Laboratory_Fume_Hoods_Manual_1.pdf
PDF 2 MB Posted
- Attached to
- Fume Hood Removal and Installation Federal contract opportunity
- Solicitation number
- FA489719QA224
About this file
manual
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| PWS-Fume_Hood.docx | DOCX document | |
| COMBOSOL-19QA244-Fume_Hood_Installation.docx | DOCX document | |
| WG-19QA224-Fume_Hood.pdf | ||
| PROTECTOR_PREMIER_DIMENSIONS.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Original instructions
Protector Premier
Laboratory Fume Hoods
Models 10040 Series 10050 Series 10060 Series 10080 Series
Copyright © 2018 Labconco Corporation. The information contained in this manual and the accompanying products are copyrighted and all rights reserved by Labconco Corporation. Labconco Corporation reserves the right to make periodic design changes without obligation to notify any person or entity of such change.
Returned or Damaged Goods
Do not return goods without the prior authorization from Labconco. Unauthorized returns will not be accepted. If your shipment was damaged in transit, you must file a claim directly with the freight carrier.
Labconco Corporation and its dealers are not responsible for shipping damages.
The United States Interstate Commerce Commission rules require that claims be filed with the delivery carrier within fifteen (15) days of delivery.
Limitation of Liability
The disposal and/or emission of substances used in connection with this equipment may be governed by various federal, state, or local regulations. All users of this equipment are required to become familiar with any regulations that apply in the user’s area concerning the dumping of waste materials in or upon water, land, or air and to comply with such regulations. Labconco Corporation is held harmless with respect to user’s compliance with such regulations.
Contacting Labconco Corporation
If you have questions that are not addressed in this manual, or if you need technical assistance, contact
Labconco’s Customer Service Department or Labconco’s Product Service Department at 1-800-821-5525 or 1-816-333-8811, between the hours of 7:30 a.m. and 5:30 p.m., Central Standard Time.
Part #9410500, Rev. G
ECO L354
Warranty
Labconco Corporation provides a warranty to the original buyer for the repair or replacement of parts and reasonable labor as a result of normal and proper use of the equipment with compatible chemicals. Broken glassware and maintenance items, such as filters, gaskets, light bulbs, finishes and lubrication are not warranted. Excluded from warranty are products with improper installation, erratic electrical or utility supply, unauthorized repair and products used with incompatible chemicals.
The warranty for Protector® Premier® Laboratory Fume Hoods will expire one year from date of installation or two years from date of shipment from Labconco, whichever is sooner. Warranty is non-transferable and only applies to the owner (organization) of record.
Buyer is exclusively responsible for the set-up, installation, verification, decontamination or calibration of equipment. This limited warranty covers parts and labor, but not transportation and insurance charges. If the failure is determined to be covered under this warranty, the dealer or Labconco Corporation will authorize repair or replacement of all defective parts to restore the unit to operation. Repairs may be completed by 3rd party service agents approved by Labconco Corporation. Labconco Corporation reserves the rights to limit this warranty based on a service agent’s travel, working hours, the site’s entry restrictions and unobstructed access to serviceable components of the product.
Under no circumstances shall Labconco Corporation be liable for indirect, consequential, or special damages of any kind. This warranty is exclusive and in lieu of all other warranties whether oral, or implied.
TABLE OF CONTENTS
CHAPTER 1: INTRODUCTION 1
About This Manual 2
Typographical Conventions 3
(Conventions Typographiques)
Your Next Step 4
CHAPTER 2: PREREQUISITES 5
Location Requirements 6
Support Requirements 6
Exhaust Requirements 6
Electrical Requirements 7 (Exigences électriques)
Service Line Requirements 8
Space Requirements 8
CHAPTER 3: GETTING STARTED 9
Unpacking Your Fume Hood 10
Removing the Shipping Skid 10
Sash Weight Release 11
Installing the Hood on a Supporting Structure & Work Surface 11
Connecting to the Hood Exhaust System 13
Connecting the Electrical Supply Source 13
Connecting the Service Lines to the Protector Hood 16
Sealing the Protector Hood to the Work Surface 17
Certifying the Protector Fume Hood 17
CHAPTER 4: PERFORMANCE FEATURES AND SAFETY
PRECAUTIONS 18
Performance Features 18
Safety Precautions 21 (Consignes de sécurité)
CHAPTER 5: USING YOUR PROTECTOR PREMIER 25
Operating the Vertical Rising Sash 25
Operating the A-Style Combination Sash 25
Operating the Blower 25
Operating the Lights 26
Working in your Protector Fume Hood 26
CHAPTER 6: MAINTAINING YOUR PROTECTOR PREMIER 28
Service Safety Precautions 28 (Mesures de sécurité des services)
Routine Maintenance Schedule 29
Routine Service Operations 30
CHAPTER 7: MODIFYING YOUR PROTECTOR PREMIER 31
Installing Work Surfaces 31
Installing Ceiling Enclosures Above the Fume Hood 32
Installing Rear Panels Behind the Fume Hood 34
Installing Additional Service Fixtures 35
Installing Guardian Airflow Monitors 35
Distillation Grids – Field Installation 35
Sash Stop Kit – Field Installation 36
Upper Bypass – Field Installation 36
Installing an Electrical Duplex Outlet 36 (Installation d'une sortie Duplex électrique)
CHAPTER 8: TROUBLESHOOTING 37
APPENDIX A: PROTECTOR PREMIER COMPONENTS 40
APPENDIX B: PROTECTOR PREMIER DIMENSIONS 43
APPENDIX C: PROTECTOR PREMIER SPECIFICATIONS 45
(ANNEXE C: SPÉCIFICATIONS PROTECTEUR DU PREMIER MINISTRE)
APPENDIX D: SERIAL NUMBER TAG DESCRIPTION 48
(ANNEXE D: DESCRIPTION DE SÉRIE TAG NUMBER)
APPENDIX E: REFERENCES 54
Product Service 1-800-522-7658 1
CHAPTER 1
INTRODUCTION
Congratulations on your purchase of a Labconco Protector® Premier® Laboratory
Fume Hood. Your Protector Laboratory Fume Hood is designed to protect you. It is the result of Labconco’s more than 60 years of experience in manufacturing fume hoods, and users like you suggested many of its features to us.
