View the file

Other files

File Type Posted
36C25821B0009 A0004.docx DOCX document
S05 MASTER - IFB 36C25821B0009 Responses to RFI 001 - 016.pdf PDF
36C25821B0009 A0003.docx DOCX document
Tyvek Bunny Suit - PPE.jpg JPG image
S04 Pre-Bid Virtual Conference Sign-In Sheet - 664-CSI-120.docx DOCX document
S04 Pre Bid Tele-Conference Agenda and Minutes.pdf PDF
36C25821B0009 A0002.docx DOCX document
S06 36C25821B0009 A0001.docx DOCX document
A1 Drawing - Part-3-4.zip ZIP file
B2 Operators Manual WD290-09.pdf PDF
Self Performance Calculations.pdf PDF
A1 Drawing - Part-9-10.zip ZIP file
A1 Drawing - Part-7-8.zip ZIP file
A1 Drawing - Part-1-2.zip ZIP file
A1 Drawing - Part-5-6.zip ZIP file
A2 Drawing - 1 of 2.zip ZIP file
WD CA20210001 Mod 1 01-08-2021.pdf PDF
A2 Drawing - 2 of 2.zip ZIP file
P01 SOW_664-CSI-120_FINAL.docx DOCX document
36C25821B0009.docx DOCX document
Show all 20

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

I n f e c t i o n C o n t r o l

Dwg. No. :

Drawn By :

Rev. :

Scale :

of

Page :

Date :

Tel: (800) 457-4117 Web: www.belimed.com

2325 Charleston Regional Parkway, Charleston, SC 29492

Typical installation of a Belimed WD290

9/19/14

3/4" = 1'

FCUT-WD290-09

BELIMED WD290

WASHER/DISINFECTOR

Washer/Disinfector.

DATA PORT (BY OTHERS)

110V GFCI RECEPTACLE (BY OTHERS)

JUNCTION BOX

(BY OTHERS)

FLEX HOSES (BY BELIMED)

SHUTOFF VALVES (BY OTHERS)

STEAM PARTICLE STRAINER

(BY OTHERS)

OPTIONAL STAINLESS

STEEL TRIM PACKAGE

(BY BELIMED)

CONDENSATE TRAP

(BY BELIMED)

FLEX DUCT

(BY BELIMED)

AIR EXHAUST DUCT

(BY OTHERS)

CONDENSATE RETURN

(BY OTHERS)

STEAM SUPPLY

(BY OTHERS)

HOT WATER

(BY OTHERS)

COLD WATER

(BY OTHERS)

DEIONIZED WATER

(BY OTHERS)

COPPER DRAINPIPE

(BY OTHERS)

OPTIONAL ENCLOSURE

PANEL (BY BELIMED)

MANUAL DAMPER

(BY OTHERS)

Drawn By :

Rev. :

Scale :

of

Page :

Date :

Tel: (800) 457-4117 Web: www.belimed.com

2325 Charleston Regional Parkway, Charleston, SC 29492

Typical installation of a Belimed WD290

9/19/14

3/8" = 1'

FCUT-WD290-09

BELIMED WD290

WASHER/DISINFECTOR

TECHNICAL DRAWING

Washer/Disinfector.

LEGEND

SERVICE SHUT-OFF VALVE, BY OTHERS

STEAM PARTICLE STRAINER, BY OTHERS

PLAN VIEW

SIDE ELEVATIONDECONTAM ELEVATION

MANUAL EXHAUST DAMPER, BY OTHERS

TYPICAL SEISMIC ANCHORAGE

Drawn By :

Rev. :

Scale :

of

Page :

Date :

Tel: (800) 457-4117 Web: www.belimed.com

2325 Charleston Regional Parkway, Charleston, SC 29492

Typical installation of a Belimed WD290

9/19/14

NTS

FCUT-WD290-09

BELIMED WD290

WASHER/DISINFECTOR

TECHNICAL SPECIFICATIONS

Washer/Disinfector.

CONTINUED ON PAGE 4

Drawn By :

Rev. :

Scale :

of

Page :

Date :

Tel: (800) 457-4117 Web: www.belimed.com

2325 Charleston Regional Parkway, Charleston, SC 29492

Typical installation of a Belimed WD290

9/19/14

NTS

FCUT-WD290-09

BELIMED WD290

WASHER/DISINFECTOR

TECHNICAL SPECIFICATIONS (CONT.)

Washer/Disinfector.

File details come from the government source that posted it. Updated .