May 30, 2018
National Center for Advancing Translational Sciences
(NIH-NCATS)
Program of Requirements Contract No. HHSN292201500003I
Task Order no. HHSN29200005
NIH-NCATS PROGRAM OF REQUIREMENTS
MAY 30, 2018
Page ii
This page has been left blank intentionally.
Page ii
Page iii
PROGRAM OF REQUIREMENTS (POR) FOR
THE NATIONAL INSTITUTES OF HEALTH
NATIONAL CENTER FOR ADVANCING TRANSLATIONAL SCIENCES
Division of Policy and Program Assessment
DHHS/NIH/ORF
Page iv
TABLE OF CONTENTS
I. EXECUTIVE SUMMARY 1
A. Project Description 1
B. Background Information 1
C. Objectives of the Program of Requirements 2
D. Introduction to the Final Program of Requirements 3
II. DESIGN STANDARDS 5
A. Applicable Codes and Standards: 5
B. Applicable Federal Regulations: 5
C. Utility Company Regulations 5
D. Industry Codes & Standards 6
III. SPACE AND OCCUPANCY REQUIREMENTS 7
A. Activity Analysis 7
B. General Space Parameters 9
C. Space Descriptions 9
D. Space Program List 17
Page v
IV. BUILDING CRITERIA AND DESIGN REQUIREMENTS 155
A. General Requirements 155
B. General Design and Construction Guidelines 155
C. Architectural Criteria 160
D. Sustainability 160
E. Signage 161
F. Laboratory Design Criteria 161
G. Structural 163
H. Mechanical Systems 164
I. Plumbing Systems 169
J. Life Safety/ Fire Prevention 172
K. Electrical 174
L. Communications/Telephone and Data 178
M. Security - Access Control System 179
N. Equipment List 180
V. APPENDIX 183
A. Abbreviations 184
B. Drawing Reference Symbols 186
C. Room Finish Class Group Key (Architectural Requirements) 188
Page vi
Page vi
I. EXECUTIVE SUMMARY Page 1
I. EXECUTIVE SUMMARY
A. Project Description
Th e National Center for Advancing Translational Sciences (NCATS) at the National Institutes of Health (NIH) was offi cially established in fi scal year 2012 to transform the translational science process so that new treatments and cures for disease can be delivered to patients faster. NCATS, one of 27 Institutes and Centers (ICs) at NIH, strives to develop innovations to reduce, remove or bypass costly and time-consuming bottlenecks in the translational research pipeline in an eff ort to speed up the delivery of new drugs, diagnostics, and medical devices to patients. 1
Translation is the process of turning observations in the laboratory, clinic and community into interventions that improve the health of individuals and the public — from diagnostics and therapeutics to medical procedures and behavioral changes. Translational science is the fi eld of investigation focused on understanding the scientifi c and operational principles underlying each step of the translational process. NCATS studies translation on a system-wide level as a scientifi c and operational problem. 2
NCATS currently occupies leased lab and offi ce space in Rockville, Maryland. Because of continued growth in programs and staffi ng, the NIH seeks to lease new clinical research space for NCATS to complement the existing laboratories and provide effi cient facilities to meet the needs of NCATS both today and into the future.
Th e new NCATS lease will occupy approximately 19,700 usable square feet of leased space in the northern metropolitan Washington, DC area. Th e new lease shall be located within the delineated area defi ned in the NIH Request for Lease Proposal (RLP).
B. Background Information
NCATS is comprised of six components; the Bioprinting Group, Soft ware Development, Small Molecule Repository,Incubator Lab, Freezer Repository and IC Storage Facility, and Intramural Administrative Services/ Acquisitions. Several of these components have outgrown their original space and future projections indicate a continuing growth in staff and program activities.
1 NCATS history from website www.ncats.NIH.gov/about/center 2 Translational Science explanation from website www.ncats.NIH.gov/about/center
I. EXECUTIVE SUMMARY Page 2
Because of space constraints the Intramural Administrative Services and Acquisitions group has been physically separated to a second leased facility. Th is separation has resulted in reduced effi ciencies and separated a vital support group of NCATS from their customers.
Because of space constraints at the main NCATS facility, additional space is also being leased at two remote locations for freezer and refrigeration storage. Th e remote freezer and refrigeration locations also have a negative impact on the effi ciency of the NCATS operations.
To meet current and future needs NCATS program components require more space to eff ectively fulfi ll their objectives and meet the increase in scientifi c demand. A new leased facility is needed to supplement the existing clinical research facility, allow for the relocation of the Intramural Administrative Services and Acquisitions group, and provide freezer storage capacity.
Th is Program of Requirements (POR) has been developed to defi ne the space needs and technical requirements for a new NCATS lease.
C. Objectives of the Program of Requirements
Th e ultimate objective of this POR is to defi ne the requirements for a lease that will provide an innovative and fl exible work environment that can evolve over the next 15 years to meet tomorrow’s NIH and NCATS workforce requirements by increasing workfl ow effi ciencies, enhancing collaboration and increasing morale.
Th e activities and requirements defi ned in this POR are for three of the groups within NCATS: Compound Management, Tissue Printing and Intramural Administrative Services and Acquisitions. Th ese three groups will be occupying the new lease.
NCATS is comprised of a diverse staff who guide the Center’s translational research activities. A total of up to 49 staff are expected to utilize the new NCATS lease as described in this POR. A breakdown of staffi ng is included in Section III of this POR.
Th e POR has been developed collaboratively with the NIH and NCATS to compile and include all necessary technical requirements including chemical lists and other features for BSL-2 Laboratories and needed offi ce space. Th e POR incorporates requirements and guidelines from the NIH Design Requirements Manual (DRM) along with other applicable codes, standards, and requirements.
Th e POR is intended to be used for the purpose of soliciting lease proposals, and as the basis of design for the fi nal design and construction of the leased space.
I. EXECUTIVE SUMMARY Page 3
D. Introduction to the Final Program of Requirements
Th e following Program of Requirements (POR) provides documentation of the programing process to identify the space needs and technical requirements of the new NCATS lease space for the National Institutes of Health (NIH). Th is POR refl ects information and criteria gathered from the NIH, additional data collection from equipment suppliers, site visits to the existing facility, and meetings and interviews with the NCATS personnel aff ected.
