This is a Draft Requirement Description for respiratory fit testing equipment being procured by the Department of Veterans Affairs, Network Contracting Office 20, on behalf of the Western States Network Consortium (WSNC) for VA medical facilities and clinics in VISNs 17, 19, 20, 21, and 22. The primary requirement is for brand name or equal to TSI 8048 PortaCount Respirator Fit Tester, a quantitative fit-testing device that must comply with OSHA and ANSI standards for tight-fitting respirators.
The equipment must be capable of fit-testing multiple respirator types including full-face and half-face elastomeric respirators, NIOSH series 100, 99, and 95 filtering-facepiece respirators (including disposable N95s), and European FFP1, FFP2, and FFP3 respirators. Key performance specifications include a maximum fit factor of 10,000 (200 for disposable N95s), test completion time of 2-5 minutes, and an onboard touchscreen with options for external tablet or laptop connectivity. The device must feature internal memory capable of storing a minimum of 1,000 fit-test results and include non-destructive N95 testing capability without probe insertion. Portability is essential, with maximum weight limits of 8 pounds for the unit and 16 pounds including all accessories. Additional requirements include NIOSH-certified mask sampling adapter kits and TRM-approved software or software capable of obtaining TRM approval.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C26026Q0330.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Draft Salient Characteristics
1. Contractor shall provide brand name or equal to TSI 8048 PortaCount Respirator Fit Tester
1. The quantitative respirator fit-tester shall include a fit-testing protocol with exercises and exercise procedures that shall be in compliance with OSHA and ANSI for tight-fitting respirators.
1. The respirator fit-tester shall be able to fit-test a wide variety of disposable N95s as well as other respirators. The quantitative respirator fit-tester at a minimum must be capable of fit-testing the following respirators:
1. Full-face elastomeric
1. Half-face elastomeric
1. NIOSH series-100 filtering-facepiece
1. NIOSH series-99 filtering-facepiece
1. NIOSH series-95 filtering-facepiece, including disposable N95s
1. Series FFP1, FFP2, FFP3 respirators
1. The respirator fit-tester must have a maximum fit factor of 10,000; and a maximum fit factor of 200 for disposable N95 respirators.
1. The respirator fit-tester shall complete the test between 2 – 5 minutes.
1. The respirator fit-tester must have an onboard touch screen for ease of use, and the option to connect to an external tablet or laptop.
1. The respirator fit-tester must have its own internal memory capability for the storage and retrieval of previous conducted test.
1. Equipment must be able to test N95s without having to insert a probe so it does not destroy respirators for future uses.
1. The respirator fit-tester needs to be capable of storing at minimum of 1000 fit-test results.
1. The respirator fit-tester needs to be portable for transporting it to various locations within a hospital as well as to off-site patient clinics and other locations, with a maximum weight limit of 8 pounds for the unit and a maximum weight limit of 16 pounds when all accessories.
1. Mask sampling adapter kits must be NIOSH certified.
1. Software needs to be TRM approved or capable of getting TRM approved.