization and analysis capabilities packaged in an intuitive interface. In the cloud, users can access a wide selection of bioinformatics tools and share data globally. Data generated using the NextSeq 1000 and NextSeq 2000 Systems are compatible with industry-standard formats for convenient import into the solution of choice.
Configuration options
Workflow management
Sequencing and primary analysis
Alignment and variant calling
Interpretation and data analytics
Local run setup with local analysis
Cloud-based run setup with local analysis
Cloud-based run setup with cloud-based analysis
Plan run in cloud and download sample sheet or generate sample sheet manually
Plan run in cloud, including analysis settings
Automated onboard secondary analysis with DRAGEN hardware-accelerated algorithms Access BaseSpace
Sequence Hub apps, instrument performance data, license management, and long-term storage solutions Automated cloud-based secondary analysis using standardized DRAGEN pipeline and workflow/task execution service
Figure 4: Flexible informatics suite—The NextSeq 1000 and NextSeq 2000 Systems feature local and cloud-based options for run setup, run management, and data analysis, enabling users to run their sequencing their way.
RFQ No. FA520924Q0071 Attachment 1 https://www.illumina.com/products/by-type/informatics-products/dragen-bio-it-platform.html
For Research Use Only. Not for use in diagnostic procedures. | 7M-NA-00008 v6.0
Proven global leader
The NextSeq 1000 and NextSeq 2000 Systems offer the certainty of a field-tested solution and a reliable sequencing partner. Since the launch in 2020, over two thousand instruments have been installed and used to generate thousands of publications. With a mature global quality and manufacturing infrastructure, Illumina provides comprehensive support and best-in-class product consistency, setting the standard for NGS solutions.
Joining this community gives access to a large ecosystem of applications, protocols, and informatics that have been built in collaboration with thousands of researchers and industry thought leaders across the globe.
Confidence in continued innovation
Illumina has a proven track record of developing genomics solutions that empower researchers to perform studies at the throughput, scale, and price meeting their research objectives. Customer experience is at the center of every innovation, making it as easy as possible to prepare samples, sequence, and analyze data.
Simple, fast support
Modular construction
The NextSeq 1000 and NextSeq 2000 Systems are built in a modular fashion, simplifying service and support.
Onboard sensors monitor system performance and alert users to possible issues. Troubleshooting and repairs are easier for the service engineer to perform, ultimately saving time.
Illumina Proactive
Illumina Proactive is a secure and remote instrument performance support service designed to detect risk failure preemptively, troubleshoot runs more efficiently, and prevent in-run failures. The complementary service helps minimize unplanned downtime and avoid unnecessary sample loss by anticipating repairs and alerting Illumina field personnel to schedule maintenance visits.
Commitment to your success
Each system purchase includes a one-year service warranty. Comprehensive maintenance, repair, and qualification solutions are also available. In addition, Illumina offers on-site training, ongoing support, phone consults, webinars, and courses at various Illumina locations globally. We’re here with all the resources that you need to accelerate progress.
Our world-class support team is composed of experienced scientists who are experts in library prep, sequencing, and analysis to help you maximize your investment and support peak performance. This dedicated team includes highly qualified field service engineers (FSE), technical applications scientists (TAS), field application scientists (FAS), system support engineers, bioinformaticians, and IT network experts, all deeply familiar with NGS and the applications that Illumina customers perform around the globe. Technical support is available via phone five days a week or access online support 24/7, worldwide and in multiple languages.
Summary
The NextSeq 1000 and NextSeq 2000 Systems expand what can be accomplished with a benchtop sequencing system. High flexibility and scalability, combined with enhanced performance using XLEAP-SBS chemistry, enable a wide range of applications. The NextSeq 1000 and NextSeq 2000 Systems offer proven NGS technology with operational simplicity, a streamlined workflow including analysis, and comprehensive support.