Th is POR was reviewed and updated in collaboration with the NIH and NCATS through an interactive process of input and reviews as the the fi nal document was developed.
A 100% cost estimate was developed for the project and forwarded to the NIH under a separate transmittal.
I. EXECUTIVE SUMMARY Page 4 Page 4
II. DESIGN STANDARDS Page 5
II. DESIGN STANDARDS
A. Applicable Codes and Standards:
Th e fi nal design and contract documents for the NCATS lease shall comply with all applicable Federal Regulations, the National Institutes of Health (NIH) Design Requirements Manual (DRM), the NIH leasing agreement, and all applicable Building Codes and Standards. In case of confl icts between adopted or selected codes or standards and the DRM, the most stringent, technically appropriate, and conservative criteria shall apply unless directed otherwise by the NIH Project Offi cer. Where it is unclear which set of requirements is applicable, consult the Authority Having Jurisdiction (AHJ) and the NIH Project Offi cer for direction.
Once a building location is selected and a lease is executed, a detailed code analysis shall be performed be the lessor’s architect and engineer.
Th e most current edition of each code, regulation and standard shall be applicable to the design, contract documents and construction of the new leased space. Th e most current edition should be interpreted as: (a) For County/City Code compliance purposes for permitting, design must comply with the jurisdiction’s most recently adopted edition of each. (b) For NIH leased facilities, design must comply with the most current published edition of Codes and applicable Federal Regulations, as of the time of design, even if not yet locally adopted
B. Applicable Federal Regulations:
Th e preparation of the design and contract documents for the NCATS leased space shall be in accordance with all applicable Federal and agency regulations and standards, including but not limited to: Volume I and Volume II of the Health and Human Services (HHS) Facilities Program Manual, the NIH Design Requirements Manual (DRM), the NIH Architectural and Engineering Design Policies and Guidelines, and the NIH Planning and Programmatic Guidelines.
C. Utility Company Regulations
Th e following industry standards are used as references: American Society of Heating, Refrigeration and Air Conditioning Engineers (ASHRAE), various sections of American National Standards Institute (ANSI), American Society of Mechanical Engineers (ASME), National Electrical Manufacturer Association (NEMA), National Electric Code (NEC), International Plumbing Code (IPC), and local utility requirements from Potomac Electric Power Company (PEPCO), Washington Suburban Sanitary Commission (WSSC), and Washington Gas (WG). Also, utilities and utility work are subject to Federal, State, and local government regulations in the State of Maryland.
II. DESIGN STANDARDS Page 6
D. Industry Codes & Standards
Th ere are numerous industry guidelines and standards that must be followed in concert with the codes and standards listed in this POR and the NIH DRM. Th e designer of the fi nal NCATS lease space must be cognizant of all requirements of the DRM and its referenced documents, and all requirements for Bio-Safety Level 2 (BSL-2) laboraotries. Several other codes, industry guidelines and standards are referenced in other sections of this POR.
III. SPACE AND OCCUPANCY REQUIREMENTS Page 7
III. SPACE AND OCCUPANCY REQUIREMENTS
A. Activity Analysis
Th e National Center for Advancing Translational Sciences (NCATS) provides leadership for a national program to advance the discipline of translational science and to catalyze the generation of innovative methods and technologies that will enhance the development, testing, and implementation of diagnostics and therapeutics across a wide range of human diseases and conditions. NCATS conducts intramural research and supports and coordinates research projects through research grants, contracts, and other mechanisms as well as supports training in translational sciences. In addition, NCATS collaborates with organizations and other institutions engaged in translational research and training activities, and collects and disseminates research fi ndings and related information.
Th e activities and spaces described in this POR for the NCATS lease will involve three specifi c groups: Compound Management, Automated Chemistry/IT (Incubator/ Bio-printing) and Intramural Administrative Services and Acquisitions.
Compound Management oversees the Molecular Libraries Program which off ers public sector biomedical researchers access to large scale screening capacity necessary to iden-tify small molecules that can be optimized as chemical probes. Th ese probes are then used to study the functions of genes, cells and biochemical pathways which leads to new discoveries in the functions of genes and signaling pathways. A key activity is collecting samples for high throughput biological screening and distributing those to respective collaborators. Th e new lease will allow NIH to bring this work in-house under NCATS.
Automated Chemistry/IT (Incubator/Bio-printing) responds to new research initiatives resulting from national and international health crises such as the British Petroleum spill, the Ebola epidemic, and recent Zika outbreaks. In all of these cases, the pilot phase, or entire body of work, was initiated quickly using existing space and without external funding. Public health emergencies will continue and this group will continue the rapid initiation of new research eff orts in response to these outbreaks.
Intramural Administrative Services/Acquisitions support the activities of the NCATS Intramural Program by providing comprehensive administrative and management support for NCATS. Th e support services provided include managing budget, personnel, procurement, space and facilities, travel, property, and other administrative functions as necessary to ensure the effi cient and eff ective implementation and operation of the programs within NCATS. Th is group also develops, analyzes and implements policies, guidelines, and procedures on matters relating to administrative management and disseminates to relevant staff ; and advises the staff of NCATS of administrative policies and practices.
