Attachment E - Salient Characteristics.pdf
PDF 206 KB Posted
- Solicitation number
- RFQ-22-PHX-056
About this file
This combined synopsis and solicitation requests offers for brand name or equal GE AISYS CS2 anesthesia equipment, accessories, installation, configuration, preventative maintenance, and repair services for the Phoenix Indian Medical Center. The Indian Health Service intends to award a firm-fixed price contract.
Offerors must provide pricing on the attachment A price schedule by September 1st, 2022. The contract will have a period of performance of one year. Technical factors will be evaluated on an acceptable/unacceptable basis and will include a description of equipment being offered as equal and past performance references. Price will be the most important factor, and award will be made to the lowest priced technically acceptable offer. The place of performance will be the Phoenix Indian Medical Center.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment A - Price Schedule.xlsx | XLSX spreadsheet | |
| Attachment F - SOW Preventative Maintenance and Repair.docx | DOCX document | |
| Attachment B - Clause Admin Data.docx | DOCX document | |
| Combined Synop RFQ-22-PHX-056.docx | DOCX document | |
| Attachment D - Equipment Service Location Listing.docx | DOCX document | |
| Attachment C - 2022 Tax Exempt Letter.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
2 3
Physical Specifications Dimensions Height: 133.9 cm/52.7 in
| Height | (with | vertical | arm): | 190.5 | cm/75.0 | in |
| 211 | cm/83.1 | in |
Width: 68 cm/26.8 in
Depth: 82 cm/32.3 in
Weight: 190 kg/419 lbs
Top shelf Weight limit: 45 kg/100 lb
Width: 55 cm/21.65 in
Depth: 51.6 cm/20.31 in
Upper shelf (optional) Weight limit: 23 kg/50 lb
Width: 54.8 cm/21.57 in
Depth: 44.4 cm/17.48 in
Work surface Height: 87.5 cm/34.4 in
Size: 2684.2 cm2/416 in2
Folding side shelf (optional) Weight limit: 12 kg/25 lb
Height: 88.17 cm/34.7 in
Width: 27.7 cm/10.91 in
Depth: 36.6 cm/14.41 in
DIN rail (optional) Side of machine: 53.9 cm/21.22 in
Drawers (internal dimensions) Small
Height: 10.5 cm/4.13 in
Width: 37.80 cm/14.88 in
Depth: 37.64 cm/14.82 in
Large
Height: 15.0 cm/5.91 in
Width: 37.80 cm/14.88 in
Depth: 37.64 cm/14.82 in
Absorber bag arm (optional) Arm length: 39.8 cm/15.67 in
| Bag | arm | height | (adjustable): | 98 | cm/38.6 | in |
| 123 | cm/48.4 | in |
Casters Diameter: 12.5 cm/5 in
Brakes: Central brake
Pendant mounting interface (optional)*
Height from floor: 76 cm/29.92 in
Suspended mass limit: 364 kg/800 lb
Ventilator Operating Specifications Modes of ventilation (standard) Volume Control Mode with tidal volume compensation
Modes of ventilation (optional) Pressure Control and PCV-VG (Pressure control volume guaran-tee)
| Synchronised | Intermittent | Mandatory | Ventilation | (SIMV) |
| (volume, | pressure | and | PCV-VG) |
PSVPro (Pressure Support with Apnea backup)
CPAP+PSV (Pressure support mode)
Notification of spontaneous breathing
| Patient-generated | breaths | will | change | pressure | and | |
| flow | waveform | color | for | immediate | clinician | notification |
Ventilation parameters Tidal volume range: 20 to 1500 mL
| (Volume | Control, | PCV-VG, | SIMV |
| and | SIMV | PCV-VG | modes) |
| Incremental | settings: | 20 | to | 50 | mL | (increments | of | 1 | mL) | |||||
| 50 | to | 100 | mL | (increments | of | 5 | mL) | |||||||
| 100 | to | 300 | mL | (increments | of | 10 | mL) | |||||||
| 300 | to | 1000 | mL | (increments | of | 25 | mL) | |||||||
