DEPARTMENT OF THE AIR FORCE

HEADQUARTERS 96TH TEST WING (AFMC)

EGLIN AIR FORCE BASE FLORIDA

MEMORANDUM FOR: AFTC/PZIOA 2 FEB 2018

FROM: 782 TS/RNWI

SUBJECT: Government Technical Requirement Document

1. The 782nd Test Squadron (782 TS) Seekers and Sensors Test Flight (RNWIA) has a requirement for Three (3) Commercial Off The Shelf (COTS) Blackbodies. The blackbodies must meet the following technical requirements.

a. The blackbodies must have a 4-inch square extended area.

b. The blackbodies must have a 50°C - 600°C Absolute Temperature Range.

c. The blackbodies must have an emissivity of 97% ± 3%.

d. The blackbodies must have NIST traceable calibration.

e. The blackbodies must have at least a 0.5°C temperature accuracy.

f. The blackbodies must have a 120VAC power supply.

2. The 782nd Test Squadron (782 TS) Seekers and Sensors Test Flight (RNWIA) has a requirement for one (1) Commercial Off The Shelf (COTS) Blackbody. The blackbody must meet the following technical requirements.

a. The blackbody must have a 4-inch square extended area.

b. The blackbody must have a (-20)°C – 150°C Absolute Temperature Range.

c. The blackbody must have an emissivity of 97% ± 3%.

d. The blackbody must have NIST traceable calibration.

e. The blackbody must have at least 0.5°C temperature accuracy.

f. The blackbody must be able to be calibrated by removable sensor replacement with NIST traceable calibration.

g. The blackbody must have a nitrogen hood for an inert atmosphere at low temperatures.

h. The blackbody must have a 120VAC power supply.

3. The 782nd Test Squadron (782 TS) Seekers and Sensors Test Flight (RNWIA) has a requirement for one (1) Commercial Off The Shelf (COTS) Calibration Kit. The calibration kit must meet the following technical requirements.

a. The calibration kit must have NIST traceable calibration.

b. The calibration kit must have a 50°C – 600°C Temperature Calibration Range.

c. The calibration kit must be able to update the high temperature blackbody controllers

(Item 1) with a new calibration.

d. The calibration kit must have a resolution of at least 0.001°C.

e. The calibration kit must have a 120VAC power supply

MEMORANDUM FOR: AFTC/PZIOA 2 FEB 2018