Activity Analysis
III. SPACE AND OCCUPANCY REQUIREMENTS Page 8
Figure III-1 NCATS Laboratory Primary and Secondary Activities Diagram
Figure III-2 NCATS Flow Diagram
Activity Analysis
SHARED
SUPPORT
(COLLABORATION)
BUILDING
LOADING
DOCK
CHEMICAL
STORAGEWASTE
OFFICE ZONE
LABORATORY ZONE
COMPOUND
MANAGEMENT
& TISSUE
PRINTING
NCATS
FRONT DOOR
COMPOUND MANAGEMENT
LABORATORY
TISSUE PRINTING
LABORATORY
SHARED LABORATORY
SUPPORT
INTRAMURAL
SERVICES
OFFICES
BBBUU
LO
CCH
S
BBBORAORAORATORTORTORY ZY ZY ZONEONEONE
LA
PRINTING
NCATSTSNCACATTTTS
BBBLLLALAA
LOL
COMOMPOG
ST
AF
F
CL
EA
N
DI
RT
Y
ST
AF
F
V
IS
IT
O
R
LABORATORY ZONE
LABORATORY SERVICE CORRIDOR
OPEN LABORATORY SPACE
LAB
SUPPORT
SPACE
LAB
SUPPORT
SPACE
LAB
SUPPORT
SPACE
LAB
SUPPORT
SPACE
LINEAR
EQUIP’T
ROOM
LINEAR
EQUIP’T
ROOM
RR
ROOROOOOOOOOOOOOMMMME RR SMMM
NEAR
ROOOOMM
LAB LAB
RR SMM
III. SPACE AND OCCUPANCY REQUIREMENTS Page 9
B. General Space Parameters
All spaces described in this POR are to be in a single facility. All spaces shall be contiguous and contained within their own secured perimeter, and within a secured perimeter for the NCATS leased space. Refer to the Space Description-Guideplates later in this section for specifi c technical requirements of the individual spaces listed below and the Appendix (Figure V-1) for Master Equipment List.
C. Space Descriptions
Th e following information describes the various space types within the NCATS lease space.
Th e space types are divided into four main categories; Laboratory, Laboratory Support, Offi ce/Offi ce Support, and Building Support.
1. Laboratory Th is is the main laboratory space that will function primarily for molecular biology.
Th is area shall be a large, open space, free of columns. Th e lab will be furnished with an adaptable casework system - similar to “Solution” by Air Master Systems, “Enterprise” by Kewaunee Scientifi c, or “EVO2” by OC River - to allow for confi guration of larger, fl oor mounted robotic equipment. Th e casework will consist of mobile benches with upper shelves, mobile or suspended cabinets below the bench top. Laboratory grade metal casework meeting the Scientifi c Equipment and Furniture Association (SEFA) standards is required. Sinks, fume hoods and biosafety cabinets requiring fi xed connections to building systems shall be located strategically at the perimeter of the space to allow the lab to function well yet allow for future adaptability.
All materials used in the lab (cabinetry, shelving, fl ooring and etc.) shall be corrosion resistant to facilitate cleaning and combat the growth of bacteria. Non-absorbing materials such as stainless steel sinks, epoxy counter tops, and acid resistant plastic laminates are required. Services provided in the space shall include: potable water (for emergency shower/eyewashes) and industrial cold and hot water, laboratory waste and vent, deionized or distilled water, carbon dioxide (CO2), vacuum, compressed air, telephone, local area network for computers and electric power. Th ese services shall be carefully organized in a repetitive, modular discipline, on overhead racks, fed from above. Points of connection to each building system shall be located at consistent, predetermined, and recurring positions to allow the laboratory to be reconfi gured easily to accept new equipment.
III. SPACE AND OCCUPANCY REQUIREMENTS Page 10
2. Laboratory Support Th e support space shall be designed using the same design parameters as are used in the laboratory, i.e. the same planning module and same fi nishes. Th e following are descriptions of the various laboratory support spaces specifi c to NCATS.
Brooks Compound Storage (Guideplate I.B.1) Th is space will house the Brooks Life Science Systems Sample Store II. Th is will be housed in the open laboratory. Refer to the “Space Description – Guideplates" for additional information. NCATS is considering an automated assay ready plate handler to be added to the front of the Brooks equipment. Space shall be provided for this additional unit. Power that feeds the equipment and all mechanical components shall be on emergency power and UPS as the Brooks System cannot lose power during an outage. Th e weight of this system is substantial and will require a location that is slab-on-grade of suffi cient slab depth to support the weight (refer to the fl oor loading data in the Appendix under Brooks Life Science Systems). If the selected building requires the lab to be on an upper fl oor, the fl oor must be reinforced to support the load and access to a freight elevator of minimum size of 10'-0" x 14'-0" is required.
Kardex Compound Storage (Guideplate I.C.2) Th is space will house the Kardex Remstar Shuttle XP. Th is space shall be immediately adjacent to the open laboratory space. Power that feeds the equipment and all mechanical components shall be on emergency power and UPS as the system cannot lose power during an outage. Th e Kardex Remstar Shuttle XP shall have redundant cooling in the event of a primary cooling system failure.
-20 Degree Freezer Room on Emergency Power (Guideplate I.C.3) Th is is a controlled environmental room with a set-point of -20 degrees Celsius.
Th is room is used for storing samples only and will not be used for any temperature controlled work. Th e fl oor of the room is insulated and is several inches thick. Th e room will require either a depressed slab or a ramp to access the room. Th e depth of the depression or size of the ramp shall be coordinated with the fi nal unit selected. Th e plan layout will maximize the amount of wire shelving units to be stored inside. Th e room shall be lockable but the door shall have the largest possible view window. All mechanical components shall be accessible and serviceable from outside the room in an adjacent space. Th e refrigerant lines shall not to exceed 100 linear feet between the mechanical equipment and the room. A high and low temperature monitoring and alarm system shall be connected to a central equipment alarm system. Provide conduit pathways within the building for this room to be connected to the Rees equipment monitoring system (a NIH system). All gaps between the room and adjacent construction shall be sealed air tight. A manual door release shall be provided inside the environmental room. Th is room shall be on emergency power backup.
III. SPACE AND OCCUPANCY REQUIREMENTS Page 11
Equipment Room (Guideplate I.C.4) Th e equipment room will house ultra-low temperature (-80 °C) freezers that are currently stored off -site. Th is space will have high heat loads and high noise levels so provisions shall be made to keep the space cool to extend freezer life. Utilities for this space include HVAC with supplemental cooling, electric in both 120 and 208 volt, on emergency power and connected to an equipment monitoring system.
Hood Room (Guideplate I.C.5) Th is room is needed for working with hazardous reagents and measuring out powder compounds. Th e room consists of four fume hoods and one vented Electrostatic Dissipative (ESD) workstation. In addition, there are mobile benches with shelving above and casework below. A sink, emergency shower, and eyewash shall be provided.
Flammable/ Chemical storage cabinets shall be provided. Refer to the Appendix for a list of chemicals and quanities to be stored in this room.