| 1000 | to | 1500 | mL | (increments | of | 50 | mL) |
Minute volume range: Less than 0.1 to 99.9 L/min
Pressure (P) Inspired range: 5 to 60 cmH2O
(increments of 1 cmH2O)
5 to 1500 mL volume delivery
| Pressure | (P) | max | range: | 12 | to | 100 | cmH2O |
| (increments | of | 1 | cmH2O) |
| Pressure | (P) | support | ||||
| range: | Off, | 2 | to | 40 | cmH2O |
(increments of 1 cmH2O)
Rate: Rate: 4 to 100 breaths per minute for
Volume Control, Pressure Control and
| PCV-VG; | 2 | to | 60 | breaths | per | minute |
| for | SIMV, | PSVPro, | SIMV | PCV-VG; | 4 | to |
| 60 | breaths | per | minute | for |
CPAP+PSV
(increments of 1 breath per minute)
Ventilator Operating Specifications (continued)
Inspiratory/expiratory ratio: 2:1 to 1:8 (increments of 0.5)
| Inspiratory | time: | Inspiratory | time: | 0.2 | to | 5.0 | seconds |
| (increments | of | 0.1 | seconds) | ||||
| (SIMV, | PSVPro, | SIMV | PCV-VG | ||||
| and | CPAP+PSV) |
Trigger window: 0 to 80% (increments of 5%)
| Flow | trigger: | 1 | to | 10 | L/min | (increments | of |
| 0.5 | L/min) |
| 0.2 | to | 1 | L/min | (increments | of |
| 0.2 | L/min) |
| Inspiration | termination | ||||||
| level: | 5 | to | 75% | (increments | of | 5%) | - |
| Rise | Rate | 1-10 | (PCV, | PCV-VG, | |
| SIMV, | PSVPro, | CPAP+PSV | and | ||
| SIMV | PCV-VG) |
Inspiratory Pause range: 0-60%
Positive End Expiratory Pressure (PEEP) Type: Integrated, electronically controlled
| Range: | OFF, | 4 | to | 30 | cmH2O |
| (increments | of | 1 | cmH2O) |
Ventilator performance Pressure range at inlet: 240 kPa to 700 kPa/35 psig to 102 psig
Peak gas flow: 120 L/min + fresh gas flow
Flow valve range: 1 to 120 L/min
| Flow | compensation | ||||
| range: | 200 | mL/min | to | 15 | L/min |
Patient Spirometry Pressure-volume loop
Flow-volume loop
Pressure-flow loop
Airway pressure and flow waveforms
Adjustable low and high alarm limits for Ppeak and MVexp
Detection through machine flow transducers.
Ventilator Accuracy Delivery/monitoring accuracy Volume delivery: > 210 mL = better than 7%
| ≤ | 210 | mL | = | better | than | 15 | mL |
| < | 60 | mL | = | better | than | 10 | mL |
Pressure delivery: ±10% or ±3 cmH2O
PEEP delivery: ±1.5 cmH2O
| Volume | monitoring: | > 210 | mL | = | better | than | 9% |
| ≤ | 210 | mL | = | better | than | 18 | mL |
| < | 60 | mL | = | better | than | 10 | mL |
Pressure monitoring: ±5% or ±2 cmH2O
Alarm settings Tidal volume (VTE): Low: OFF, 1 to 1500 mL
High: 20 to 1600 mL, OFF
| Minute | volume | (VE): | Low: | OFF, | 0.1 | to | 10 | L/min |
| High: | 0.5 | to | 30 | L/min, | OFF |
| Inspired | oxygen | (FiO2): | Low: | 18 | to | 99% |
| High: | 19 | to | 100%, | OFF |
| Apnea | alarm: | Mechanical ventilation ON: | ||
| < | 5 | mL | breath | measured |
| in | 10 | to | 30 | seconds, |
| increments | of | 1 | second |
Mechanical ventilation OFF:
| < | 5 | mL | breath | measured |
| in | 10 | to | 30 | seconds, |
| increments | of | 1 | second |
Low airway pressure: 4 cmH2O above PEEP
| High | pressure: | 12 | to | 100 | cmH2O |
| (increments | of | 1 | cmH2O) |
| Sustained | airway |
| pressure: | Mechanical ventilation ON: |
| (P) max | < | 30 | cmH2O, | ||||
| the | sustained | limit | is | 6 | cmH2O | ||
| (P) max | 30 | to | 60 | cmH2O, | |||
| the | sustained | limit | is | 20% | of | (P) | max |
| (P) max | > | 60 | cmH2O, | ||||
| the | sustained | limit | is | 12 | cmH2O |
PEEP and mechanical ventilation ON:
| Sustained | limit | increases | ||
