This document is a Salient Characteristics specification for a Refrigerated/Heating Circulator (H26002), outlining the technical requirements for a single unit to be procured.

The refrigerated/heating circulator must be a circulating bath system with a stainless-steel tank of 50-liter capacity featuring insulated construction, integrated secure cover and drain, and a usable bath opening of 13.8" W x 15.7" L x 11.8" D. The unit shall have overall dimensions of 17.7" W x 25.2" L x 37.4" H and operate across a temperature range of -40 to 100° C with temperature stability of ±0.01° C or better and resolution of 0.01° C or better. The circulator must be capable of 3-point absolute temperature calibration and include a color TFT display with programmable controller and rotary knob interface. The system shall be switchable between internal and external circulation modes with integrated stainless steel pump connections and a port for external temperature control. The unit must utilize environmentally friendly R1270 (propane) refrigerant at 90 grams with a Global Warming Potential of 2, deliver minimum cooling capacity of 1.1 kW at 20° C and heating capacity of 2 kW, operate from 230 VAC at 60 Hz, and meet or exceed Class III (FL) certification per DIN 12876-1. Additional requirements include standard RS-232 and USB ports, integrated safety features for operator and equipment protection, all necessary software for setup and operation, and all essential startup components including hose adapters, barbs, fittings, and US power cord.

View the file

Other files

File Type Posted
B08_ATCH_2_508.pdf PDF
B08_ATCH_3_FISMA.pdf PDF
Sol_140G0126Q0084.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristics H26002

Refrigerated/Heating Circulator

Salient Characteristics

1 ea. Refrigerated/Heating Circulator

1. Shall consist of a refrigerated/heating style circulating bath with color display, programmable controller, and fluid circulation capability.

2. Shall have a stainless-steel bath tank with insulated construction. Tank shall also have an integrated secure cover and drain, and have a capacity of 50 L.

3. Shall have a usable bath opening of 13.8” W x 15.7” L x 11.8” D.

4. Shall have overall dimensions of 17.7” W x 25.2” L x 37.4” H.

5. Shall have an operational temperature range of at least -40 to 100o C.

6. Shall be switchable between internal and external circulation

7. Shall include integrated stainless steel pump connections.

8. Shall include an integrated programmer with a TFT display and central rotary knob user interface.

9. Shall use environmentally friendly natural refrigerant: R1270 (propane), 90 g, GWP = 2.

10. Shall include a port for external temperature control (sensor) capability.

11. Shall include a standard RS-232 and USB port.

12. Shall be capable of a minimum 3-point absolute temperature calibration.

13. Shall have temperature stability of +/-0.01o C, or better and resolution of .01o C, or better.

14. Shall have a cooling capacity of at least 1.1 kW at 20o C and a heating capacity of at least 2 kW.

15. Shall be capable of operating from 230 VAC, 60 Hz, power.

16. Shall meet or exceed Class III (FL) per DIN 12876-1 certification.

17. Shall include features that detect and protect the safety of the operator and the equipment during periods of abnormal or unsafe operating conditions.

18. Shall include all applicable software, as required, for setup and operation.

19. Shall include all essential parts/hardware including hose adapters, barbs, fittings, US power cord, etc. as needed for start-up and operation.