4 Degree Cold Room on Emergency Power (Guideplate II.C.1) Th is is a controlled environmental room with a set-point of 4 degrees Celsius. Th is room is used for storing samples and reagents as well as long-term experiments requiring temperature controlled work and will include a stainless steel bench on one side and wire shelving units on the opposite wall. Th e fl oor of the room is insulated and is several inches thick. Th e room will require either a depressed slab or a ramp to access the room.
Th e depth of the depression or size of the ramp shall be coordinated with the fi nal unit selected. Th e room shall be lockable. All mechanical components shall be accessible and serviceable from outside of the room in an adjacent space. Th e refrigerant lines shall not to exceed 100 linear feet between the mechanical equipment and the room.
A high and low temperature monitoring and alarm system shall be connected to a central equipment alarm system. Provide conduit pathways within the building for this room to be connected to the Rees equipment monitoring system (a NIH system). All gaps between the room and adjacent construction shall be sealed air tight. A manual door release shall be provided inside the environmental room. Th is room shall be on emergency power backup.
Tissue Culture Room (Guideplate II.C.3) Th e tissue culture room is a shared space that includes two Bio-Safety Cabinets (BSCs), multiple stacked CO2 incubators with central CO2 access, an adjacent sink, deep shelf storage space for plastic ware storage, two refrigerators and a tabletop centrifuge. Th ere is working bench space near the sink to prepare media. Th e bench space between the BSCs will hold microscopes for viewing slides. A house vacuum is required to serve this space.
Histology Room (Guideplate II.C.4) Th e histology room is a shared space that will be similar to the tissue culture space but is used for the preparation of tissues for study. Th ere is working bench space throughout the room to provide directional fl ow for the processing of tissue specimens. A house vacuum is required to serve this space.
III. SPACE AND OCCUPANCY REQUIREMENTS Page 12
Tissue Printing Room (Guideplate II.C.5) Th e tissue printing room is similar in design to the tissue culture room in that it is a shared space. Th e room includes eight BSCs; four to house tissue printers and four BSC for general culture work. Similar to the tissue culture room, this space will house multiple stacked CO2 incubators, a sink, deep shelf storage, 2-3 refrigerators, a tabletop centrifuge, and a microscope bench. A house vacuum is required to serve this space.
Consumables Storage (Guideplate IV.C.1) Th is area will provide space for high density wire shelving to store kits for the various robotic instruments, gloves, pipettes and other consumable supplies and equipment.
Hazardous Waste Room (Guideplate IV.C.2) Th is room is a collection point for all chemical waste that will leave the facility. Th e room shall be located close to the building’s loading dock, yet still within the NCATS controlled lease space. Th e room shall have card key access, proper HVAC, spill containment, and an emergency shower/ eye wash. If NCATS is located on a fl oor above the ground fl oor then access to a service elevator of minimum 10'-0" x 14'-0" size is required.
Chemical Storage Room (Guideplate IV.C.3) Th is room is where all incoming fl ammable chemicals will be stored before being distributed to the laboratory spaces. Refer to the Appendix for a list of chemicals and quanities expected to be stored.
LN2 and CO2 Room (Guideplate IV.C.4) Th is room houses the in-use cylinders and dewars used as the source for nitrogen and CO2. Space for empty cylinders and spares shall also be provided.
Glasswash / Autoclave Room (Guideplate IV.C.5) Th is space serves as a central glass wash facility and will house two glassware washers and two medium-sized sterilizers. Th e intent for duplicating units is to provide redundancy so the facility can still operate if one of each type unit goes down for servicing. All joints between adjacent materials and fi nishes in this space shall to be thoroughly caulked and sealed. Th e space shall have a fi xed (non-accessible) ceiling of gypsum wallboard with an epoxy, water resistant fi nish. Th e space shall also have non-slip epoxy fl oors, and have cleanable walls to withstand moisture and prevent pest infestation. Discharge from Autoclave chamber and from Sterilizer chamber must fi rst be cooled to 140 degrees F or below before being allowed to enter fl oor drain thru open-site drain. Provide funnel on fl oor drain grate to allow drain to receive spillage. Drain piping is lab waste piping to Acid Neutralizing Tank. Acid neutralization tank requires space, chemicals storage, controls, monitoring, discharge to sanitary, and ventilation. Space is provided for staging clean and dirty glassware.
III. SPACE AND OCCUPANCY REQUIREMENTS Page 13
3. Offi ce and Offi ce Support NCATS is not in need of a formal reception area but the lease space shall have a secured “front door” where visitors, guests, and vendors can be met and escorted accordingly.
Th is space can be combined with the Transmitting (Shipping/Receiving) Station noted below. NCATS shall have one offi ce suite for all lab and non-lab users to facilitate sharing of resources. Th e offi ce suite shall be adjacent to the laboratory spaces. If the selected leased space does not allow all offi ces to be located together then the offi ces for laboratory staff shall be adjacent to their labs. Daylighting of interior occupied work spaces is required. Open workstations shall be located near the exterior windows to maximize daylight and views for all staff . Private offi ces should have glazing at doors and walls to allow daylight into offi ce space yet still provide some acoustic privacy.
Private Offi ce Private offi ces shall be provided for the NCATS staff , guest researchers, and other visitors to the leased space as indicated in the Space Program. Offi ces are required for group or team leaders, area supervisors and specifi c administrative staff . Provisions shall be made for privacy at offi ces as required by the NIH-DRM, including full height partitions, doors and baffl ing of mechanical systems that penetrate the space. Ergonomic systems furniture shall be used in all offi ce spaces.
Workstation Open offi ce space is provided for Research Associates, purchasing agents, program assistants and other administrative personnel. Th ese personnel will be located in open offi ce style workstations with 54 inches partitioning, built in work surfaces, and built in electrical, data, and communications distribution with outlets. Appropriate provisions shall be made for privacy by using sound absorbing fi nishes on all workstation partitions, and utilization of a sound absorbing ceiling system. Ergonomic systems furniture shall be used in all open offi ce spaces.