The Labconco Protector Premier Fume Hood has been engineered to provide maximum visibility in a laboratory, and effectively contain toxic, noxious, or other harmful materials when properly installed. The Protector Premier offers many unique features to enhance safety, performance, and visibility. To take full advantage of them, please acquaint yourself with this manual and keep it handy for future reference. If you are unfamiliar with how fume hoods operate, please review Chapter 4: Performance Features and Safety Precautions before you begin working in the fume hood. Even if you are an experienced fume hood user, please review Chapter 5: Using Your Protector Premier, which describes your
Protector Hood’s features so that you can use the hood efficiently.
Figure 1-1
Chapter 1: Introduction
Product Service 1-800-522-7658 2
About This Manual
This manual is designed to help you learn how to install, use, and maintain your laboratory fume hood. Instructions for installing optional equipment on your hood are also included.
Chapter 1: Introduction provides a brief overview of the laboratory fume hood, explains the organization of the manual, and defines the typographical conventions used in the manual.
Chapter 2: Prerequisites explains what you need to do to prepare your site before you install your laboratory fume hood. Electrical and service requirements are discussed.
Chapter 3: Getting Started contains the information you need to properly unpack, inspect, install, and certify your laboratory fume hood.
Chapter 4: Performance Features and Safety Precautions explains how the
Protector operates and the appropriate precautions you should take when using the fume hood.
Chapter 5: Using Your Protector Premier discusses the basic operation of your fume hood. Information on how to prepare, use and shut down your Protector
Hood are included.
Chapter 6: Maintaining Your Protector Premier explains how to perform routine maintenance on your fume hood.
Chapter 7: Modifying Your Protector Premier explains how to modify the fume hood or add accessories.
Chapter 8: Troubleshooting contains a table of problems you may encounter while using your laboratory fume hood including the probable causes of the problems and suggested corrective actions.
Appendix A: Protector Premier Components contains labeled diagrams of all of the components of the fume hoods.
Appendix B: Protector Premier Dimensions contains comprehensive diagrams showing all of the dimensions for the laboratory fume hoods.
Appendix C: Protector Premier Specifications contains the electrical requirements for laboratory fume hood. Wiring diagrams are also included.
Appendix D: Serial Number Tag Description provides current rating code used on serial number tag.
Appendix E: References lists the various resources available that deal with laboratory fume hoods.
Product Service 1-800-522-7658 3
Typographical Conventions
Recognizing the following typographical conventions will help you understand and use this manual:
Book, chapter, and section titles are shown in italic type (e.g., Chapter 3:
Getting Started).
Steps required to perform a task are presented in a numbered format.
Comments located in the margins provide suggestions, reminders, and references.
Critical information is presented in boldface type in paragraphs that are preceded by the exclamation icon. Failure to comply with the information following an exclamation icon may result in injury to the user or permanent damage to fume hood.
Les informations critiques sont présentées en gras dans les paragraphes qui sont précédés par l'icône d'exclamation. Ne pas se conformer aux informations qui suivent une icône d'exclamation peut résulter à la blessure de l'utilisateur ou à des dommages irréversibles de la hotte aspirante.
Critical information is presented in boldface type in paragraphs that are preceded by the wrench icon. These operations should only be performed by a trained certifier or contractor. Failure to comply with the information following a wrench icon may result in injury to the user or permanent damage to your hood.
Les informations critiques sont présentées en gras dans les paragraphes qui sont précédés par l'icône de clé plate. Ces opérations devraient être seulement exécutées par un professionnel agrée. L'échec pour se conformer aux informations qui suivent une icône de clé plate peut résulter à la blessure de l'utilisateur ou à des dommages irréversibles de la hotte.
Important information is presented in capitalized type in paragraphs that are preceded by the pointer icon. It is imperative that the information contained in these paragraphs be thoroughly read and understood by the user.
A number icon precedes information that is specific to a particular model of laboratory fume hood. The 4' icon indicates the text is specific to the 4-foot wide model. The 5' icon indicates the text is specific to the 5-foot model, etc.
The S icon indicates the text is specific to the standard model.
The A icon indicates the text is specific to the A-Style Combination Sash
Model.
4' 5 '
6 '
S
A
8'
Product Service 1-800-522-7658 4
CAUTION – See Manual. When this symbol is on a fume hood it indicates a caution that is detailed in this manual.
PRUDENCE – Consulter le Manuel. Quand ce symbole est sur une hotte aspirante, il indique une prudence qui est détaillée dans ce manuel.
CAUTION – Hot Surface.
AVERTIR – Surface Chaude
CAUTION – See Manual. This symbol on the fume hood indicates the possibility of a pinch hazard.
PRUDENCE – Consulter le Manuel. Ce symbole sur la hotte indique la possibilité d'un risque de pincement.
Your Next Step If your Fume Hood needs to be installed, proceed to Chapter 2: Prerequisites to ensure your installation site meets all of the requirements. Then, go to Chapter 3:
Getting Started for instructions on how to install your laboratory fume hood and make all of the necessary connections.
If you would like to review how laboratory fume hoods operate, go to Chapter 4:
Performance Features and Safety Precautions.
For information on the operational characteristics of your laboratory fume hood, go to Chapter 5: Using Your Protector Premier.
If your laboratory fume hood is installed and you need to perform routine maintenance on the cabinet, proceed to Chapter 6: Maintaining Your Protector
Premier.
For information on making modifications to the configuration of your fume hood, go to Chapter 7: Modifying Your Protector Premier.
Refer to Chapter 8: Troubleshooting if you are experiencing problems with your
Product Service 1-800-522-7658 5
CHAPTER 2
PREREQUISITES
Before you install your laboratory fume hood, you need to prepare your site for installation. Carefully examine the location where you intend to install your hood. You must be certain that the area is level and of solid construction. In addition, a dedicated source of electrical power must be located near the installation site.
Carefully read this chapter to learn the requirements for your installation site:
The location requirements.
The support requirements.
The exhaust requirements.
The electrical power requirements.
The service line requirements.
The space requirements.
