C07_PR 1301408430 Performance Work Statment (PWS).pdf
PDF 109 KB Posted
- Attached to
- NSWCCD Emergency repairs, corrective maintenance, preventive maintenance (PM) of TA Instruments equipment
- Solicitation number
- N00167-26-SIMACQ-CD-61-0159
About this file
This is a Performance Work Statement (PWS) for calibration and maintenance services of Waters Technologies (TA Instruments) thermal analysis equipment at the Naval Surface Warfare Center, Carderock Division (NSWCCD), Non-Metallic and Low Observable Materials Branch (Code 6170).
The contractor shall provide professional preventative maintenance and annual calibration services for four primary thermal analysis systems: Discovery DSC 2500 System (Part No. 999800.705, Serial No. DSC24-03162), Discovery TGA 5500 System (Part No. 999800.559, Serial No. 5500-1116), Discovery Q1000/Q2000 DSC250 System (Part No. 999800.558, Serial No. 00034), and Discovery Q500 TGA550 System (Part No. 999800.560, Serial No. 0500-0016). Additionally, the contractor shall provide 4 hours of on-site technical support labor for connection setup, diagnostic checks, and system validation (Part No. 999800.500). All calibrations must be certified and valid for one year from the service date. Key maintenance actions include complete system inspection and cleaning, verification of gas cooling accessories and auto-sampler alignments, calibration using NIST-traceable reference standards, and validation testing to verify baseline flatness, sensitivity, and measurement precision. The contractor must deliver two formal deliverables: an Inspection Report (A001) and a Calibration Report (A002).
The contractor and assigned field service engineers must be factory-trained, authorized, and certified directly by Waters Technologies Corporation/TA Instruments, with demonstrated competence in servicing Discovery series DSC and TGA modules. All services shall be performed on-site at NSWCCD, 9500 MacArthur Boulevard, West Bethesda, Maryland 20817-5700, with a one-year period of performance from contract award. The security classification is Unclassified. Any replacement parts or materials must be shipped to the Central Receiving building (Building 143) during standard receiving hours (0600-1400 Monday-Thursday; 0600-1330 Friday).
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Performance Work Statement (PWS)
Waters Technologies (TA Instruments) Calibration and Maintenance Services
1.0 Description/Background:
The Non-Metallic and Low Observable Materials Branch, Code 6170, of the Naval Surface Warfare
Center, Carderock Division (NSWCCD) requires calibration, performance maintenance, and technical support services for government-owned Waters Technologies Corporation (TA Instruments) thermal analysis equipment currently utilized by the Branch. This specialized equipment is essential for testing qualified materials and supporting program requirements for multiple NAVSEA projects. The procurement ensures that all systems are properly verified, serviced, and calibrated to original manufacturer specifications to prevent project delays and ensure continuous fleet support.
2.0 Objective:
The Contractor shall provide professional preventative maintenance and calibration services for the specified thermal analysis equipment. All calibrations performed under this contract shall be certified and valid for a period of one (1) year from the date of service.
The contract scope covers the following equipment modules, aligned with the manufacturer's quote
(Quote # 24579363) options and matching active government contract inventories (excluding Section 2.4)
2.1 Discovery DSC 2500 System:
2.1.1 Part Number: 999800.705 (PM Visit Charge DDSC DSC2500 DSC-X3)
2.1.2 Serial Number: DSC24-03162
2.1.3 Service Required: Performance Maintenance Visit (PMV) including labor and certified calibration with traceable standards.
2.2 Discovery TGA 5500 System:
2.2.1 Part Number: 999800.559 (PM Visit Charge Q5000IR DTGA TGA5500)
2.2.2 Serial Number: 5500-1116
2.2.3 Service Required: Performance Maintenance Visit (PMV) including labor and certified calibration with traceable standards.
2.3 Discovery Q1000/Q2000 DSC250 System:
2.3.1 Part Number: 999800.558 (PM Visit Charge Q1000/Q2000 DSC250)
2.3.2 Serial Number: 00034
2.3.3 Service Required: Performance Maintenance Visit (PMV) including labor and certified calibration with traceable standards.
2.4 Discovery Q500 TGA550 System:
2.4.1 Part Number: 999800.560 (PM Visit Charge Q500 TGA550)
2.4.2 Serial Number: 0500-0016
2.4.3 Service Required: Performance Maintenance Visit (PMV) including labor and certified calibration with traceable standards.
2.5 Thermal Analysis Modules (On-Site Labor):
2.5.1 Part Number: 999800.500 (Labor Thermal Analysis)
2.5.2 Quantity: 4 Hours
2.5.3 Service Required: On-site technical support for connection setup, diagnostic checks, and system validation of the thermal analysis modules.
3.0 Performance Maintenance & Annual Calibration/Certification:
The Contractor shall perform on-site performance maintenance and annual calibration services on the equipment listed in Section 2.0. The services performed by the Contractor’s service engineer must guarantee that all instruments meet original factory performance specifications.
Key maintenance actions include:
3.1 Complete system inspection, cleaning, and replacement of standard worn items.
3.2 Verification of gas cooling accessories (GCA), furnaces, and auto-sampler alignments.
3.3 Calibration using NIST-traceable reference standards.
3.4 Validation testing to verify baseline flatness, sensitivity, and measurement precision.
3.5 Provision of formal calibration certificates and comprehensive maintenance check sheets.
4.0 Special Qualifications:
The Contractor and any field service engineers assigned to execute this PWS must be factory-trained, authorized, and certified directly by the original equipment manufacturer (Waters Technologies
Corporation / TA Instruments). The engineers must possess demonstrated competence in servicing high-precision thermal analysis instrumentation, including Discovery series DSC and TGA modules, Horizontal Gas Purge systems.
5.0 Deliverables:
The Contractor shall deliver the physical services and corresponding documentation for each piece of serviced equipment. All electronic reports shall be submitted directly to the Technical Point of Contact
(TPOC) upon completion of the on-site visits.
Number Name
A001 Inspection Report
A002 Calibration Report
6.0 Period, Place of Performance, and Delivery Information:
6.1 Period of Performance: One (1) year from the date of contract award.
6.2 Place of Performance: All services, including the 4-hour technical support labor and
Performance Maintenance Visits (PMV), shall be performed on-site at the government facility:
All services, including the 4-hour technical support labor and Performance Maintenance Visits
(PMV), shall be performed on-site at the government facility:
6.3 Delivery & Shipping Information: Any replacement parts, accessories, or materials shipped in support of this contract must be routed through the central receiving building during standard receiving hours (0600-1400 Monday through Thursday; 0600-1330 Friday):
6.4: Shipping Address: NSWCCD, Central Receiving (Attn: Code 6170 / Bldg. 143), 9500 MacArthur Blvd, West Bethesda, MD 20817-5700.
7.0 Security Requirements: Unclassified