Salient Characteristics

General

Zeiss OS HD-OCT 6000, with AngioPlex OCT Angiography Version 9.5 - Brand Name or Equal

Package Items Summary

Item Number
Item Description
Quantity
0001
PRO Zeiss OS Cirrus HD-OCT Model (6000) with AngioPlex oCT Angiography Version 9.5 – with one (1) year mfg. warranty - Brand Name or Equal
1
0002
Installation
1
0003
Training
1

Delivery Requirements:

Need Date: March 1, 2021 – May 30, 2021

Items shall be delivered no earlier than the estimated delivery start date and no later than the delivery end date which is subject to change based upon the current construction schedule. Vendors shall not increase pricing in the event of a government-caused delivery delay for up to 30 calendar days.

The vendor will Coordinate delivery prior to scheduling delivery of items.

Delivery shall be (FOB) – Destination.

COVID-19 Contractor employees that will be on-site at the VA Palo Alto Campus must comply with all safety and security requirements to include COVID-19 safety requirements.

Deliver to:

VA Palo Alto Health Care System 3801 Miranda Ave.

Bldg. 710 Loading Dock Palo Alto, CA 94304

Salient Characteristics

· Angiography version 9.5.

· To include GCGPA, Smart HD Scans, FastTrac, Adv RPE, GCA, Ant Seg Basic.

· Capability to capture and provide OCT Angiography provided by a 1 single OCT scan that allows for visualization of both micro vascular and structural information in a noninvasive dye free scan by the OCT.

· Ability to import and integrate the Cirrus model HD OCT 4000 raw data. Raw data must be interchangeable between the Cirrus 4000 and Cirrus 5000 unit.

· Unit must have exact Macular Scan Pattern of 512x128 for retina scan with exact scanning speed at 27,000 A-scans a second with a 6mmx6mm. Must be able to provide on Company Literature.

· Unit must exact scan pattern of 200x200 in a 6mmx6mm cube for Glaucoma Scan with exact scanning speed at 27,000 A-scans a second. The scan pattern must be able to capture and provide RNFL, ONH, Neuro Retinal Rim Tissue, and 3-D Cube capabilities all in one scan Scanning speeds: Scan at exactly 27,000 A-scans/sec, as well as the ability to scan at 68,000 A-scans/sec.

· Unit must be able to provide a report for Ganglion Cell Analysis that is derived from a scan must be comprised of 6mmx6mm cube scan with a minimum be 200x200 line scan patter with no more than 30 microns between each slice.

· Minimum Pupil Requirements must be 2.0mm or less.

· Unit must utilize an LSO Fundus Camera (Line Scanning Laser Ophthalmoscope).

· Unit must Automatically Track to Prior Scan: The unit must be able to do this without technician needing to align to prior scan: Anterior Segment scan that is at minimal 15.5mm in length x 5.8mm in depth. Must be able to provide on Company Literature.

· Global Pachymetry report that is comprised of minimum of 24 radial scans and must be at a minimum 9mm in length.

· The unit must be a 90 degree configuration, so that the technician must sit to the side (90 degrees) of the patient.

· Fundus Imaging: Methodology – Line scanning ophthalmoscope (LSO).

· Iris Imaging: Methodology – CCD Camera.

· Fixation: Internal fixation focus adjustment – 20D to +20D (diopters).

· Internal Computer: Operating system/processor – Windows 10, i& Intel processor.