Refer to Appendix B: Protector Premier Dimensions for complete fume hood dimensions.
Refer to Appendix C: Protector Premier Specifications for complete laboratory fume hood electrical and environmental conditions, specifications and requirements.
Chapter 2: Prerequisites
Product Service 1-800-522-7658 6
Location Requirements
The fume hood should be located away from traffic patterns, doors, windows, fans, ventilation registers, and any other air-handling device that could disrupt its airflow patterns. All windows in the room should be closed.
La hotte aspirante devrait être localisé loin des voies de circulation, des portes, des fenêtres, des ventilateurs, des bouches de ventilation, et de tout appareil qui pourrait interrompre ses voies de flux d'air. Toutes les fenêtres dans la pièce devraient être fermées.
Support Requirements
DO NOT install the fume hood on a cart, dolly, or mobile bench. ALL Protector Hood installations must be permanent and stationary. The supporting structure usually consists of a base cabinet and chemically resistant work surface.
NE PAS installer la hotte aspirante sur un chariot ou un banc mobile. TOUTES les installations de la Hotte Protecteur doivent être permanentes et fixes. La structure de soutien consiste habituellement en un meuble doté d’une surface de travail chimiquement résistante.
Exhaust Requirements
The exhaust duct connection has been designed for 12" nominal duct (12.75" OD) to allow for minimum static pressure loss while operating at 60 to 100 fpm face velocities. The 12" diameter exhaust duct also allows for proper transport velocities away from the hood in the 1000 fpm to 2500 fpm range. The proper exhaust volume and static pressure loss are listed next for each hood model:
Product Service 1-800-522-7658 7
Proper blower selection can be determined from these exhaust requirements and the total system static pressure loss. Contact Labconco Customer Service for assistance in sizing a blower system.
Electrical Requirements The Protector Hood models feature internal wiring for the fluorescent light assembly and light switch. All internal wiring is terminated at the single point wiring junction box for hook-up by a qualified electrician. The blower switch, and light switch wires are also terminated at the single point wiring junction box for hook-up by a qualified electrician. Refer to Chapter 3: Getting Started and
Appendix C: Protector Premier Specifications for the wiring diagram for proper electrical installation.
Les modèles Protège-capot disposent câblage interne de l'assemblage de lumière fluorescente et interrupteur de lumière. Tout le câblage interne est terminé à la boîte de jonction point de câblage unique pour le raccordement par un électricien qualifié. Le bouton du ventilateur, et fils de l'interrupteur de lumière sont
Projected Airflows and Static Pressure
Eco Energy Saving Air Foil Flush Air Foil
Hood Type Face Velocity (fpm) Airflow Volumetric Rate (CFM) @ Airflow Volumetric Rate (CFM) @
Sash at Sash at 62.5% Static Pressure (inches of water) Static Pressure (inches of water)
Full open open 4' 5' 6' 8' 4' 5' 6' 8' at 28" at 18" Hood Hood Hood Hood Hood Hood Hood Hood
PREMIER 100 160 725, 0.22" 955, 0.31" 1180, 0.41" 1640, 0.28" 760, 0.23" 1000, 0.33" 1235, 0.43" 1715, 0.30"
80 128 580, 0.14" 765, 0.20" 945, 0.26" 1310, 0.18" 605, 0.15" 800, 0.21" 990, 0.28" 1370, 0.19"
60 96 435, 0.08" 575, 0.11" 710, 0.15" 985, 0.10" 455, 0.08" 600, 0.12" 745, 0.16" 1030, 0.11"
50 80 365, 0.06" 480, 0.08" 590, 0.10" 820, 0.07" 380, 0.06" 500, 0.08" 620, 0.11" 860, 0.08"
40 64 290, 0.04" 380, 0.05" 470, 0.07" 655, 0.05" 305, 0.04" 400, 0.05" 495, 0.07" 685, 0.05"
Eco Energy Saving Air Foil Flush Air Foil
Hood Type Face Velocity (fpm) Airflow Volumetric Rate (CFM) @ Airflow Volumetric Rate (CFM) @
Sash at Sash at Static Pressure (inches of water) Static Pressure (inches of water)
62.50% Full open 4' 5' 6' 8' 4' 5' 6' 8' open at 18" at 28" Hood Hood Hood Hood Hood Hood Hood Hood
PREMIER 100 62 450, 0.09" 595, 0.12" 735, 0.16" 1025, 0.11" 485, 0.09" 640, 0.13" 790, 0.17" 1100, 0.12"
80 50 365, 0.06" 480, 0.08" 590, 0.10" 820, 0.07" 390, 0.06" 515, 0.08" 635, 0.11" 880, 0.08"
60 38 270, 0.03" 360, 0.04" 440, 0.06" 615, 0.04" 290, 0.03" 385, 0.05" 475, 0.06" 660, 0.04"
Projected Airflows and Static Pressure
Eco Energy Saving Air Foil Flush Air Foil
Hood Type Face Velocity (fpm) Airflow Volumetric Rate (CFM) @ Airflow Volumetric Rate (CFM) @
Sash at Sash at 62.5% Static Pressure (inches of water) Static Pressure (inches of water)
Full open open 4' 5' 6' 8' 4' 5' 6' 8' at 28" at 18" Hood Hood Hood Hood Hood Hood Hood Hood
PREMIER 100 160 705, 0.21" 930, 0.29" 1150, 0.39" 1600, 0.27" 740, 0.22" 975, 0.31" 1205, 0.42" 1675, 0.29"
W/Bypass 80 128 565, 0.13" 745, 0.19" 920, 0.25" 1280, 0.17" 590, 0.14" 780, 0.20" 965, 0.27" 1340, 0.19"
Block 60 96 425, 0.08" 560, 0.10" 690, 0.14" 960, 0.10" 445, 0.08" 585, 0.11" 725, 0.15" 1005, 0.10"
50 80 350, 0.05" 465, 0.07" 575, 0.10" 800, 0.07" 370, 0.06" 490, 0.08" 605, 0.11" 840, 0.07"
40 64 280, 0.03" 370, 0.05" 460, 0.06" 640, 0.04" 295, 0.04" 390, 0.05" 480, 0.07" 670, 0.05"
Eco Energy Saving Air Foil Flush Air Foil
Hood Type Face Velocity (fpm) Airflow Volumetric Rate (CFM) @ Airflow Volumetric Rate (CFM) @
Sash at Sash at Static Pressure (inches of water) Static Pressure (inches of water)
62.50% Full open 4' 5' 6' 8' 4' 5' 6' 8' open at 18" at 28" Hood Hood Hood Hood Hood Hood Hood Hood
PREMIER 100 62 440, 0.08" 580, 0.11" 720, 0.15" 1000, 0.10" 475, 0.09" 625, 0.12" 775, 0.16" 1075, 0.11"
W/Bypass 80 50 350, 0.05" 465, 0.07" 575, 0.10" 800, 0.07" 380, 0.06" 500, 0.08" 620, 0.11" 860, 0.07"
Block 60 38 265, 0.03" 350, 0.04" 430, 0.06" 600, 0.04" 285, 0.03" 375, 0.05" 465, 0.06" 645, 0.04"
Product Service 1-800-522-7658 8 également mis fin à la boîte de jonction point de câblage unique pour le raccordement par un électricien qualifié. Reportez-vous au Chapitre 3: Mise en route et à l'Annexe C: Spécifications Premier Protector pour le schéma de câblage pour l'installation électrique correcte.
