19M-227A-MA
8 November 2018
Revision 0
PURCHASE DESCRIPTION
FOR
Mid Range Balance
1.0 SCOPE: This purchase description specifies a Mid Range Balance to be used in the calibration of mass standards by selected Precision Measurement Equipment Laboratories (PMELs) to calibrate both Avoirdupois and Metric field masses and torque calibration mass sets in the range of 0.5 to 5 pounds.
2.0 PERFORMANCE REQUIREMENTS:
2.1 Display Resolution: The Mid Range Balance shall be readable to at least 1 milligram.
2.2 Measuring Range: The Mid Range Balance shall have a continuous metric range, which covers 0.5 to 5 pounds minimum.
2.3 Repeatability: The repeatability of the Mid Range Balance shall not exceed 2 milligrams throughout the Measuring Range (see PD paragraph 2.2).
2.3.1 Repeatability Definition: Repeatability for this instrument is defined as the standard deviation of 6 sets of 2 pairs of differences between two master weights of the same nominal value (6 runs using ABBA method).
2.4 Reliability: The Mid Range Balance shall maintain its repeatability specification over a period of 12 months, minimum.
2.5 Shift Test/Eccentricity: The Shift Test error of the Mid Range Balance shall not exceed 15 milligrams throughout the Measuring Range (see PD paragraph 2.2).
2.6 Span: The Span error of the Mid Range Balance shall not exceed 50 milligrams at 5 pounds.
2.7 Stabilization Time: The stabilization time for an individual reading shall be no more than 20 seconds.
3.0 FUNCTIONAL REQUIREMENTS:
3.1 Manual Control: Mid Range Balance shall be self-contained and manually controlled via push button(s) or switch.
3.1.1 Internal/External Adjustment: Mid Range Balance shall have an internal, motorized substitution weight(s) and the ability to use an external substitution weight(s). Item’s internal adjustment function shall have the option to be both automated (settings configured by user) or user-initiated.
3.1.1.1 Internal Substitution Weight: When using the internal substitution weight(s) for adjustment, operator shall have the ability to edit the programmed value(s) for the internal weight(s).
3.1.1.2 External Substitution Weight: When using an external substitution weight(s) for adjustment, operator shall have the ability to enter (or edit) the actual mass value of the external weight(s).
3.2 Ancillary Equipment: Mid Range Balance shall have a rigid, separate, external draft shield, which provides adequate space around the Mid Range Balance and easy access to the weighing chamber. See Section 4.8 for additional requirements.
Attachment 1
Revision 0
3.2.1 Materials: External draft shield shall be composed primarily of at least 5 clear panes of Lexan, glass, or equivalent transparent material with ¼ +/- 1/16 inch thickness. Shield shall not be composed of Plexiglas due to inherent static properties.
3.2.2 Assembly: External draft shield shall come fully assembled, or be assembled in-situ by the manufacturer’s trained technician.
3.2.3 Functionality: External draft shield shall be placed on the Mid Range Balance’s supporting table and shield entire weighing portion of the Mid Range Balance, and allow for cords, such as power cords, to be positioned as needed. External shield shall function such that the operator is able to access the Mid Range Balance’s weighing pan through the internal draft shield in order to add and remove weights as necessary to perform measurements through the use of a sliding door or hinge mechanism.
3.2.4 Surface Area: Assembled closed draft shield shall not cover a surface area of more than 6 square feet.
3.3 Calibration Adjustment: Any special equipment, fixtures, tools, standards, etc., required for calibration and are not common to a secondary calibration laboratory must be supplied with the Mid Range Balance. A unit that fails calibration and is recommended to be aligned or adjusted in order to meet PD Paragraphs 2.5 and 2.6, either manually or using alignment/adjustment software, shall include any special hardware, software, software licenses needed to perform the alignment/adjustment process.
4.0 MECHANICAL & PHYSICAL REQUIREMENTS:
4.1 Components: Commercial parts and materials that are used shall have properties consistent with a precision laboratory instrument.
