This is a product specification document for the ENDODRY™ Storage and Drying System, a Class I medical device manufactured by STERIS Corporation (Cantel Medical Italy S.r.l.). The system is designed for efficient drying and horizontal storage of up to eight endoscopes, utilizing filtered compressed air to achieve external drying in three hours and internal channel drying in one hour. The cabinet meets European standard EN 16442:2015 and provides validated endoscope storage for up to 31 days while maintaining patient-ready condition and reducing post-reprocessing microbial growth.

The ENDODRY system is available in single-sided or pass-through configurations with left or right door swing options (models DRY-1001 through DRY-2004). Key specifications include dimensions of 600 x 1900 x 540 mm with an operating weight of 170 kg, compressed air requirements of 6 mm ID tubing at 2–10 bar pressure, and electrical power of 100–240 V AC, 100 W. The system features a 6-inch color touch screen interface, SMART LED status indicators, barcode identification for traceability, locking doors, climate sensors, and integration capability with MEDIVATORS ADVANTAGE PLUS and DSD EDGE automated endoscope reprocessors. Additional features include optional SMART Connect licensing for enhanced data transfer between devices, network integration via Ethernet, report generation capabilities, and preventive maintenance support through STERIS service specialists. The document includes available accessories such as transport cassettes, silicon oil, O-rings, drawer bumpers, and USB adapter cables.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

ENDODRY™— 11059789 Rev. A 1

ENDODRY™

Series

APPLICATION

The ENDODRY™ Storage and Drying System is designed to ensure the efficient drying of endoscopes, both externally and internally, while providing horizontal storage for up to eight endoscopes.

Drying is achieved through a continuous flow of filtered compressed air, ensuring validated storage of endoscopes for up to 31 days. EN 16442:2015 validation and a peer-reviewed study in the American Journal of Infection Control showed the drying of external surfaces of the endoscope in three hours, and internal channels in one hour. This maintains endoscopes in a patient-ready condition, reducing the risk of post-reprocessing microbial growth.1

The ENDODRY Storage and Drying System integrates with the MEDIVATORS™ ADVANTAGE PLUS™ Automated Endoscope Reprocessor and transport cassette to minimize handling throughout the reprocessing workflow, enhancing infection prevention practices and reprocessing efficiency.

DESCRIPTION

• Efficiently stores up to eight endoscopes horizontally using cassettes for organized and accessible storage.

Reduced operator handling minimizes endoscope recontamination and optimizes infection control.

• Meets European standard EN 16442:2015, ensuring secure storage and infection control compliance.

• Achieves external drying in three hours and internal channel drying in one hour, improving endoscope turn-around time.

• Provides validated endoscope storage for up to 31 days, offering flexibility in endoscope use.

• Available in single-sided or pass-through versions with left or right door swing configurations.

SPECIFICATIONS

Configurations Single-sided Pass-through

Size (W x H x L) 600 x 1900 x 540 mm (23.6 x 74.8 x 21.3″)

Weight Operating Weight: 170 kg (375 lb) Standard Weight: 145 kg (320 lb) Shipping Weight: 220 kg (485 lb)

Compressed Air Requirements 6 mm (1/4″) ID

Pressure: 2–10 bar (29–145 psi) Pressure “one-click”: 5–10 bar (72.5–145 psi)

1. Perumpail, Ryan B., Neil B. Marya, Betty L. McGinty, and V. Raman Muthusamy. "Endoscope reprocessing: comparison of drying effectiveness and microbial levels with an automated drying and storage cabinet with forced filtered air and a standard storage cabinet." American journal of infection control 47, no. 9 (2019): 1083-1089.

ENDODRY™— 11059789 Rev. A 2

Demand: 100 L/min (SRA); 17 L/min at 5 bar (3.5 SCFM); 0.6 CFM at 72.5 psi Demand “one-click”: 200 L/min (SRA); 33 L/min at 5 bar (7 SCFM); 1.2 CFM at 72.5 psi2

Quality: ISO 8573-1:2010 class 1.4.1

Electricity Power Supply: 100–240 V AC ± 10%, 50/60 Hz, 1-phase Power Input: 100 W