| by | PEEP | minus | 2 | cmH2O |
Mechanical ventilation OFF:
| (P) max | ≤ | 60 | cmH2O, | ||||
| the | sustained | limit | is | 50% | of | (P) | max |
| (P) max | > | 60 | cmH2O, | ||||
| the | sustained | limit | is | 30 | cmH2O |
Subatmospheric pressure: Paw < -10 cmH2O
| Alarm | silence | ||||
| countdown | timer: | 120 | to | 0 | seconds |
| * Interface | compatible | with | Kreuzer, | Dräger | and | ceiling | columns. | |
| Contact | your | local | GE | Healthcare | representative | for | solutions | to |
| other | ceiling | column | manufacturers. Anaesthesia | delivery | screen |
Attachment E - Salient Characteristics
4 5
Ventilator Components Flow transducer Type: Variable orifice flow sensor
Dimensions: 22 mm OD and 15 mm ID
| Location: | Inspiratory | outlet | and |
| expiratory | inlet |
(Optional autoclavable sensor available)
Oxygen sensor
| Type: | Optional | galvanic | fuel | cell | or |
| paramagnetic | with | Respiratory | |||
| Module | option |
Ventilator screen Display size: 38 cm/15 in
Pixel format: 1024 (H) x 768 (V)
Communication ports RS-232C compatible serial interface
Ethernet
Datex-Ohmeda device interface solutions port
USB port
VGA Output
Aladin2 Cassette Anaesthetic agent delivery
| Vaporizer: | Aladin2 | Cassette | - | Available |
| with | Isoflurane, | Desflurane, | ||
| Sevoflurane | and | Enflurane |
| Number | of | |
| active | positions: | 1 |
Dimensions Height: 7 cm/2.76 in
Depth: 24 cm/9.45 in
Width: 14 cm/5.51 in
Empty weight: 2.8 kg/6.2 lb
Cassette handling No restriction for tilting during storage or handling
Agent capacity Total (Enf, Iso, Sev): 220 mL
Total (Des): 240 mL
| When | cassette | indicator | ||||||||
| shows | empty | (Enf, | Iso, | Sev): | 125 | mL | (95 | mL | residual | volume) |
| When | cassette | indicator | ||||||
| shows | empty | (Des): | 140 | mL | (100 | mL | residual | volume) |
Accuracy
| All | agents | in | typical | operating | conditions. | Fresh | gas | flow | |
| range | 1.0 | to | 10 | L/min. | Ambient | temperature | 18° | to | 25°C/ |
| 64.4° | to | 77°F. |
| Enflurane, | Isoflurane, | ||||||
| Sevoflurane: | ±0.2% | v/v | of | full | scale | or | ±10% |
| of | setting | (whichever | is | greater) |
| Desflurane: | ±0.5% | v/v | of | full | scale | or | ±10% |
| of | setting | (whichever | is | greater) |
| In | other | operating | conditions. | Fresh | gas | flow | range | 0.2 | to |
| 10 | L/min. | Ambient | temperature | 10° | to | 35°C/50° | to | 95°F. |
| Enflurane, | Isoflurane, | ||||||
| Sevoflurane: | ±0.4% | v/v | of | full | scale | or | ±20% |
| of | setting | (whichever | is | greater) |
| Desflurane: | ±1.0% | v/v | of | full | scale | or | ±20% | of |
| setting | (whichever | is | greater) |
Agent setting ranges
| Enflurane | and | Isoflurane: | OFF, | 0.2 | to | 5% | in | fresh | gas | flow, |
| resolution | 0.1% |
| Sevoflurane: | OFF, | 0.2 | to | 8% | in | fresh | gas | flow, |
| resolution | 0.1% |
| Desflurane: | OFF, | 1.0 | to | 18% | in | fresh | gas | flow, |
| resolution | 0.2% |
CARESCAPE Respiratory Modules General specifications
E-sCAiO, E-sCAiOV, E-sCAiOE, E-sCAiOVE
Size (W x D x H): 3.8 x 20.5 x 11.3 cm/1.5 x 8.1 x 4.4 in
Weight: 0.7 kg/1.5 lb
Sampling rate: 120 ±20 mL/min
| Automatic | compensation | for | atmospheric | pressure | variation | ||||
| (500 | to | 800 | mmHg), | temperature, | and | CO2, | O2, | N2O | and |
| anaesthetic | agent | cross | effects. |
Desflurane
| Measurement | range: | 0 | to | 20 | vol% | ||||