Service Line Requirements All service lines to the laboratory fume hood should be ¼ inch outside diameter, copper (brass for natural gas), and equipped with an easily accessible shut-off valve, should disconnection be required. Recommended operating pressure is 40
PSI, with a maximum allowable pressure of 200 PSI. Consider a pressure regulator to reduce line pressure to 40 PSI. Please check with local codes for other requirements.
Space Requirements The dimensions for the different models are shown in Appendix B: Protector
Premier Dimensions.
Product Service 1-800-522-7658 9
CHAPTER 3
GETTING STARTED
Now that the site for your laboratory fume hood is properly prepared, you are ready to unpack, inspect, install, and certify your unit. Read this chapter to learn how to:
Unpack and move your Protector Hood.
Set up the fume hood with the supporting structure and work surface.
Connect to an exhaust system.
Connect the electrical supply source.
Connect the service lines.
Sealing the Protector Hood to the work surface.
Arrange certification of your Protector Hood.
Depending upon which model you are installing, you may need common plumbing and electrical installation tools in addition to 5/16", 3/8", 7/16", and
1/2" wrenches, ratchets, sockets, a nut driver set, a flat-blade screwdriver, a
Phillips screwdriver, and a carpenter level to complete the instructions in the chapter.
The Protector Premier Hood models weigh between 400 to 800 lbs. (182-363 kg). The shipping skid allows for lifting with a mechanical lift truck or floor jack. If you must lift the fume hood manually, follow safe-lifting guidelines. Normally, the fume hood can be slid off a hydraulic lift table and be placed into position on top of the work surface. Do not lift by the front air foil.
Les premiers modèles de la Hotte Protecteur pèsent entre 400 à
800 livres. (182-363 Kg). La palette bois d’envoi permet le soulèvement par un camion muni d’un élévateur mécanique ou par un cric rouleur. Si vous devez soulever manuellement la hotte aspirante, respectez les règles de sécurité du soulèvement.
Normalement, la hotte aspirante peut être glissée d'une table munie d’un élévateur hydraulique et être placée en position sur la surface de travail. Ne pas soulever par l’écoulement d'air du devant.
Chapter 3: Getting Started
Product Service 1-800-522-7658 10
Unpacking Your Fume Hood
Carefully remove the shrink-wrap or carton on your fume hood and inspect it for damage that may have occurred in transit. If your unit is damaged, notify the delivery carrier immediately and retain the entire shipment intact for inspection by the carrier.
DO NOT RETURN GOODS WITHOUT THE PRIOR
AUTHORIZATION OF LABCONCO. UNAUTHORIZED
RETURNS WILL NOT BE ACCEPTED.
IF YOUR HOOD WAS DAMAGED IN TRANSIT, YOU MUST
FILE A CLAIM DIRECTLY WITH THE FREIGHT CARRIER.
LABCONCO CORPORATION AND ITS DEALERS ARE NOT
RESPONSIBLE FOR SHIPPING DAMAGES.
Do not discard the shipping skid or packing material for your fume hood until you have checked all of the components and installed and tested the unit. The Premier fume hood baffle is shipped installed as shown in Figure 1-1. Do not remove the fume hood from its shipping skid until it is ready to be placed into its final location.
Move the unit by placing a flat, low dolly under the shipping skid, or by using a floor jack.
Do not move the hood by tilting it onto a hand truck.
Ne pas déplacer la hotte en le penchant sur un diable.
Removing the Shipping Skid
LEAVE THE FUME HOOD ATTACHED TO ITS SHIPPING
SKID UNTIL IT IS AS CLOSE TO ITS FINAL LOCATION AS
POSSIBLE. MOVE THE HOOD BY USING A SUITABLE
FLOOR JACK, OR BY PLACING A FURNITURE DOLLY
UNDERNEATH THE SKID. DO NOT MOVE THE HOOD BY
TILTING IT ONTO A HAND TRUCK.
After you verify the fume hood components, move your hood to the location where you want to install it. Follow the steps listed next to remove the shipping skid from your unit.
1. Remove the side panels by unscrewing the Phillips screws.
2. Find the hardware (bolts, washers, nuts) that attach the fume hood to the skid and remove the hardware. Some hardware is on the sides and some is on the back.
The United
States
Interstate
Commerce
Commission rules require that claims be filed with the delivery carrier within fifteen
(15) days of delivery.
Product Service 1-800-522-7658 11
Sash Weight Release To protect the fume hood from damage in shipment, the sash weight has been secured to the back of the fume hood with screws. Simply remove the screws and make sure the sash cables or chains are on the pulleys or sprockets before operation of the sash. On models with more than one sash, the sash weights have been secured to the shipping skid with lag screws. Remove the weights from the skid and attach them to the respective sash cables using the hooks provided.