Temperature and Humidity Requirements

Operating Temperature: 5°C–40°C (41°F–104°F) Relative Humidity: 10%–75%

Transportation and Storage

Temperature: −10°C–60°C (14°F–140°F) Humidity: 10%–90% Altitude: Up to 2000 m

Noise Level 49.8–50.3 dBA3

Interface IP20, Electrical Protection Class I, Ethernet RJ45 (10/100 Mbit)

REGULATIONS AND STANDARDS

The ENDODRY Storage and Drying System meets applicable requirements of the following listings and standards, and carries the appropriate symbols:

• Class I medical device under Regulation (EU) 2017/745, as amended

• UNI EN ISO 9001 – Quality management systems – Requirements

• EN ISO 13485 – Medical devices – Quality management systems – Requirements for regulatory purposes

• EN ISO 15223–1 – Medical devices – Symbols to be used with information to be supplied by the manufacturer – Part 1: General requirements

• EN 16442 – Controlled environment storage cabinet for processed thermolabile endoscopes

• IEC 61010–1 – Safety requirements for electrical equipment for measurement, control, and laboratory use – Part 1: General requirements

• IEC 61326–1 – Electrical equipment for measurement, control and laboratory use – EMC requirements – Part 1:

General requirements

• DIN EN 62366–1 – Medical devices – Part 1: Application of usability engineering to medical devices

• IEC 62304 – Medical device software – Software life cycle processes

FEATURES

Intuitive Touch Screen Interface: The ENDODRY Storage and Drying System features a 6″ color touch screen for quick access to essential information. Whether using the standard or pass-through version with dual touch screens, this intuitive interface displays drawer numbers, stored endoscope details, drying times and status indicators, making it user-friendly.

Enhanced Endoscope Drying: The ENDODRY Storage and Drying System uses filtered compressed air to efficiently dry endoscopes. It achieves external drying in three hours and inner channel drying in one hour.4 Each endoscope connects to the drying circuit through a hookup connection, ensuring thorough drying and reducing microbial risks.

Effective Filtration System: Endoscope drying occurs as air circulates through its channels and external surfaces. The ENDODRY Storage and Drying System is equipped with a powerful air filtration system that achieves a filtration result of over 99.995% for particles under 0.3 μm. For comparison, the EN 16442 standard describes filters with a 99.95% retention rate.

SMART Light Indicators: These LED status indicators provide at-a-glance information on endoscope storage status and potential issues. Green indicates the storage phase, blue signifies the drying phase and red alerts to alarms or the expiration of storage time. The LED colors match those displayed on the touch screen interface, providing an intuitive user experience.

Space-Saving Design: The ENDODRY Storage and Drying System, equipped with tempered glass doors, features a space-conscious design, tailored to fit the needs of any healthcare facility. Available in single-sided or pass-through versions (front and back door) for separation of wet and dry areas.

Endoscope Transport Cassettes: Specially designed endoscope transport cassettes not only safeguard stored

2. SRA: Standard Reference Atmosphere.

3. Sound pressure level (SPL) was determined according to IEC 61010-1.

4. Perumpail, Ryan B., Neil B. Marya, Betty L. McGinty, and V. Raman Muthusamy. "Endoscope reprocessing: comparison of drying effectiveness and microbial levels with an automated drying and storage cabinet with forced filtered air and a standard storage cabinet." American journal of infection control 47, no. 9 (2019): 1083-1089.

ENDODRY™— 11059789 Rev. A 3 endoscopes but also prevent free swinging of the distal tip, reducing the risk of damage, repairs and infection transmission. Using same cassettes for both the automated endoscope reprocessor and the ENDODRY Storage and Drying System minimizes contamination risks, streamlines workflow and decreases the likelihood of endoscope damage and related repairs. These transport cassettes provide an additional layer of protection, preventing recontamination and ensuring efficient endoscope handling.

Efficient Endoscope Connection: Each endoscope connects to the drying circuit through an ergonomic hookup connection. This user-friendly design simplifies the connection process for operators. Depending on the model, the endoscope cabinet shares the same connection system as MEDIVATORS™ ADVANTAGE PLUS™ Automated Endoscope Reprocessor, MEDIVATORS™ ADVANTAGE PLUS™ Pass-Thru Automated Endoscope Reprocessor and MEDIVATORS™ DSD EDGE™ Automated Endoscope Reprocessor. Using a common endoscope connection for the reprocessing phases streamlines operations and offers advantages such as improved efficiency, reduced handling, lower contamination risk and quick connections.

