This specification document details equipment requirements for a Water Quality Multiprobe Datalogger Instruments procurement for the State of Maine. The agency seeks two multiprobe/datalogger instruments capable of long-term, unattended water quality monitoring. Each datalogger must include comprehensive sensor packages for temperature (range -5°C to 50°C, accuracy ±0.1°C), optical dissolved oxygen (luminescent sensor, accuracy ±0.1 mg/l for 0-20 mg/l range, drift <1% per year), pH measurement with separate reference and reading electrodes (±0.2 pH units accuracy), four-electrode specific conductance (1% accuracy), and depth measurement 0-25 meters (0.2% accuracy range with barometric pressure capability). The instruments must support internal memory storage of at least two months of data at 15-minute intervals or up to 80 MB capacity, feature diagnostic LED indicators, include one 5-meter and one 30-meter underwater communication cable, possess sealed waterproof connections, and include RS-232 interfaces with calibration cups, weighted sensor guards, detachable cable-lock systems, and bail assemblies. Power requirements include USB connectivity with download cables and internal battery packs for unattended operation.

Each datalogger package includes an integrated handheld computer featuring a crushproof and waterproof case with unconditional lifetime guarantee, integrated GPS, Pocket PC platform, 1 GB data expansion capacity, color display with 80 MB or greater memory, data averaging capability, and a 3.5-inch transflective screen. All equipment must include appropriate software, rechargeable battery packs, and supporting accessories for complete field deployment and data collection functionality.

View the file

Other files

File Type Posted
Solicitation_terms.doc DOC document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Two Water Quality Multiprobe/Datalogger Instruments. Each Datalogger needs to be capable of long term, unattended monitoring. Datalogger should be equipped with the following instrumentation/capabilities;

· All Dataloggers shall be supplied with download cables and must be able to connect via USB and be powered by USB.

· Ports capable of supporting the water quality sensors requested below and have the ability to add parameter sensors at a future time.

· Temperature Sensor - The Temperature Sensor shall have a range of -5 degrees C to 50 degrees C with an accuracy of ± 0.1 Degrees C.

· Optical Dissolved Oxygen Sensor - The Dissolved Oxygen Sensor shall be a luminescent oxygen sensor with an accuracy of ± 0.1 mg/l for readings from 0 – 20 mg/l. The optical dissolved oxygen sensor shall not require replaceable caps or membranes and have a drift of < than 1% per year.

· pH Sensor - The pH measurement system shall have separate reference and reading electrodes. The reference electrode shall be able to be refilled with pH electrolyte and have a replaceable flowing junction. The sensor shall be calibrated using two or three point calibration method with an accuracy of ± 0.2 pH units.

· Conductivity, 4 electrode - The Specific Conductance sensor shall be a graphite sensor with an accuracy of ± 1% of the reading.

· Depth Sensor (0-25 meters) - The Depth sensor shall be of range 0 – 25 meters with accuracy of 0.2% of range and be capable of measuring Barometric Pressure in mmHg for DO calibration.

· Internal memory capable of storing at least two months of data collecting at 15 minute intervals in the multiprobe or a 1 second interval within the display up to 80 MB

· Dataloggers shall have diagnostic LED’s to confirm connection

· Underwater Communication Cables (One 5 meter cable, One 30 meter cable)

· All connections on the Datalogger shall be sealed and waterproof

· RS-232 Interface

· Calibration and storage cups

· Weighted Sensor Guard

· Detatachable Cable-Lock system

· Bail Assembly.

· Internal Battery Pack for unattended monitoring.

· Integrated Handheld Computer w/ following specifications/functionality;

· Crushproof and waterproof case covered by unconditional lifetime guarantee

· Integrated GPS

· Pocket PC Platform

· 1 GB data expansion capacity

· Color Display

· Data averaging capable

· Display must have 80MB of memory or greater be equipped with all software, rechargeable battery pack and a 3.5” transflective screen