NOTE: THE SASH WEIGHT ITSELF WAS INDIVIDUALLY
MATCHED FOR THIS SPECIFIC HOOD AND SHOULD NOT
BE EXCHANGED ON ANY OTHER UNIT.
Installing the Hood on a Supporting Structure and Work Surface
The Protector Hood is heavy! Use caution when lifting or moving the unit.
La Hotte Protecteur est lourd ! Prudence en soulevant ou en déplaçant l'objet.
When installing the Protector Hood onto a chemically-resistant work surface or benchtop, ensure that the structure can safely support the combined weight of the fume hood and any related equipment. The 30" work surface should be at least as wide as the hood to properly support it. The work surface is aligned flush with the back of the fume hood for good airflow: this will provide the correct spacing under the air foil for proper bypass airflow. The lower base cabinets are placed flush with the front of the work surface as shown in Figure 3-1.
WARNING: It is important to support the rear of the work surface and fume hood. The cross support provides support for the bottom of the work surface. Install the cross support after the base cabinets and work surface are leveled and before installing the hood.
AVERTISSEMENT : Il est important de soutenir l'arrière de la surface de travail et la hotte aspirante. Le support tranversal soutient le bas de la surface de travail. Installer le support transversal après que les meubles et la surface de travail soient nivelés et avant d'installer la hotte.
Product Service 1-800-522-7658 12
The following are instructions for mounting a cross support:
1. Level the base cabinets and the work surface. Work surface should be placed flush with the back of the fume hood as shown in Figure 3-1.
2. Scribe a line on the wall or back of the base cabinet to locate the support under the work surface.
3. Mount the support by attaching it to the wall or base cabinet.
4. Place the hood on top of the work surface and cross support.
The work surface should be smooth and durable, such as a chemically-resistant epoxy resin. The surface should be nonporous and resistant to the acids, solvents, and chemicals used in conjunction with the Protector Fume Hood. The work surface should also contain a dished recessed area for containing primary spills.
Figure 3-1
Dimensions Shown in Inches
Hood Depth “B”
Hood Internal Depth “D”
Filler Panel Depth
“E” Work Surface
Depth
31.7" 24" 8" 30"
Place Cross
Support Here
1.7
Product Service 1-800-522-7658 13
Connecting to the Hood Exhaust System
WARNING: The weight of the exhaust ductwork system must be supported independently of the hood superstructure. Do not allow this weight to be supported by the hood structure as damage to the hood may occur.
AVERTISSEMENT : Le poids du système d'aspiration de ductwork doit être soutenu d'une manière indépendante de la superstructure de la hotte. Au cas où ce poids est supporté par la structure de la hotte, des dommages à la hotte peuvent arriver.
The exhaust connection should be installed by a qualified
HVAC contractor. The exhaust connection on your hood has been designed for 12" nominal pipe (12.75" OD) to allow for minimum static pressure loss with proper transport velocities away from the hood. Consult Labconco Customer Service should you require help sizing your blower for the exhaust volume and total system static pressure loss.
La connexion d'échappement devrait être installée par un professionnel de CVC agrée. La connexion d'aspiration sur votre hotte a été conçue pour un tuyau nominal de 12 pouces
(12,75 pouces de diamètre externe) afin d’avoir une perte minimale de pression statique avec les correctes flux de transport loin de la hotte. Consulter le Service Clientèle de
Labconco si la calibration de votre soufflerie pour le volume d'aspiration et la perte de pression statique du système le requiert.
The selected exhaust duct material should match the hood procedures and chemicals used to ensure compatibility.
Le matériel de conduit d'aspiration choisi devrait être en conformité avec les procédures de la hotte et les produits chimiques qui sont utilisés pour garantir la compatibilité.
Connecting the Electrical Supply Source to the Protector Fume Hood
Prior to connecting any electrical wiring to the fume hood structure, refer to the hood identification plate for the proper electrical requirements of your specific model.
Product Service 1-800-522-7658 14
WARNING: The building electrical supply system for
Protector Hoods is required to include overload protection. A switch or circuit breaker shall be in close proximity to the equipment and within easy reach of the operator. The switch or circuit breaker is to be marked as the disconnecting device for the equipment. Consult the NEC-2002 for proper installation.
AVERTISSEMENT : Le système d’alimentation électrique de la
Hotte Protecteur doit inclure la protection contre la surcharge. Un commutateur ou disjoncteur doit être tout près de l'équipement et à portée facile de l'opérateur. Le commutateur ou le disjoncteur doit être marqué comme l'appareil débranchant pour l'équipement. Consulter le NEC-2002 pour une installation correcte.
The identification plate, model number, serial number, and electrical connection boxes are accessible from the front of the fume hood by removing the front panel.
The Protector Premier Hood is normally wired for 115 Volt, 50/60 Hz, 20 Amp or
230Volt, 50/60 Hz, 10 Amp electrical service. Check the I.D. plate behind the front panel for voltage verification. The number of circuits varies depending on the model. All of the electrical connections are terminated at the field wiring terminal box for hook-up by a qualified electrician. We recommend each circuit be a dedicated branch circuit. However, if wired together the maximum load allowed is the sum of individual outlets plus the rating of the unit (i.e. 2 Amps remote, 8 Amps integral blower) not to exceed the 10A or 20A electrical service.
The single point internal junction box is used for the connection of the lights, blower, and duplex outlets. Refer to the wiring diagram for your Protector
Premier in Appendix C: Protector Premier Specifications.
The fume hood is required to be grounded to the MAINS protective earthing ground for safe operation. Using a ring terminal sized for a 10-24 machine screw, connect the MAINS ground conductor to the grounding lug marked with the protective earthing symbol, . Only MAINS ground conductors should be connected to the protective earthing ground lug, no other conductors should be connected to this grounding lug. Using wire nuts connect the MAINS supply conductors to the fume hood supply wires. Insure that the wires are connected as per the appropriate wire color codes for the input voltage. For 115V Phase (Hot) is black and Neutral is white, for 230V Phase1 is brown and Phase2 is blue. Refer to the wiring diagram for your Protector Premier in Appendix C: Protector Premier Specifications.
