C04a Statement of Work 26Q0556.pdf

PDF 342 KB Posted

Attached to
Consolidated Sterilizer System (Brand Name SR-24D or Equal) Federal contract opportunity
Solicitation number
1232SA26Q0556
Issued by
Department of Agriculture Agricultural Research Service

About this file

This is a Statement of Work (SOW) for the U.S. Department of Agriculture (USDA) for Project Autoclave FY26-004859.

The scope of work requires procurement of an autoclave in the style of a Consolidated Sterilizer Systems SR-24D, including removal and disposal of the existing autoclave and installation of the new unit with full utility connections. The existing autoclave at the National Laboratory for Agriculture and the Environment is located on the 3rd floor and is reaching the end of its lifecycle. The equipment must fit through a 46½" x 84" door opening and will be transported via a freight elevator to the 3rd floor. The facility has a loading dock with a dock leveler but no forklifts on-site. Utility connections include potable (softened) water and compressed air drops from the ceiling, with steam and floor drain located on the right side of the existing autoclave position.

The technical specifications require a Consolidated Sterilizer System SR-24D or equivalent with the following key features: 24" x 36" x 48" chamber constructed of 316L stainless steel with carbon steel jacket, passivated finish, right-swing open door, floor-standing mounting, thermal printer, electronic USB data storage, configurable cycle parameters, temperature range of 100°C to 135°C, hi-vacuum system with water ejector and booster pump, brass/bronze/copper piping, water conservation features, automatic jacket blowdown, load control probe temperature control, one stationary upper shelf, one stationary lower shelf, and emergency stop button. The contract must include removal and disposal of the existing autoclave, installation, and complete utility hook-up of the new equipment. No key personnel, travel, deliverables schedule, security requirements, data rights, or Section 508 compliance specifications are applicable to this SOW.

View the file

Other files for this federal contract opportunity

Other files attached to Consolidated Sterilizer System (Brand Name SR-24D or Equal), newest first.
File Type Posted
C04 Sol_1232SA26Q0556.pdf PDF
C04b 52.225-2BuyAmericanCert.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

US Department of Agriculture (USDA)

Statement of Work

Contents Project

General Information

1.0 Scope of Work

2.0 Background

Contractor Requirements

3.0 Technical Requirements / Tasks

4.0 Government Furnished ..................................................................................................................... Error! Bookmark not defined.

5.0 Deliverables / Schedule

6.0 Travel

7.0 Contractor’s Key Personnel

8.0 Security Requirements

9.0 Data Rights

10.0 Section 508 – Electronic and Information Technology Standards

Attachment

Project Autoclave FY26-004859

General Information

1.0 Scope of Work

Procurement of autoclave in the style of a Consolidated Sterilizer Systems SR-24D, removal and disposal of existing autoclave and installation of new including full utility connections.

2.0 Background

The National Laboratory for Agriculture and the Environment existing autoclave is approaching the end of its lifecycle.

The autoclave resides on the 3rd floor of the laboratory. There is a freight elevator capable of transporting the autoclave to the 3rd floor. There is a loading dock with a dock leveler. There are no forklifts on site, nor would a forklift be able to traverse to the 3rd floor. The door to the autoclave room is 46 ½” x 84”. See photos below for utility connections. The potable water (softened) and compressed air drop from the ceiling. The steam and floor drain are located on the right side of the existing autoclave.

The basis of this request is a Consolidated Sterilizer System SR-24D or similar.

Equipment Requirements

3.0 Technical Requirements / Tasks

Consolidated Sterilizer System SR-24D (Brand Name or Equal)

• Chamber Dimension: 24” x 36” x 48”

• Chamber Construction: 316L Stainless Steel with Carbon Steel Jacket

• Finish: Passivated

• Door: Right swing open

• Mounting: Floor-standing

• Thermal Printer

• Electronic Data Storage via USB

• Green® Features

• Configurable Cycle Parameters

• Battery Backup of Program, Memory & Config Data

• Temp Range: 100°C - 135°C

• Hi-Vacuum System with Water Ejector and Booster Pump

• Piping: Brass, Bronze and Copper

• Water Conservation

• Automatic Jacket Blowdown

• Temp Controlling Feature: Load Control Probe

• Shelves: One (1) Stationary Upper, One (1) Stationary Lower

• Emergency Stop Button Installation

• Contract to include removal and disposal of existing autoclave

• Installation and utility hook up of new autoclave

• Autoclave must fit through 46 ½” x 84” door opening

5.0 Deliverables / Schedule

N/A

6.0 Travel

N/A

7.0 Contractor’s Key Personnel

https://consteril.com/wp-content/uploads/2018/08/CSS-Medium-Series-Spec-Sheet-high-res.pdf

8.0 Security Requirements

9.0 Data Rights

N/A

10.0 Section 508 – Electronic and Information Technology Standards

Attachment existing (control panel was removed for installation)

File details come from the government source that posted it. Updated .