This document is Attachment A, a detailed specification for a Hydrogen Sulfide Analyzer, which appears to be part of a state or local government procurement opportunity. The specification outlines comprehensive technical requirements for a portable analyzer capable of detecting hydrogen sulfide in concentrations ranging from 3 parts per billion (ppb) to 10 parts per million (ppm), with high-precision measurement capabilities. The analyzer must have a resolution of 20 parts per trillion (ppt), offer multiple measurement modes (survey, manual, auto), provide onboard data logging for 20,000 samples, include a liquid crystal display screen, and feature digital and analog output options.
The analyzer specification includes additional critical requirements such as a lightweight design (not exceeding 5.4 pounds), a rechargeable battery with an 18-hour life that charges in 3 hours, and a hard carrying case for transport. The device must provide multiple data output methods, including USB connectivity to computers, printers, or flash drives, and an analog 4-20 milliamp current loop output. The specification also mandates a one-year warranty for parts and labor, and requires vendors to detail the cost of any additional accessories needed for equipment operation. These precise technical requirements suggest a specialized scientific or environmental monitoring procurement, likely for a government agency or research institution involved in environmental testing or industrial safety.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Attachment A
Hydrogen Sulfide Analyzer Specification Hydrogen Sulfide Analyzer to include:
a. Detection range of 3 parts per billion(ppb) to 10 parts per million(ppm) with a resolution of 20 parts per trillion(ppt) dependent on range
b. Measures in ppm, ppb with accuracies:
i. Plus or minus 1 ppb at 5 ppb
ii. Plus or minus 3 ppb at 50 ppb
iii. Plus or minus 0.03 ppm at 0.5 ppm
iv. Plus or minus 0.3 ppm at 5.0 ppm
c. Response Time
i. Survey Mode: 12-27 seconds
ii. Manual Mode: 12-27 seconds
iii. Auto Mode: 12-53 seconds
d. Flow Rate of 150 Milliliters(mL)/min
e. Automatic date/time stamp
f. Ability to Autosample: 1, 2, 5, 10, 15, 30, 45, 60, 90, or 120 minutes
g. Onboard data logging – 20,000 samples
h. Liquid Crystal Display Screen
i. Output
i. Digital: Universal Serial Bus to personal computer, printer, or flash drive
ii. Analog: 4-20 Milliamps(mA) passive current loop; accurate to 0.3 percent of output
j. Not to exceed 5.4 pounds
k. Battery
i. Rechargeable battery
ii. 18 hour life
iii. charges in 3 hours
l. Hard carrying case for transport
m. Detail cost of any accessories needed to allow equipment to operate (if not included)
n. Warranty of 1 year for parts and labor