Product Service 1-800-522-7658 15
Figure 3-2
All wiring for the fume hood SHOULD be performed by a licensed electrician and conform to all local codes. In most cases, the hood will require the use of shielded conduit to protect the wiring into the hood. The grounding connection shall not be made to the terminal box cover.
Tout le câblage électrique pour la hotte aspirante devrait être exécuté par un électricien agrée et être conforme à tous les règles en vigueur. Dans la plupart des cas, la hotte exigera l'usage de conduit blindé pour protéger le câblage électrique dans la hotte. La prise de terre ne sera pas faite à la couverture de la boîte du terminal.
Internal Junction
Box
Product Service 1-800-522-7658 16
The fluorescent light has been mounted outside the top liner panel and is sealed from vapors inside the hood structure. To change the fluorescent light bulbs in your hood, you must first remove the front panel from the hood. Next remove the knock out plugs holding the light fixture in place. Lift fixture up and replace any deflective bulbs. Reverse order to reassemble.
Connecting the Service Lines to the Protector Fume Hood
The hoods with service fixtures have been plumbed from the valve to the hose connector or gooseneck for your installation convenience. Supply tubing shall be provided by the qualified installer. Tubing can enter the hood from above, through the back, or through the work surface to make these connections to the service fixtures.
NOTE: Inspect all fittings for leakage. Tighten the fittings slightly if needed.
NOTE : Inspecter toutes les installations à la recherche de fuite. Resserrer les installations légèrement si nécessaires.
CAUTION: Do not use oxygen with any standard service fixture. Contact Labconco Customer Service for oxygen fixture information.
PRUDENCE : Ne pas utiliser de l'oxygène avec l'accessoire de service standard. Contacter le Service Clientèle de Labconco pour les informations d'accessoire d'oxygène.
Should access to the hood plumbing fixture bodies be required, remove the service access plate on the hood front corner posts by loosening their individual screws (see item 11, Figure A-1 in Appendix A). The valve body will now be fully exposed for any service work that may be necessary. The service fixtures supplied on your laboratory hood are designed for use with the following services:
Air Hot Water Vacuum
Cold Water Natural Gas – See caution below
WARNING: Contact Labconco Customer Service directly before using any service other than those listed above in these valves to assure full compatibility.
AVERTISSEMENT : Contacter le Service Clientèle de
Labconco directement avant d'utiliser n'importe quel service autre que ceux énumérés au-dessus dans ces soupapes pour assurer une pleine compatibilité.
Product Service 1-800-522-7658 17
CAUTION: Natural gas should be used only in the service fixture that has been pre-plumbed with brass tubing. Sulfur content of the gas could cause deterioration of standard copper supply lines.
PRUDENCE : Le gaz naturel devrait être seulement utilisé dans l'accessoire de service qui a été pré soudé avec des tuyaux de cuivre. Le contenu soufré du gaz pourrait causer la détérioration des lignes d’alimentation en cuivre standard.
Sealing the Protector Hood to the Work Surface
When the hood has been set in place, ducted, wired, and plumbed, it should be sealed at the work surface to prevent spilled materials from collecting under the walls of the hood. Materials such as silicone sealants are recommended to seal the hood structure.
Certifying the Protector Fume Hood
The combination of your laboratory hood, exhaust ductwork, and exhaust blower gives you the flexibility to change the airflow at the sash opening of your hood.
To determine the actual face velocity at the sash opening, airflow velocity readings will need to be taken. This should be done across the sash opening of the hood in accordance with the Industrial Ventilation Manual section on laboratory hoods (see Appendix E: References). Labconco recommends an average face velocity at the sash opening of 60 to 100 feet per minute. Consult
Chapter 2 for proper airflow volumes for your particular model.
Your Protector Fume Hood has been tested at the factory per ASHRAE 110-1995.
All hoods achieve an “as manufactured rating” of less than 0.05 part per million
(ppm) at 4 liters per minute (lpm); AM<0.05 (consult Labconco for individual fume hood ratings). For “field use” ASHRAE testing contact Labconco Sales
Engineering Team or Customer Service for a certified on-site contractor.
NOTE: Face velocity profiles and smoke testing should be done periodically to ensure safe performance.
NOTE : Les profils de flux frontal et les tests de fumée devraient être régulièrement faits pour garantir une utilisation en toute securité.
Product Service 1-800-522-7658 18
CHAPTER 4
PERFORMANCE FEATURES
AND SAFETY PRECAUTIONS
Performance Features:
The Protector Premier Laboratory Hood is designed to meet the needs of the laboratory scientist at OSHA approved velocities as low as 60 feet per minute.
The laboratory fume hood has been designed to effectively contain toxic, noxious, or other harmful materials when properly installed. A fully featured by-pass hood with baffle and air foil, this hood maintains safe airflow. Optional energy saving
A-Style Combination Sash models are also available. The hood features by-pass airflow design that promotes full containment as the sash is moved. Airflow is diverted behind the front panel and under the air foil to help control fluctuations in face velocity, which occur as the sash is closed.
1. Unique sash provides maximum visibility of 37.5" high while conserving energy by limiting sash travel to 28". Vertical-rising sash may be raised from a closed to 28" operating height. The fume hood features a containment-enhancing sash handle. Optional sash stops are available to limit sash height and reduce energy usage.
2. By-pass airflow design ensures containment as the sash moves.
3. Large usable interior work depth and interior height of 48" provides ample working space.
4. Baffle (not shown) directs airflow to the rear of the interior to provide efficient airflow. The baffle may be removed for cleaning purposes only.
5. Exterior access cover plates are removable for easy access to plumbing valves and sash adjustment hardware when access through the sides is not available.
Chapter 4: Performance Features & Safety Precautions
Product Service 1-800-522-7658 19
6. Lift-Away™ front panel provides easy access to electrical wiring, sash weights, and lighting fixtures.
7. Energy efficient fluorescent lighting is located behind a laminated safety glass shield mounted to the top of the hood. The factory-wired instant start T8 lighting is serviceable from outside the hood cavity. If needed, more energy efficient, 50,000 hour LED direct replacement T8 bulbs can be ordered. See Appendix A.