Learn more
NextSeq 1000 and NextSeq 2000 Sequencing Systems
RFQ No. FA520924Q0071 Attachment 1 Date: 15 Aug 2024 https://www.illumina.com/company/contact-us.html https://www.illumina.com/systems/sequencing-platforms/nextseq-1000-2000.html
For Research Use Only. Not for use in diagnostic procedures. | 8M-NA-00008 v6.0
1.800.809.4566 toll-free (US) | +1.858.202.4566 tel techsupport@illumina.com | www.illumina.com
© 2024 Illumina, Inc. All rights reserved. All trademarks are the property of Illumina, Inc. or their respective owners. For specific trademark information, see www.illumina.com/company/legal.html.
M-NA-00008 v6.0
NextSeq 1000 and NextSeq 2000 System specifications
Specifications Instrument configuration Self-contained, dry instrument with integrated DRAGEN field-programmable gate array (FPGA) secondary analysis
Instrument control computer Base unit: 2u Microserver located inside the instrument Memory: 288 GB Hard Drive: 3.8 TB SSD Operating System: Linux CentOS 7.6
Operating environment Temperature: 15°C–30°C Humidity: 20%–80% relative humidity, noncondensing Altitude: 0–2000 meters For Indoor use Only
Laser Wavelengths: 449 nm, 523 nm, 820nm Safety: Class 1 Laser Product
Dimensions W × D × H: 55 cm × 65 cm × 60 cm Weight: 141 kg Crated Dimensions Crated W × D × H: 92 cm × 120 cm × 118 cm Crated Weight: 232 kg
Power requirements Instrument Input Voltage: 100 VAC to 240 VAC Instrument Input Frequency: 50/60 Hz
Bandwidth for network connection 200 MB/s/instrument for internal network uploads 200 MB/s/instrument for BaseSpace Sequence Hub uploads 5 MB/s/instrument for Instrument Operational Data uploads
Product safety and compliance NRTL certified IEC 61010-1 CE marked FCC/IC approved
Ordering information
Product Catalog no.
NextSeq 2000 Sequencing System 20038897
NextSeq 1000 Sequencing System 20038898
NextSeq 1000 to NextSeq 2000 upgrade 20047256
NextSeq 1000/2000 P1 XLEAP-SBS Reagent Kit (100 cycles)a,b 20100983
NextSeq 1000/2000 P1 XLEAP-SBS Reagent Kit (300 cycles)a,b 20100982
NextSeq 1000/2000 P1 XLEAP-SBS Reagent Kit (600 cycles)a,b 20100981
NextSeq 1000/2000 P2 XLEAP-SBS Reagent Kit (100 cycles)a,b 20100987
NextSeq 1000/2000 P2 XLEAP-SBS Reagent Kit (200 cycles)a,b 20100986
NextSeq 1000/2000 P2 XLEAP-SBS Reagent Kit (300 cycles)a,b 20100985
NextSeq 1000/2000 P2 XLEAP-SBS Reagent Kit (600 cycles)a,b 20100984
NextSeq 2000 P3 XLEAP-SBS Reagent Kit (100 cycles)a,b 20100990
NextSeq 2000 P3 XLEAP-SBS Reagent Kit (200 cycles)a,b 20100989
NextSeq 2000 P3 XLEAP-SBS Reagent Kit (300 cycles)a,b 20100988
NextSeq 2000 P4 XLEAP-SBS Reagent Kit (50 cycles)a 20100995
NextSeq 2000 P4 XLEAP-SBS Reagent Kit (100 cycles)a 20100994
NextSeq 2000 P4 XLEAP-SBS Reagent Kit (200 cycles)a 20100993
NextSeq 2000 P4 XLEAP-SBS Reagent Kit (300 cycles)a 20100992
NextSeq 1000/2000 XLEAP-SBS Read and Index Primersa 20112856
NextSeq 1000/2000 XLEAP-SBS Index Primer Kita 20112858
NextSeq 1000/2000 XLEAP-SBS Read Primer Kita 20112859
a. XLEAP-SBS reagent kits for the NextSeq 1000 and NextSeq 2000 instruments are shipped and stored at the same temperature as standard SBS reagent kits.
b. XLEAP-SBS reagents for P1, P2, and P3 flow cells available Q2 2024.
RFQ No. FA520924Q0071