4.1.1 Weighing Pan: Mid Range Balance’s weighing pan/surface shall be made of stainless steel, or similar material, which resists rusting, chipping, flaking, or scraping. Weighing pan/surface shall not be composed of, or covered in, a soft material, such as cork, which may flake, disintegrate, or rub off over time.
4.2 Use of Best Commercial Practices: The equipment shall be designed and constructed of materials that meet the requirements of this specification and in accordance with the best commercial practices for industry approved equipment. It shall be designed and manufactured to maximize reliability, maintainability, accuracy of operation, and ease of operation. All parts (which include, but are not limited to, shafts, bearings, regulators, switches, controls) shall have proper clearance and adjustments. They shall work together so that the equipment will supply the rated requirements without unnecessary strains, vibrations, or overheating. They shall be able to withstand the conditions met in shipping, storage, installation, and service. All materials and parts of this equipment shall be capable of withstanding the environmental requirements of this specification.
4.3 Matched and Selected Components: These shall be kept to an absolute minimum. Use of specially built, nonstandard components shall be avoided.
4.4 Model and Serial Number Identification: Mid Range Balance’s Part Number, Manufacturer, and Serial Number shall be identified clearly on the item in a location visible to the operator. Unique item identification shall follow Item Identification and Valuation requirements as stated in the contract.
4.5 Operating Environment/Storage:
Revision 0
4.5.1 Operating Temperature: The Mid Range Balance shall be capable of operating in an ambient temperature of 67 to 79 degrees F without loss of required performance.
4.5.2 Operating Humidity: The Mid Range Balance shall be capable of operating in an atmosphere from 20% to 50% relative humidity without loss of required performance.
4.6 Facility Electrical Power and Connections: The Mid Range Balance shall be capable of operating on standard 110 V AC 60 Hz electrical power.
4.7 Safety: Documentation will be provided to support the means by which electrical safety considerations have been addressed. The safety of all personnel will be assured in conductive or nonconductive contact with the outside of the equipment, and during the changing of the fuse with the power to the system and its components are turned on. This documentation shall consist of a minimum of a citation of a UL or ANSI identified standard or equivalent United States Standard to which the equipment conforms.
4.8 Working Height: The Mid Range Balance’s weighing pan surface shall be located no more than 8 inches higher than the Mid Range Balance’s supporting table. In addition, when moving a 14.5 inch tall weight from the Mass Comparator’s supporting table and onto the weighing pan, the weight shall be lifted no more than 9 inches vertically, with no obstructions from the External Draft Shield (para 3.2).
5.0 OPERATOR’S MANUAL: Manual shall be IAW DD1423s and Data Item Descriptions, and shall include, at a minimum, the following information:
A. Setup/installation of balance including external draft shield B. Procedure on how to adjust the balance, if necessary, per section 3 C. Drawing/Diagrams needed for adjustment D. A list of peripheral support equipment required to accomplish the adjustments E. Information on any internal adjustments made during operation F. Definitions on changeable settings G. Procedure on how to package, store and transport the balance
6.0 TESTING: This item will be inspected and accepted at destination as stated in the contract. The item will be checked with test equipment with parameters equal to or better than those required to assure that specified performance parameters are met. Test equipment used shall have first echelon certification directly traceable to the National Institute of Standards Technology (NIST) or other AFMETCAL-approved equivalent international laboratory. Tests will be performed to assure that the item meets the requirements of this purchase description. Such testing at destination does not relieve the contractor of performing all inspections and quality checks at the point of manufacture as necessary to assure performance as specified herein.
Attachment:
Data Requirements
Attachment to Purchase Description 19M-227A-MA
Mid Range Balance
DATA REQUIREMENTS
MANUALS: A complete user and service manual and calibration procedure shall be provided with each unit in contractor format. Manuals shall be on CD/DVD-ROM in Indexed Portable Document Format (iPDF). The manuals shall comply with Data Item Description (DID) DI-TMSS-80527C and the Contract Data Requirements List (DD Form 1423).