8. Low mounted, factory-wired light and blower switches are ADA compliant.
9. All Air Foils allow air to sweep the work surface for maximum containment. Airfoils are offered with either Eco-Energy Saving or Flush
Foil. The Clean-Sweep™ openings create a constant barrier from contaminants. In addition, should the operator inadvertently block the airflow entering, the air continues to enter from under the air foil and through the Clean-Sweep openings. See Appendix A for alternate PVC
Eco-Air Foil as PVC material resists corrosion from mineral acids.
10. Streamlined corner posts provide maximum visibility and the flexibility to add services after installation.
11. All hoods are factory prepared for up to 8 service fixtures.
12. Duplex electrical receptacles are mounted on the right and left corner posts as requested. Receptacles are factory-wired to hood single point junction box.
13. Shipped fully assembled and eliminates the need for costly onsite assembly.
14. Accessory Guardian™ Digital Airflow Monitor or Guardian Airflow
Monitor continuously monitors face velocity. An audio/visual alarm alerts the user to low airflow conditions. The right corner post is factory prepared to accommodate the Guardian Monitor (sold separately).
15. Optional Sash Models. Sashes are offered in smooth anti-racking cable, chain, auto-return, or Intelli-Sash.
16. Frame of epoxy-coated steel and aluminum is durable and corrosion resistant.
17. Exhaust Connection. The hood features 12" (12.75" OD pipe) exhaust connections sized to allow for a minimum static pressure loss through the hood structure while providing a good transport velocity through the exhaust system. 4' Integral Blower models use 10" (10.75" OD).
18. Spillstopper™ Solid Epoxy Work Surface is dished to contain spills.
(Work surface is sold separately).
19. Optional Ceiling Enclosure Kits are available for a decorative facade between the hood and the ceiling. Integral Blower models have a facade as shown in Figure 4-1.
Product Service 1-800-522-7658 20
20. Optional Sash Stops provide a means of controlling the operating height of the sash.
21. Integral Blower Models available as shown in Figure 4-1.
Figure 4-1
Note:
Integral Blower
Model Shown
Product Service 1-800-522-7658 21
See Diagram on Page 20.
22. CAUTION – Current rating of receptacle is specified in Appendix D.
AVERTIR – Classification des prises de courant est spécifié dans l’Annexe D.
23. CAUTION – See Appendix C and D for complete current rating.
AVERTIR – Voir Annexe C et D pour la classification complete du courant.
24. CAUTION – Hot Surface. Blower motor can attain temperature that could cause burns.
AVERTIR – Surface Chaude. Le moteur de soufflerie peut atteindre la température qui pourrait causer des brûlures.
25. CAUTION – Flammable Gas.
AVERTIR – Gaz Inflammable.
Safety Precautions Although the laboratory hood has been engineered to maintain optimum operator safety, caution should always be used while working in the hood. Prior to using the hood, check to make sure that the exhaust blower is operating and that air is entering the hood at its specified face velocity.
Bien que la hotte de laboratoire ait été réglée pour maintenir la sécurité optimale de l’opérateur, la prudence devrait toujours être utilisée en travaillant sous la hotte. Avant utiliser la hotte, le contrôle pour s'assurer que la soufflerie d'aspiration fonctionne et que cet air entre dans la hotte au flux spécifié.
USE GOOD HOUSEKEEPING IN THE HOOD AT ALL TIMES.
CLEAN UP SPILLS IMMEDIATELY WITH A MILD DETERGENT.
PERIODICALLY CLEAN HOOD INTERIOR, INCLUDING
FLUORESCENT LIGHT GLASS PANEL. REPLACE BURNED OUT
LIGHT BULBS TO MAINTAIN MAXIMUM ILLUMINATION.
THE LAMP(S) IN THIS PRODUCT CONTAIN MERCURY
Manage in accordance with local disposal laws. DO NOT place lamps in trash.
Dispose as a hazardous waste. For information regarding safe handling, recycling and disposal, consult www.lamprecycle.org
CETTE LAMPE DANS CE PRODUIT CONTIENT DU MERCUE
Éliminez ou recyclez conformément aux lois applicables. Pour de l‘information concernant des pratiques de manipulation sécuritaires et l’élimination sécuritaire et le recyclage, veuillez consulter www.lamprecycle.org
Product Service 1-800-522-7658 22
DO NOT OVERLOAD THE WORK SURFACE WITH
APPARATUS OR WORK MATERIAL. THE SAFE
OPERATION OF THE LABORATORY HOOD IS BASED
UPON HAVING PROPER AIRFLOW THROUGH THE
STRUCTURE. DO NOT PLACE LARGE, BULKY OBJECTS
SUCH AS BLOCK HEATERS, DIRECTLY ON THE HOOD
WORK SURFACE. INSTEAD, ELEVATE THE OBJECT 2" TO
3" ON BLOCKS TO ALLOW A FLOW OF AIR UNDER THE
OBJECT AND INTO THE LOWER REAR BAFFLE EXHAUST
SLOT. ENSURE BLOCKS ARE LEVEL AND SECURED IN
PLACE.
Blocking the bottom of the baffle at rear of hood will change the airflow pattern in the hood causing turbulence and possible leakage at the face of the hood. (Don’t store containers or supplies against baffles, as this will affect airflow through the hood).
Avoid placing your head inside hood. Keep hands out of hood as much as possible.
Bloquer le fond du déflecteur à l'arrière de la hotte changera le modèle du flux d'air dans la hotte causant de la turbulence et une fuite possible devant la hotte. (Ne pas emmagasiner des récipients ou des provisions contre les déflecteurs, car ceci affectera le flux d'air à travers la hotte).
Eviter de placer votre tête à l’intérieur de la hotte. Garder les mains à l’extérieur de la hotte le plus possible.
Always work as far back in hood as possible. It is best to keep all chemicals and apparatus 6" inside the front of the hood.
Toujours travailler aussi loin que possible de la hotte. Il est recommandé de garder tous les produits chimiques et appareils à 6 pouces à l’intérieur de l’avant de la hotte.
This hood does not feature explosion-proof electrical components, unless ordered separately. Therefore, use of flammable or explosive materials in quantities above the explosive limit are not recommended.