Mail / Sorting Station (Shared with Copy Room Guideplate IV.A.3) Th is is a location where incoming mail and outgoing packages can be sorted and processed. Th is station could be combined with the Transmitting Station below. Th e space should include a counter with mail cubbies for incoming mail, space for outgoing mail and space for a mobile mail cart. Th e NIH has specifi c requirements , rules and regulations governing mail. Th e mail sorting and distribution equipment must be compliant with all NIH rules and regulations. Requirements for NIH mail distribution equipment can be found in Section 4.6.1.7 of the NIH DRM.
III. SPACE AND OCCUPANCY REQUIREMENTS Page 14
Conference Room (Guideplate IV.A.1) A conference room is needed to house up to 30 people for formal and informal meetings of staff and for periodic training. Th e conference area shall be utilized on a shared basis and be designed in accordance with National Fire Protection Association (NFPA) occupant loads. Th e conference room shall be equipped to accommodate fl exible seating arrangements. Provide white boards, electrical connections for audio visual equipment, a audio visual screen, adjustable overhead lighting, and data and telephone lines. Th e minimum ceiling height shall be 9'-0" with lay-in acoustical ceiling tiles.
Transmitting Station (Shipping/Receiving - Guideplate IV.A.2) Th is is a multi-purpose station that processes all incoming and outgoing packages. Th is station should be located near the “front door” of the facility and can include reception duties as well.
Copy / Supply Room (Shared with Mail/Sorting Station Guideplate IV.A.3) Space shall be provided for copying machines, FAX machines, fi les, shelves and other routine offi ce equipment. In addition, space is required for central computer systems.
Th is area does not have the stringent air change requirements that the laboratory areas have. File rooms shall be lockable.
III. SPACE AND OCCUPANCY REQUIREMENTS Page 15
Space Descriptions - Guideplates
4. Building Support While the selected building will have spaces that support NCATS and other tenants, the following areas are intended to be exclusive spaces for the function of NCATS within the leased space.
Break Room (Guideplate IV.B.1) A break room shall be provided for NCATS staff to take coff ee breaks and eat lunches.
Th ere shall be a plastic laminate fi nished counter top with built-in stainless steel sink, and overhead and below counter cabinets for storage. Th ere shall be space provided for a residential size refrigerator with ice maker (the refrigerator to be provided by others).
Th ere shall also be space provided for two microwave ovens (provided by others), and under-counter waste bins and recycling receptacles (provided by others). Seating and tables for up to 12 NCATS staff shall also be provided (seating and tables to be provided by others).
Mechanical Room HVAC and plumbing systems, with their respective redundancies, and all utilities are required to be provided as base building systems located in the main building mechanical equipment room(s). Th e main building mechanical equipment room(s) shall be located outside of and in addition to the proposed NCATS lease space of 19,7000 SF, but is a requirement as part of the overall leased facility. Primary HVAC and plumbing system requirements and required redundancies are further described in the Mechanical Systems summary in Section IV.
Compressed air and laboratory vacuum with redundancy are two systems required to be provided as base building systems through all NCATS lab and lab support spaces (the nitrogen and carbon dioxide systems are listed under Laboratory Support). If the required compressed air and laboratory vacuum systems in the base building are not adequate to handle the needs of NCATS lease, mechanical space is required to be provided in the base building mechanical room(s) located outside of the NCATS lease space to house these supplemental systems, so they are not located in laboratory space.
Similarly, a RO/DI water system is required in some NCATS labs and lab support spaces.
Th e RO/DI water system equipment shall be located in the base building mechanical room(s), outside of the NCATS leased space.
III. SPACE AND OCCUPANCY REQUIREMENTS Page 16
Space Descriptions - Guideplates
Electrical/UPS Room Th e NCATS lease space has a large number of robotic equipment, both large and small, that cannot experience loss of power in the event of a power outage. Th e building containing the NCATS lease space may have UPS power but typically these systems are specifi c to the needs of the tenants and may not be adequate for NCATS.
Th e electrical NCATS UPS room is meant to house a UPS system dedicated to NCATS. In addition any electrical sub-panels or transformers dedicated to NCATS lease space will be located here.
MDF/IDF
NCATS will be connected to the NIH IT network using a cloud-based strategy. Th is room will house any cabling and equipment needed to connect to the cloud as well as equipment and cabling needed for a wireless network. Th is shall be a secured space for NCATS access only.
III. SPACE AND OCCUPANCY REQUIREMENTS Page 17
Space Descriptions - Guideplates
NCATS Space Program List Continued
D. Space Program List
III. SPACE AND OCCUPANCY REQUIREMENTS Page 18
Space Descriptions - Guideplates
III. SPACE AND OCCUPANCY REQUIREMENTS Page 19
III. SPACE AND OCCUPANCY REQUIREMENTS Page 20
III. SPACE AND OCCUPANCY REQUIREMENTS Page 21
III. SPACE AND OCCUPANCY REQUIREMENTS Page 22
*FOR INFORMATION ONLY
OFFICE - GROUP
LEAD
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
OFFICE - GROUP
LEAD
OFFICE - HOTELING
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
COPY/ SUPPLY/ MAIL
PHONE/ HUDDLE ROOM
KITCHENETTE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
*FOR INFORMATION ONLY
GUIDEPLATE SKETCH
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
1" = 10'-0"
COMPOUND MANAGEMENT - CONCEPTUAL ORGANIZATION AND
ADJACENCY PLAN FOR OFFICES
I.A.0
10217_NIH NCATS R17 Guide Plates
10217.000
0 10' 20'
OFFICE SUITE ENTRANCE
COMPOUND MANAGEMENT OFFICES
ADMIN SERVICES CORRIDOREXIT TO LABORATORY ZONE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
OFFICE - GROUP
LEAD
III. SPACE AND OCCUPANCY REQUIREMENTS Page 23
Space Descriptions - Guideplates
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
OFFICE - GROUP LEAD - 3D VIEW
I.A.1-3D
10217_NIH NCATS R17 Guide Plates
10217.000
III. SPACE AND OCCUPANCY REQUIREMENTS Page 24
Space Descriptions - Guideplates
LATERAL FILE STORAGE
WHITE BOARD
WORKSTATION
CREDENZA
TACK BOARD
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
1/4" = 1'-0"
OFFICE - GROUP LEAD
I.A.1
10217_NIH NCATS R17 Guide Plates
0 4' 8'
III. SPACE AND OCCUPANCY REQUIREMENTS Page 25
III. SPACE AND OCCUPANCY REQUIREMENTS Page 26
III. SPACE AND OCCUPANCY REQUIREMENTS Page 27
Space Descriptions - Guideplates
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
WORKSTATION - RESEARCH ASSOCIATE - 3D VIEW
I.A.2-3D
III. SPACE AND OCCUPANCY REQUIREMENTS Page 28
Space Descriptions - Guideplates
PD
LATERAL FILE STORAGE BELOW
WORKSTATION
WORKSTATION PANEL
COLLABORATION SPACE
BEYOND
WHERE INDICATED IN AN EXISTING FLOOR, PROVIDE CORE DRILL PACKAGE TO MAINTAIN FIRE
RATING OF FLOOR. EXTEND CONDUIT AND WIRING IN CEILING SPACE OF FLOOR BELOW TO
ELECTRICAL CLOSET OR PANELBOARD.