| Accuracy: | ± | (0.15 | vol% | + | 5% | of | the | reading) |
Waveform displayed
MAC value displayed
Identification threshold: 0.15 vol %
Agent mixture detection
GE Datex-Ohmeda infrared sensor
Adjustable high and low alarm limits for EtAA and FiAA
Patient Spirometry (available in GE Datex-Ohmeda Anaesthesia Monitor module bay)
Note: For ventilation parameters reference the ventilator operating specifications
Pressure-volume loop
Flow-volume loop
Pressure-flow loop
Airway pressure and flow waveforms
Adjustable low and high alarm limits for Ppeak and MVexp
| Detection | through | Adult | D-lite | or | D-lite(+) | and | Pediatric | |
| Pedi-lite | or | Pedi-lite(+) | flow | and | gas | sampling | sensor | with |
| following | specifications: |
D-lite Pedi-lite and D-lite(+) and Pedi-lite(+) Respiration rate: 4 to 35 bpm 4 to 70 bpm
Tidal volume
| Measurement | range: | 150 | to | 2000 | mL | 5 | to | 300 | mL |
| Accuracy: | greater | of | greater | of | |||||
| (±6% | or | 30 | mL) | (±6% | or | 4 | mL) |
Minute volume Measurement range: 2 to 20 L/min 0.1 to 5 L/min
Airway pressure
| Measurement | range: | -20 | to | +100 | -20 | to | +100 |
| cmH2O | cmH2O |
Accuracy: ±1 cmH2O ±1 cmH2O
Display units: cmH2O, mmHg, kPa, mbar, hPa
Flow
| Measurement | range: | -100 | to | +100 | -25 | to | +25 |
| L/min | L/min |
I:E I:E ratio: 1:4.5 to 2:1
Compliance
| Measurement | range: | 4 | to | 100 | 1 | to | 100 |
| ml/cmH2O | ml/cmH2O |
Airway resistance Measurement range: 0 to 200 cmH2O/L/s
CARESCAPE Respiratory Modules (continued)
Non-disturbing gases
| Ethanol, | acetone, | methane, | nitrogen, | nitric | oxide, | |
| carbon | monoxide, | water | vapor, | isopropanol, | freon | R134A. |
| Maximum | effect | |||||||||
| on | readings: | CO2 | < | 0.2 | vol%; | O2, | N2O | < | 2 | vol%; |
| Anaesthetic | agents | < | 0.15 | vol% |
Carbon dioxide (CO2) EtCO2: End-tidal CO2 concentration
FiCO2: Inspired CO2 concentration
CO2 waveform
| Measurement | range: | 0 | to | 15 | vol% | ||||
| (0 | to | 15 | kPa, | 0 | to | 113 | mmHg) |
Accuracy: ± (0.2 vol% +2% of the reading)
GE Datex-Ohmeda infrared sensor
Adjustable low and high alarm limits for EtCO2 and FiCO2
Respiration rate (RR) Measurement range: 4 to 100 breaths per minute
Detection criteria: 1% variation in CO2
| Accuracy: | ±1 | breaths | per | minute | |||
| (at | 4 | to | 20 | breaths | per | minute) |
| ± | 5% | (at | 20 | to | 100 | breaths |
| per | minute) |
| Adjustable | low | and | high | alarm | limits | for | respiration | rate; |
| alarm | for | apnea |
Patient Oxygen (O2) FiO2: Inspired O2 concentration
EtO2: End-tidal O2 concentration
FiO2-EtO2: Inspired-expired difference
O2 waveform Measurement range: 0 to 100 vol%
Accuracy: ± (1 vol% + 2% of the reading)
GE Datex-Ohmeda differential paramagnetic sensor
Adjustable low and high alarm limits for FiO2 and EtO2
Nitrous Oxide (N2O) Measurement range: 0 to 100 vol%
| Accuracy: | ± | (2 | vol% | + | 2% | of | the | reading) |
| (0%<N2O<85%) |
Anaesthetic Agent (AA) Isoflurane and Enflurane
| Measurement | range: | 0 | to | 6 | vol% | |||
| Accuracy: | ± | (0.15 | vol% | + | 5% | of | the | reading) |
Sevoflurane
| Measurement | range: | 0 | to | 8 | vol% | |||
| Accuracy: | ± | (0.15 | vol% | + | 5% | of | the | reading) |
Aladin2 Cassettes
6 7
CARESCAPE Respiratory Modules (continued)
Sensor specifications
D-lite Pedi-lite and D-lite(+) and Pedi-lite(+)
Dead space: 9.5 mL 2.5 mL
