Criteria and Description of Water Purification.pdf
PDF 429 KB Posted
- Attached to
- Water Purification
- Solicitation number
- Req-348491
- Issued by
-
Department of Energy Office of Science
About this file
This document is a Criteria and Description for a Water Purification requirement under a federal contract opportunity.
The key details are:
- SLAC National Accelerator Laboratory, operated by Stanford University for the Department of Energy, has a requirement for 1 unit of water purification equipment that must meet specific criteria, including water purity, dispense flow rate, organic content, power requirements, and feed water compatibility.
- The contract opportunity is being competed using a Lowest Price Technically Acceptable (LPTA) evaluation method. Offerors must be registered in the System for Award Management (SAM) and comply with SLAC's Terms and Conditions for Fixed Price Commercial Supplies and Services.
- Proposals should include the quoted price, delivery timeline, technical compliance statement, and applicable certifications. The closing date and time are not specified.
- The resulting purchase order will be issued by SLAC and funded by the Department of Energy Office of Science, a civilian agency.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Quan ty 1, Water Purifica on Must meet the following criteria
Criteria Description
1. Water Purity The purity of the water at the output should be of DI quality with a resistivity at room temperature (25°C) of at least ≥18 MΩ-cm.
2. Dispense Flow Rate The water dispense flow rate at the output of the purifier should at least be ≥15 litres/minute
3. Organics The organic content of the dispensed water should be <10 ppb
4. Particles The particle filtration size should be upto 0.2 μm
5. Daily Usage (max) The maximum allowable amount of water dispensed per day should be >5000 litres/day.
6. Mains Input 110 Volts, 60 Hz
7. Power Consumption 22000 VA
8. Protection Rating 20 amps
9. Inlets & Outlets ¾" BSP
10. Power Supply The unit should be able to be connected to a local isolator with a single-phase industrial 20amp water-resistant plug/socket and cable arrangement
11. Feed water requirements The system should be compatible with feed water with a maximum fouling index: 10, Free Chlorine: <0.1ppm, Maximum Pressure (max): 5.5 bar (80psi) and Minumum Pressure (min): 3.5 bar (50 psi)
12. Storage tanks and fill rate Unit should have an integral storage tank of at ≥300 litres with a filling rate of ≥180 litres/hour
File details come from the government source that posted it. Updated .