COMBINATION POWER AND TELECOMMUNICATIONS FLOORPD
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
1/4" = 1'-0"
WORKSTATION - RESEARCH ASSOCIATE
I.A.2
10217_NIH NCATS R17 Guide Plates
III. SPACE AND OCCUPANCY REQUIREMENTS Page 29
III. SPACE AND OCCUPANCY REQUIREMENTS Page 30
III. SPACE AND OCCUPANCY REQUIREMENTS Page 31
Space Descriptions - Guideplates
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
OPEN LABORATORY - 3D VIEW
I.B.1-3D
III. SPACE AND OCCUPANCY REQUIREMENTS Page 32
Space Descriptions - Guideplates
OVERHEAD SERVICE
CARRIER (TYP) - (2)
CAPPED EXHAUST, UPS
POWER, (3) CA/SIDE,
QUICK CONNECT FITTINGS
ENTRANCE TO
COMPOUND
STORAGE
(KARDEX)
ACOUSTIC REFORMATTER STATION
FUTURE WORKFLOW AUTOMATION (TECAN) -
FRONT OF UNIT
COMPOUND STORAGE (BROOKS)
(4) N2
(4) N2
(3) LCMS
OPEN
EQUIPMENT
ZONE
OPEN
EQUIPMENT
ZONE (TYP)
DRY ICE
STORAGE
MOVABLE TABLE
AND CART ZONE
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
1/16" = 1'-0"
OPEN LABORATORY
I.B.1
10217_NIH NCATS R17 Guide Plates
0 16' 32'
III. SPACE AND OCCUPANCY REQUIREMENTS Page 33
III. SPACE AND OCCUPANCY REQUIREMENTS Page 34
III. SPACE AND OCCUPANCY REQUIREMENTS Page 35
Space Descriptions - Guideplates
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
COMPOUND STORAGE (KARDEX) - 3D VIEW
I.C.2-3D
III. SPACE AND OCCUPANCY REQUIREMENTS Page 36
Space Descriptions - Guideplates
KARDEX REMSTAR
SHUTTLE XP - HD 500
NOTES:
- KARDEX UNIT REQUIRES MINIMUM CLEAR SPACE - CONFIRM
WITH FINAL SELECTION BY GOVERNMENT
- DUE TO HIGH DENSITY STORAGE, STRUCTURAL
ACCOMMODATION OF UNIT WEIGHT, UP TO 27,088
(HORIZONTAL ORIENTATION) TOTAL POUNDS IS REQUIRED
- LOCATED IN CLOSE PROXIMITY TO THE FH ROOM AND
IDEALLY ACCESSED FROM THE LINEAR EQUIPMENT ROOM OFF
OF THE OPEN LAB
FLOOR SINK - UNIT
CONDENSATE
FOOTPRINT OF
LINEAR EQUIPMENT
ROOM
OPEN LAB
BROOKS
FUME
HOOD
ROOM KARDEX
ENTRANCES
LER
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
As indicated
COMPOUND STORAGE (KARDEX)
I.C.2
III. SPACE AND OCCUPANCY REQUIREMENTS Page 37
III. SPACE AND OCCUPANCY REQUIREMENTS Page 38
III. SPACE AND OCCUPANCY REQUIREMENTS Page 39
Space Descriptions - Guideplates
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
-20 DEGREE FREEZER ROOM - 3D VIEW
I.C.3-3D
III. SPACE AND OCCUPANCY REQUIREMENTS Page 40
Space Descriptions - Guideplates
WIRE SHELVING UNITS (GFGI)
NOTES:
- COMPRESSOR/CONDENSER SHALL BE LOCATED IN CLOSET OR
MECHANICAL SPACE WITHIN 100 LF OF ROOM FOR WATER COOLED
UNIT
- AIR COOLED UNIT CAN BE PLACED ON THE ROOF WITHIN 100 LF OF
ROOM
- ROOM AND EQUIPMENT SHALL BE ON BACK-UP POWER AND
CONNECTED TO REES EQUIPMENT MONITORING SYSTEM
- LONG TERM STORAGE
- COULD BE ACCESSED FROM WITHIN THE 4-DEGREE FREEZER ROOM
TO REDUCE TEMPERATURE SWINGS
COLD ROOM EVAPORATOR
4 DEGREE FREEZER ANTE-
ROOM (SHOWN GREY FOR
REFERENCE
COMPRESSOR /
CONDENSER
UNIT
FLOOR SINK FOR CONDENSATE
THE DOOR BETWEEN THE 4
DEGREE AND -20 TO HAVE A
LARGE VIEW WINDOW WITH
DEFROST HEATING
ELEMENT
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
1/8" = 1'-0"
-20 DEGREE FREEZER ROOM
I.C.3
10217_NIH NCATS R17 Guide Plates
0 8' 16'
III. SPACE AND OCCUPANCY REQUIREMENTS Page 41
III. SPACE AND OCCUPANCY REQUIREMENTS Page 42
III. SPACE AND OCCUPANCY REQUIREMENTS Page 43
Space Descriptions - Guideplates
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
EQUIPMENT ROOM (-80FREEZERS) - 3D VIEW
I.C.4-3D
III. SPACE AND OCCUPANCY REQUIREMENTS Page 44
Space Descriptions - Guideplates
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80
-80 -80
FREEZERS (GFGI)
ELECTRICAL
RACEWAY ON
BOTH WALLS
AND ON
OHSC
OVERHEAD SERVICE CARRIER WITH
120V DEDICATED AND 208V SINGLE