Climate Sensors: These sensors are used to maintain the storage environment for endoscopes within specified temperature and humidity ranges, while also ensuring controlled airflow to facilitate the drying process. They play a crucial role in creating a controlled and optimal storage environment for the endoscopes.

SAFETY FEATURES

Error Alerts: The system alerts users in case of errors or if endoscopes have exceeded their designated storage time.

Barcode Identification: To enhance traceability and reduce risk of errors, the system incorporates a barcode scanner that accurately identifies both the endoscope and the operator.

Locking Doors: The system's locking doors serve as an added layer of security, safeguarding the stored endoscopes and preventing unauthorized access or tampering.

CONTROL SYSTEM

The ENDODRY Storage and Drying System records essential information, including endoscope storage location, duration, drying time and process data. The control system offers traceability through:

• Barcode Scanner: Captures and stores data for users, endoscopes and drawers via barcode technology.

• Report Generation: All relevant process data can be printed using a network printer. An application on a desktop PC can send print requests to available printers (optional network printing). Optional label printer is available.

• Network Integration: ENDODRY Storage and Drying System interfaces with the hospital network through LAN cable (optional SMART Connect add-on).

• SMART Connect License: This option enhances communication between the endoscope reprocessor and the ENDODRY Storage and Drying System via a central server. By enabling SMART Connect between the MEDIVATORS ADVANTAGE PLUS Automated Endoscope Reprocessor, MEDIVATORS ADVANTAGE PLUS Pass-Thru Automated Endoscope Reprocessor and the ENDODRY Storage and Drying System, data transfer between the devices is streamlined for improved workflow efficiency. Additionally, this option offers the following advantages:

– Automatic endoscope/user data transfer from the automated endoscope reprocessor to the ENDODRY Storage and Drying System, ensuring up-to-date information.

– Real-time verification of proper endoscope reprocessing upon insertion.

MOUNTING ARRANGEMENT

The ENDODRY Storage and Drying System is available in single-sided or pass-through versions, offering access from both front and back doors and is designed for freestanding installation.

PREVENTIVE MAINTENANCE

A global network of skilled service specialists can provide periodic inspections and adjustments to help ensure low-cost peak performance. A STERIS representative can provide information regarding annual maintenance programs.

NOTES

Supplied ENDODRY Storage and Drying System configurations may vary by market. Consult your local sales representative for more information.

For options on connecting ENDODRY Storage and Drying System with STERIS connectivity solutions, contact your sales representative.

ENDODRY™— 11059789 Rev. A 4

Selections Checked Below Apply To This Equipment

CONFIGURATIONS FOR MEDIVATORS

ADVANTAGE PLUS AER PLATFORMS

Pass-Thru Cabinet with Operator Door Opening Left and User Door Opening Left (DRY-2001) Pass-Thru Cabinet with Operator Door Opening Right and User Door Opening Left (DRY-2002) Pass-Thru Cabinet with Operator Door Opening Left and User Door Opening Right (DRY-2003) Pass-Thru Cabinet with Operator Door Opening Right and User Door Opening Right (DRY-2004) Single-Sided Cabinet with Door Opening Left (DRY-1001) Single-Sided Cabinet with Door Opening Right (DRY-1002)

CONFIGURATIONS FOR DSD AER

PLATFORMS

Single-Sided Cabinet with Door Opening Left (DRY-1003) Single-Sided Cabinet with Door Opening Right (DRY-1004)

ACCESSORIES

ADVANTAGE PLUS Transport Cassette (CAS-1000) Silicon Oil (17599496) O-Rings for Manifold (one each) (38110) Drawer Self-Adhesive Silicone Bumpers (one each) (15974) USB Adapter Cable (one each) (15376)

Item:

Locations:

For Further Information, Contact Your Local STERIS Representative.

Cantel Medical (Italy) S.r.l.

Via Laurentina 169 00071 Pomezia (RM), Italy www.steris.com

The base language of this document is ENGLISH. Any translations must be made from the base language document.

CUSTOMER IS RESPONSIBLE FOR

COMPLIANCE WITH APPLICABLE LOCAL

AND NATIONAL CODES AND

REGULATIONS.

©2025, STERIS Corporation.

All rights reserved.

This document is intended for the exclusive use of STERIS Customers, including architects or designers. Reproduction in whole or in part by any party other than a Customer is prohibited.

www.steris.com

check_0: Off