| Resistance: | at | 30 | L/min | at | 10 | L/min |
| 0.5 | cmH2O | 1.0 | cmH2O |
Et Control**
| Using | CARESCAPE | respiratory | module | E-sCAiOE | or | ||||
| E-sCAiOVE, | Et | Control | allows | you | to | set | the | desired | patient |
| End | Tidal | Oxygen | and | End | Tidal | Agent | concentrations. | ||
| The | Aisys | CS2 | then | automatically | adjusts | the | fresh | gas | |
| concentrations | to | quickly | and | efficiently | achieve | and | |||
| maintain | these | End | Tidal | concentrations. |
Isoflurane: OFF, Purge, 0.2 - 2.5%, resolution 0.1%
Sevoflurane: OFF, Purge, 0.2 - 4.0%, resolution 0.1%
Desflurane: OFF, Purge, 1.0 - 12.0%, resolution 0.2%
| When | OFF | is | selected, | no | additional | agent | is | added | to | the |
| system | and | flows | are | controlled | only | based | on | End | Tidal | |
| Oxygen | concentration. |
| When | Purge | is | selected, | agent | is | driven | out | of | the | system |
| as | fast | as | possible | by | elevating | fresh | gas | flows. |
| O2 | concentration | |||
| range: | 25 | to | 80%, | Max |
| When | Max | is | selected, | the | Aisys | CS2 | will | control | the | End | Tidal |
| Oxygen | concentration | as | high | as | efficiently | possible. |
| Flow | range: | 0.5 | to | 10 | L/min, | Minimum | flow | |
| can | be | controlled | by | a | user | setting, | ||
| 0.5 | to | 6 | L/min |
Compact Airway Modules
| M-CAiO, | M-CAiOV, | M-CAiOVX | module | software | version | 3.2 |
| or | higher; | E-CAiO, | E-CAiOV, | E-CAiOVX |
Size (W x D x H): 7.5 x 22.8 x 11.2 cm/3.0 x 9.0 x 4.4 in
Weight: 1.6 kg/3.5 lb
Sampling rate: 200 ±20 mL/min
| Automatic | compensation | for | atmospheric | pressure | variation | |||
| (500 | to | 800 | mmHg) | and | CO2/N2O | and | CO2/O2 | collision |
| broadening | effect. |
Note: For more details on Compact Airway Modules, reference the User Manual
Compact airway module gas exchange*
| (available | with | Mgas | or | Egas | Compact | Airway | Module |
| in | GE | Datex-Ohmeda | Anaesthesia | Monitor | module | bay) |
VO2: Oxygen consumption
VCO2: Carbon dioxide production
Measurement range: 50 to 1000 mL/min
| Respiration | rate | range: | 4 | to | 35 | bpm | (adults) |
| 4 | to | 50 | bpm | (pediatric) |
Compact airway module accuracy
FiO2 < 65%: ±10% or 10 mL/min
65% < FiO2 < 85%: ±15% or 15 mL/min
| Detection | through | D-lite | flow | sensor | or | Pedi-lite | flow | and | gas |
| sampling | sensor | (see | the | measurement | ranges | and | sensor | ||
| specifications | above). |
Electrical Specifications Current leakage
100/120 V: < 300µA
220/240 V: < 500µA
Power
| Power | input: | 100-120 | Vac, | 50/60 | Hz |
| 220-240 | Vac, | 50/60 | Hz |
| Power | cord: | Length: | 5 | m/16.4 | ft | |||||
| 10A | @ | 250 | Vac | or | 15A | @ | 125 | Vac |
Battery backup
| Backup | power: | Demonstrated | battery | time | under | |
| typical | operating | conditions | is | |||
| 90+ | minutes | when | anaesthesia | |||
| machine | is | fully | charged. | Battery | ||
| time | under | extreme | conditions | |||
| is | 30 | minutes | with | monitor. |
Battery type: Internal rechargeable sealed lead acid
Inlet/outlet modules
100-120 V System circuit breakers: 15A
| Outlets: | 4 | outlets | on | back, | 3-2A, |
| 1-3A | individual | breakers, | |||
| isolation | transformer |
220-240 V System circuit breakers: 8A
| Outlets | (optional): | 4 | outlets | on | back, | 3-1A, |
| 1-2A | individual | breakers, | ||||
| isolation | transformer |