Cette hotte ne possède pas de composants électriques anti-explosion, à moins que commandé séparément. Donc, l'usage de matériels inflammables ou explosifs dans les quantités au-dessus de la limite explosive n'est pas recommandé.
Product Service 1-800-522-7658 23
Do not work with chemicals in this hood without the exhaust system running. Do not store chemicals in a fume hood.
Ne pas travailler avec les produits chimiques sous cette hotte sans le système de d’aspiration en marche. Ne pas stocker des produits chimiques sous une hotte aspirante.
Perchloric acid use in this hood is prohibited.
L'usage d’acide perchlorique sous cette hotte est interdit.
High level radioisotope materials are prohibited for usage in this hood.
Les matériels d'isotope radioactif de haut niveau sont interdits à l'usage sous cette hotte.
AVOID CROSS DRAFTS AND LIMIT TRAFFIC IN FRONT OF
THE HOOD. AIR DISTURBANCES CREATED MAY DRAW
FUMES OUT OF THE HOOD.
The use of heat-generating equipment in this hood without the exhaust system operating properly can cause damage to the hood.
L'usage d'équipement chaleur-produiant dans ce capuchon sans l'opération de système d'aspiration peut causer convenablement des dommages à la hotte.
The Protector Laboratory Hood should be certified by a qualified certification technician before it is initially used. The hood should be re-certified whenever it is relocated, serviced or at least annually thereafter.
La Hotte de Laboratoire Protecteur devrait être certifié par un technicien de certification qualifié avant qu'elle soit utilisée au début. La hotte devrait être re-certifiée quand elle est réinstallée, entretenue ou du moins annuellement par la suite.
Ensure that the unit is connected to electrical service in accordance with local and national electrical codes. Failure to do so may create a fire or electrical hazard. Do not remove or service any electrical components without first disconnecting the hood from electrical service.
Product Service 1-800-522-7658 24
Garantir que l'unité est connectée au service électrique conformément aux règles électriques en vigueur. Le non-respect peut créer un feu ou un risque d'origine électrique. Ne pas enlever ou entretenir des composants électriques sans débrancher premièrement l’alimentation électrique de la hotte.
Proper operation of the fume hood depends largely upon the hood’s location and the operator’s work habits. Consult the
Reference Manual in Appendix E.
Le correct fonctionnement de la hotte d'aspiration dépend principalement de son emplacement et les habitudes de travail de l'opérateur. Consulter le Manuel de Référence dans l'Annexe E.
If the unit is not operated as specified in this manual it may impair the protection provided by the unit.
Si l'unité n'est pas utilisée comme spécifié dans ce manuel il peut diminuer la protection fournie par l'unité.
Do not touch the blower motor. The surfaces of the motor can become hot and could cause burns.
Ne pas toucher le moteur de soufflerie. Les surfaces du moteur peuvent devenir chaudes et pourrait causer des brûlures.
Do not position the fume hood so that it is difficult to operate the main disconnect device.
Ne pas positionner la hotte de sorte qu'il est difficile de faire fonctionner le dispositif principal de déconnexion.
To prevent the possibility of minor injury keep hands and fingers clear of sprockets at the top of the four corners.
Pour éviter les risques de blessures mineures garder les mains et les doigts de pignons en haut de la quatre coins.
Product Service 1-800-522-7658 25
CHAPTER 5
USING YOUR PROTECTOR
PREMIER
Operating the Vertical-Rising Sash
Because of the Protector Hood counterbalanced sash mechanism, it will take only a few pounds of force to move the sash up or down, and you can operate the sash smoothly with one or two hands positioned any where along the handle. The vertical-rising sash may be raised to a maximum 28" operating height. The airflow requirements should be sized for the 28" operating height; if using sash stops then the airflow requirements can be reduced by approximately 40% at 18".
See Chapter 2 for airflows. Optional models with auto-return to 18" or electric
Intelli-Sash are available.
Operating the A-Style Combination Sash Some other hood models have additional energy saving sashes called A-Style
Combination Sashes in place of vertical-rising sashes. These combination sashes allow the operator to use the hood with sashes either half open horizontally or vertically to conserve energy. The horizontal sashes are used in normal operating mode. Optional sets of sash stops can be installed to prevent raising the vertical sash above the half-open and fully-closed positions unless manually defeated by the operator. The airflow requirements are sized for the 50% open sash condition.
Operating the Blower Your Protector Fume Hood utilizes a remote style blower, which can be activated by turning the blower switch to “ON.” You can validate the hood performance by watching smoke drawn into the hood face opening.
S
A S
A
Chapter 5: Using Your Protector Premier
Product Service 1-800-522-7658 26
Operating the Lights Your Protector Fume Hood utilizes a factory-wired fluorescent light to illuminate the hood interior. Optional direct replacement LED T8 bulbs are available per
Appendix A. Simply turn the light switch to “ON” to operate.
Working in your Protector Fume Hood Planning
Thoroughly understand procedures and equipment required before beginning work.
Arrange for minimal disruptions, such as room traffic or entry into the room while the hood is in use.
Start-up
Turn on light and hood blower.
Slowly raise the sash.
Check the baffle air slots for obstructions.
Allow the hood to operate unobstructed for 5 minutes.
Wear a long sleeved lab coat and rubber gloves. Use protective eyewear.
Wear a protective mask if appropriate.
Loading Materials and Equipment
Only load the materials required for the procedure. Do not overload the hood.
Do not obstruct the front air foil, or rear baffle slots.
Large objects should not be placed close together and spaced above the work surface to permit airflow to sweep under the equipment.
After loading the hood, wait one minute to purge airborne contaminants from the work area.
Work Techniques
Keep all materials at least 6 inches inside of the sash, and perform all contaminated operations as far to the rear of the work area as possible.
Segregate all clean and contaminated materials in the work area.
Avoid using techniques or procedures that disrupt the airflow patterns of the hood.
Final Purging
Upon completion of work, the hood should be allowed to operate for two to three minutes undisturbed, to purge airborne contaminants from the work area before shutting down blower.
A S
Chapter 5: Using Your Protector Premier
Product Service 1-800-522-7658 27
Unloading Materials and…
This is the start of the file's text. The full file is on GovTribe.
File details come from the government source that posted it.