PHASE RECEPTACLES
-80
NOTES:
- ROOM TO STORE APPROXIMATELY (18) -80
FREEZERS THAT ARE CURRENTLY STORED
OFF-SITE
- DEPENDING ON SELECTED BUILDING, THIS
ROOM COULD BE RECONFIGURED INTO A
LINEAR LAYOUT
- ALL POWER FOR FREEZERS TO BE ON UPS
AND REES EQUIPMENT MONITORING
FLOOR DRAIN
FOR DEFROSTING
FREEZERS
FUTURE
FREEZERS
(GFGI)
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
1/8" = 1'-0"
EQUIPMENT ROOM (-80 FREEZERS)
I.C.4
10217_NIH NCATS R17 Guide Plates
III. SPACE AND OCCUPANCY REQUIREMENTS Page 45
III. SPACE AND OCCUPANCY REQUIREMENTS Page 46
III. SPACE AND OCCUPANCY REQUIREMENTS Page 47
Space Descriptions - Guideplates
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
HOOD ROOM - 3D VIEW
I.C.5-3D
III. SPACE AND OCCUPANCY REQUIREMENTS Page 48
Space Descriptions - Guideplates
ELECTRICAL
RACEWAY - TYP AT
FIXED BENCH
OPEN STORAGE
SINK,
CW, HW, EW
(1) FUME HOOD - ANALITICAL
BALANCE, N2, CA, POWER (TYP)
(3) VENTED WORKSTATION
WITH ESD WORKTOP -
CONNECT TO BUILDING
EXHAUST
CART STORAGE
CONSUMABLE STORAGE
CONSUMABLE STORAGE
EWES
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
1/8" = 1'-0"
HOOD ROOM
I.C.5
10217_NIH NCATS R17 Guide Plates
III. SPACE AND OCCUPANCY REQUIREMENTS Page 49
III. SPACE AND OCCUPANCY REQUIREMENTS Page 50
OFFICE - GROUP
LEAD
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
OFFICE - GROUP
LEAD
OFFICE - HOTELING
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
COPY/ SUPPLY/ MAIL
PHONE/ HUDDLE ROOM
KITCHENETTE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
*FOR INFORMATION ONLY
GUIDEPLATE SKETCH
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
1" = 10'-0"
TISSUE PRINTING - CONCEPTUAL ORGANIZATION AND
ADJACENCY PLAN FOR OFFICES
II.A.0
10217_NIH NCATS R17 Guide Plates
10217.000
0 10' 20'
OFFICE SUITE ENTRANCE
TISSUE PRINTING OFFICES
ADMIN SERVICES CORRIDOREXIT TO LABORATORY ZONE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
WORKSTATION -
RESEARCH
ASSOCIATE
OFFICE - GROUP
LEAD
OFFICE - HOTELING
*FOR INFORMATION ONLY
III. SPACE AND OCCUPANCY REQUIREMENTS Page 51
Space Descriptions - Guideplates
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
OFFICE - GROUP LEAD - 3D VIEW
II.A.1-3D
III. SPACE AND OCCUPANCY REQUIREMENTS Page 52
Space Descriptions - Guideplates
LATERAL FILE STORAGE
WHITE BOARD
WORKSTATION
CREDENZA
TACK BOARD
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
1/4" = 1'-0"
OFFICE - GROUP LEAD
II.A.1
10217_NIH NCATS R17 Guide Plates
III. SPACE AND OCCUPANCY REQUIREMENTS Page 53
III. SPACE AND OCCUPANCY REQUIREMENTS Page 54
III. SPACE AND OCCUPANCY REQUIREMENTS Page 55
Space Descriptions - Guideplates
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
OFFICE - HOTELING - 3D VIEW
II.A.2-3D
III. SPACE AND OCCUPANCY REQUIREMENTS Page 56
Space Descriptions - Guideplates
WORKSTATION
WHITE BOARD
WORKSTATION
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
1/4" = 1'-0"
OFFICE - HOTELING
II.A.2
10217_NIH NCATS R17 Guide Plates
III. SPACE AND OCCUPANCY REQUIREMENTS Page 57
III. SPACE AND OCCUPANCY REQUIREMENTS Page 58
III. SPACE AND OCCUPANCY REQUIREMENTS Page 59
Space Descriptions - Guideplates
OPEN TO WORKSTATIONS
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
WORKSTATION - RESEARCH ASSOCIATE - 3D VIEW
II.A.3-3D
III. SPACE AND OCCUPANCY REQUIREMENTS Page 60
Space Descriptions - Guideplates
PD
LATERAL FILE STORAGE BELOW
WORKSTATION
WORKSTATION PANEL
COLLABORATION SPACE
BEYOND
WHERE INDICATED IN AN EXISTING FLOOR, PROVIDE CORE DRILL PACKAGE TO MAINTAIN FIRE
RATING OF FLOOR. EXTEND CONDUIT AND WIRING IN CEILING SPACE OF FLOOR BELOW TO
ELECTRICAL CLOSET OR PANELBOARD.