Pneumatic Specifications Auxiliary common gas outlet (optional)
Connector: ISO 22 mm OD and 15 mm ID
Gas supply
| Pipeline | input | range: | 280 | kPa | to | 600 | kPa/ |
| 41 | psig | to | 87 | psig |
| Pipeline | connections: | DISS-male, | DISS-female, | |||
| DIN | 13252, | AS4059, | BSPP | 3/8, | ||
| S90-116, | or | NIST |
| All | fittings | available | for | O2, | N2O, |
| and | Air, | and | contain | pipeline | |
| filter | and | check | valve |
| Cylinder | input: | Pin | indexed | in | accordance | ||
| with | CGA-V-1 | or | DIN | (nut | and | ||
| gland); | contains | input | filter | and | |||
| check | valve |
Note: Maximum 3 cylinders
| Primary | regulator | diaphragm | minimum | |
| burst | pressure: | 2758 | kPa/400 | psig |
| Primary | regulator | ||||||
| nominal | output: | ≤ | 345 | kPa/50 | psig | Pin | indexed |
| cylinder | and | ≤ | 414 | kPa/60 | psig | ||
| DIN | cylinder | connections |
O2 controls
| Method: | N2O | shut | off | with | loss | of |
| O2 | pressure |
Supply failure alarm: Range: < 252 kPa/37 psig
| Sounds | at | maximum | volume |
| every | 10 | seconds |
O2 flush: Range: > 25 L/min
Alternate O2 (safety flow)
| Range: | 500 | mL/min | minimum | |
| to | 10 | L/min |
Indicator: Flow tube
Indicator accuracy: ±5% full scale
Fresh gas
| Flow | range: | 0 | and | 200 | mL/min | to | 15 | L/min |
| (minimal | flow | capable) |
| Total | flow | accuracy: | ±10% | or | ±40 | mL/min | of | setting |
| (whichever | is | greater) |
| O2 | flow | accuracy: | ±5% | or | ±20 | mL/min | of | setting |
| (whichever | is | greater) |
| Balance | gas | ||||||
| flow | accuracy: | ±5% | or | ±20 | mL/min | of | setting |
| (whichever | is | greater) | Air/N2O |
* Measurement not valid with O2 and N2O mixtures
| O2 | concentration | range: | 21% | to | 100% |
| (when | Air | is | available) |
| O2 | concentration | |||||||
| accuracy: | ±5% | V/V | for | flows | < | 1 | L/min* | |
| ±2.5% | setting | for | flows | > | 1 | L/min |
| Electronic | mixer | |||||||
| response | time: | 500 | msec | (10% | to | 90% | flow | step) |
| Compensation: | Temperature | and | atmospheric | ||
| pressure | compensated | to | |||
| standard | conditions | of | 20°C | ||
| and | 101.3 | kPa |
Hypoxic guard: Electronic
Materials
| All | materials | in | contact | with | patient | breathing | |
| gases | are | not | made | with | natural | rubber | latex |
Environmental Specifications System operation
Temperature: 10° to 35°C/50° to 95°F
| Humidity: | 15 | to | 95% | relative | humidity |
| (non-condensing) |
| Altitude: | -440 | to | 3000 | m/ | |
| 537 | to | 800 | mmHg |
System storage
Temperature: -25° to 60°C/-13° to 140°F
Humidity: 15 to 95%
| Altitude: | -440 | to | 4880 | m/ | |
| 425 | to | 800 | mmHg |
| Oxygen | cell | storage: | -15° | to | 50°C/5° | to | 122°F |
| 10 | to | 95% | relative | humidity | |||
| 500 | to | 800 | mmHg |
Electromagnetic compatibility
| Immunity: | Complies | with | all | requirements | |
| of | EN | 60601-1-2 |
| Emissions: | CISPR | 11 | ||
| Group | 1 | |||
| Class | A |
| Approvals: | AAMI | ES60601-1, | CSA | C22.2 | #601.1, |
| EN/IEC | 60601-1, | ISO | 80601-2-13 |
| * | The | stated | concentration | accuracy | may | not | be | met | for | |
| total | flows | between | 200 | and | 400 | mL/min. | At | least | 21% | |
| O2 | will | be | maintained. |
** Aisys CS2 and Et Control are not available for sale in the United States. Not cleared or approved by the FDA. Not available in all markets.
Breathing Circuit Specifications Operational modes
| Breathing | circuit | is | circle | mode; | SCGO | option | converts |
| to | open | circuit | mode |
Carbon dioxide absorbent canister
Absorbent capacity: 800 g
Integrated expiratory limb water reservoír
Ports and connectors
Exhalation: 22 mm OD ISO/15 mm ID taper
Inhalation: 22 mm OD ISO/15 mm ID taper
Bag port: 22 mm OD/22 mm ID (Australia)
Bag-to-Ventilator switch
Type: Bi-stable
| Control: | Controls | ventilator | and | direction | |||
| of | breathing | gas | within | the | circuit |
Integrated Adjustable Pressure Limiting (APL) valve
Range: 0.5 to 70 cmH2O
| Range | of | rotation: | 0.5 | to | 30 | cmH2O | (0 | to | 230°) |
| 30 | to | 70 | cmH2O | (230 | to | 330°) |
Materials
| All | materials | in | contact | with | exhaled | patient | gases | are |
| autoclavable, | except | disposable | flow | sensors, | O2 | cell, | and | |
| Respiratory | Modules. | (Autoclavable | flow | sensors | optional) |
| All | materials | in | contact | with | patient | gas | are | not | made | with |
| natural | rubber | latex. |
Breathing circuit parameters
Compliance: Bag mode: 1.82 mL/cmH2O
Expiratory resistance: Pexp Pexp Bag Mode Vent Mode Flow rate Pressure drop Pressure drop
5 L/min 0.46 cmH2O 0.46 cmH2O
30 L/min 1.47 cmH2O 1.55 cmH2O
60 L/min 3.80 cmH2O 4.09 cmH2O Note: Values include patient circuit tubing and wye piece (0.3 cmH2O at 60 L/min)
Anaesthetic gas scavenging
AGSS Type Hospital extract Machine system required connection
| High | vacuum, | High | vacuum | DISS | evac | ||||
| low | flow | with | 36 | L/min | @ | 12 | in | Hg | |
| indicator: | (305 | mmHg) |
| High | vacuum, | High | vacuum | DISS | evac | ||
| variable | flow | with | 30 | L/min | extract | ||
| bag | indicator: | flow | @ | 12 | in | Hg | |
| (305 | mmHg) |
| Passive: | Passive | or | external | 30 | mm/1.2 | in |
| active | system | M | ISO | taper | ||
| with | air | break |
© 2013 General Electric Company – All rights reserved.
| General | Electric | Company | reserves | the | right | to | make | |
| changes | in | specifications | and | features | shown | herein, | ||
| or | discontinue | the | product | described | at | any | time | without |
| notice | or | obligation. | Contact | your | GE | Representative | ||
| for | the | most | current | information. |
| GE | and | GE | Monogram | are | trademarks |
| of | General | Electric | Company. |
| Aisys, | Carestation, | ABS, | PSVPro, | Patient | Spirometry, | |
| CARESCAPE, | INview, | Aladin2, | D-lite | and | Pedi-lite | are |
| trademarks | of | Datex-Ohmeda, | Inc. |
| Datex-Ohmeda, | Inc., | a | General | Electric | company, |
| doing | business | as | GE | Healthcare. |
| NOT | COMMERCIALLY | AVAILABLE | IN | ALL | MARKETS. | ||
| NOT | CLEARED | OR | APPROVED | BY | THE | U.S. | FDA. |
DOC1261976 rev3 09/13
| GE | Healthcare | |
| P.O. | Box | 7550 |
| Madison, | WI | 53707-7550 |
USA
www.gehealthcare.com
File details come from the government source that posted it. Updated .