COMBINATION POWER AND TELECOMMUNICATIONS FLOORPD
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
1/4" = 1'-0"
WORKSTATION - RESEARCH ASSOCIATE
II.A.3
10217_NIH NCATS R17 Guide Plates
III. SPACE AND OCCUPANCY REQUIREMENTS Page 61
III. SPACE AND OCCUPANCY REQUIREMENTS Page 62
III. SPACE AND OCCUPANCY REQUIREMENTS Page 63
Space Descriptions - Guideplates
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
OPEN LABORATORY - 3D VIEW
II.B.1-3D
III. SPACE AND OCCUPANCY REQUIREMENTS Page 64
Space Descriptions - Guideplates
CEILING PREFERENCE:
1. NO CELING WITH SERVICE PLATE (21 TOTAL)
2. CEILING WITH SERVICE STRIP
SINK,
HW, CW, EWES
EXHAUST PROVISIONS:
- FUTURE CAPACITY FOR (8) 6' FUME HOODS
- MAIN TRUNK TO THE OPEN LAB, NO
ADDITIONAL DISTRIBUTION WITHIN THE LAB
OVERHEAD SERVICES (TYP):
- CO2
- VACUUM
- CA (110 PSI) WITH REGULATOR
- (2) 4" EXHAUST PORTS PER MODULE
- (1) DEDICATED OUTLET ON UPS
- 120V AND 208V POWER
- DATA
SINK,
HW, CW, RO, DI
SINK,
HW, CW, EWES
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
1/16" = 1'-0"
OPEN LABORATORY
II.B.1
10217_NIH NCATS R17 Guide Plates
0 16' 32'
III. SPACE AND OCCUPANCY REQUIREMENTS Page 65
III. SPACE AND OCCUPANCY REQUIREMENTS Page 66
III. SPACE AND OCCUPANCY REQUIREMENTS Page 67
Space Descriptions - Guideplates
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
4 DEGREE COLD ROOM - 3D VIEW
II.C.1-3D
III. SPACE AND OCCUPANCY REQUIREMENTS Page 68
Space Descriptions - Guideplates
STAINLESS STEEL WIRE
SHELVING UNITS (GFGI)
FREEZER CONTROL PANEL
(4 DEGREE COLD ROOM)
NOTES:
- COMPRESSOR/CONDENSER SHALL BE
LOCATED IN CLOSET OR MECHANICAL SPACE
WITHIN 100' LF OF ROOM
- ROOM AND EQUIPMENT SHALL BE ON BACK-
UP POWER AND CONNECTED TO REES
EQUIPMENT MONITORING SYSTEM
COLD ROOM EVAPORATOR
COMPRESSOR /
CONDENSER UNIT
DOOR TO -20 COLD ROOM
-20 COLD ROOM (SHOWN GREY
FOR REFERENCE)
FREEZER CONTROL PANEL
(-20 DEGREE COLD ROOM)
THE DOOR BETWEEN THE 4
DEGREE AND ADJACENT
SPACE TO HAVE A LARGE
VIEW WINDOW WITH
DEFROST HEATING ELEMENT
FLOOR SINK FOR
CONDENSATE
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
1/8" = 1'-0"
4 DEGREE COLD ROOM
II.C.1
10217_NIH NCATS R17 Guide Plates
III. SPACE AND OCCUPANCY REQUIREMENTS Page 69
III. SPACE AND OCCUPANCY REQUIREMENTS Page 70
III. SPACE AND OCCUPANCY REQUIREMENTS Page 71
Space Descriptions - Guideplates
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
TISSUE CULTURE - 3D VIEW
II.C.3-3D
III. SPACE AND OCCUPANCY REQUIREMENTS Page 72
Space Descriptions - GuideplatesSpace Descriptions - Guideplates
ELECTRICAL RACEWAY
6' BSC CLASS 2 TYPE A2
EXIT TO OPEN LAB
MOVABLE TABLES
6' BSC CLASS 2 TYPE A2
INCUBATOR (STACKED) -
GFGI, CO2
FIXED BENCH
REFRIGERATOR - GFGI
SINK, CW, HW, RO, DI,
EWES
CONSUMABLE STORAGE
MICROSCOPE - GFGI
OVERHEAD SERVICE
CARRIER
INCUBATOR (STACKED) -
GFGI, CO2
INCUBATOR (STACKED) -
GFGI, CO2
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
1/8" = 1'-0"
TISSUE CULTURE
II.C.3
10217_NIH NCATS R17 Guide Plates
III. SPACE AND OCCUPANCY REQUIREMENTS Page 73
Space Descriptions - GuideplatesSpace Descriptions - Guideplates
No
Yes
Yes
Yes
No
No
Yes
No No
No
No
No No
Yes
Yes
No
No
Yes
No
No
NoYes
No
No
No Yes
No
Yes
No
Yes
No
No
Yes
TISSUE CULTURE II.C.3
TISSUE CULTURE
470 SF
Open Lab
OFFICE AREA
TISSUE PRINTING
21' - 4"
10' - 0" 22' - 0"
GROUP 4
Yes - Key Card
No No
RJ45 CAT6 Ethernet, Wireless
No n/a
73°F (SUMMER), 70°F (WINTER) +/- 2°F. ALL LABS TO BE
NEGATIVELY PRESSURIZED WITH SINGLE PASS AIR. (6) ACH IS
MINIMUM. A HIGHER ACH MAY BE REQUIRED DEPENDING ON
USE.
YES
CO2 IS REQUIRED TO ALL INCUBATORS
Negative
50% +/- 5% RH
Standard
(Per DRM Below)
75-100 FC
III. SPACE AND OCCUPANCY REQUIREMENTS Page 74
III. SPACE AND OCCUPANCY REQUIREMENTS Page 75
Space Descriptions - Guideplates
GUIDEPLATE SKETCH
SMITHGROUP JJR 1700 NEW YORK AVENUE NW, SUITE 100, WASHINGTON, DC 20006 T. 202.842.2100 www.smithgroupjjr.com
DRAWING TITLE
DRAWING SCALEDRAWING NUMBER
PROJECT NAME
PROJECT NUMBER
DRAFT
HISTOLOGY
II.C.4-3D
III. SPACE AND OCCUPANCY REQUIREMENTS Page 76
Space Descriptions - Guideplates
ELECTRICAL RACEWAY
6' FUME HOOD - CONNECT
TO BUILDING EXHAUST
EXIT TO OPEN LAB
MOVABLE TABLES
6' FUME HOOD - CONNECT
TO BUILDING EXHAUST
INCUBATOR (STACKED) -
GFGI, CO2
FIXED BENCH
REFRIGERATOR - GFGI
SINK, CW, HW, EWES
CONSUMABLE STORAGE
GROSSING STATION
MICROSCOPE - GFGI
OVERHEAD SERVICE
CARRIER
INCUBATOR (STACKED) -
GFGI, CO2
INCUBATOR (STACKED) -
GFGI,…