This document is an Attachment D to a solicitation, containing the NIST Handbook 105-3 Volumetric Field Standard Specification Checklist. It outlines the evaluation criteria and requirements for volumetric field standards used in weights and measures applications. The checklist covers identification markings, construction materials, shape and design, scale plate requirements, and other technical specifications. It also addresses non-retroactive requirements for sight flow indicators, bleed lines, and vapor recovery systems. The document provides guidelines for evaluating the suitability of volumetric field standards for weights and measures usage.
The checklist covers 40 specific evaluation criteria across categories such as identification, construction, scale plate, support, drainage, and workmanship. Each requirement is referenced to the applicable section in NIST Handbook 105-3, and the evaluator must indicate whether the standard meets the requirement or is rejected, or note if the requirement is not applicable. Comments can be provided to explain any rejections or special application exemptions.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
NIST Handbook 105-3 35
Appendix 1. Optional HB 105-3 Volumetric Field Standard Specifications Evaluation Form
Submitted by: Date:
Test Number:
Metrologist:
Model Number:
Serial Number:
Seals intact: Yes/No Description:
Instructions: Evaluate each item. Check either the approved or rejected box. If not applicable or allowed for a special application, draw a line through both boxes. Explain all rejections and special application exemptions in the comments section. File with calibration documentation.
Requirement HB 105-3 Ref. Evaluation Tips Appd Rej
Identification 4.3 Check individual requirements for overall approval
1 Identification placement 4.3.1 Permanently marked, weather & liquid resistant, contrasting color to background?
2 Nominal capacity 4.3.2.1 Nominal capacity as listed?
3 Manufacturer information 4.3.2.2 Name and address?
4 Model number 4.3.2.3 5 Serial number and year 4.3.2.4 Year may be combined in serial number if clear.
Material ID & cubical coefficient of expansion
4.3.2.5 Material identified?
CCE marked on standard?
7 Delivery and/or drain time 4.3.2.6 Marked?
8 Reference temperature PO or Cal Cert Reference?
9 Limitations on use 4.3.3 Note any limitations on use:
10 Construction material 4.4 Stainless steel or low-carbon steel
11 Body and shape 4.5.1, 4.5.2 Circular cross section in horizontal plane?
No fillers, adjusting plugs or cavities?
Shape ensures emptying, drainage; no air trapped?
12 Top and bottom cones 4.5.3 Adequate pitch, anti-vortex fin, cone trough drain?
Uniform, reinforcement roll or band, attachment?
Neck (top)
Neck & fill pipe cross sectional area? 13 Neck (bottom)
4.5.4 Drain valve, accessibility?
Glass, clear, unbroken, mounting, leak free?
14 Sight glass 4.5.5
Inside diameter and wall thickness?
Location: secant, less than 6 mm from tube?
Mounting: secure, corrosion resistant, same metal?
Adjustment: stable, < 25 % smallest scale division?
15 Scale plate construction 4.5.6
Adequate provisions for sealing?
Permanent and legible?
Nominal volume & proper unit abbreviations?
�To Contain� or �To Deliver� marked?
Nominal and zero lines?
Graduation spacing ( 1.6 mm) & width?
16 Scale plate markings 4.5.7
Bottom scale plate � zero line?
Primary _______ Secondary _______ Scale unit?
17 Scale plate units 4.5.8
Lack of confusion?
RFQ Attachment D
NIST Handbook 105-3 36
Requirement HB 105-3 Ref. Evaluation Tips Appd Rej
Span? __________ % of nominal? _______ 18 Scale plate graduations 4.5.9
Acceptable for W&M use (nominal near center)?
19 Support 4.5.10 Reinforced at leg attachment, reinforcing bands?
Number of wells:
20 Thermometer wells 4.5.11
Placement of wells, provision for probe calibration?
Discharge piping at least 5° slope?
21 Drain, piping, and valves 4.5.12 Fast acting full port valve (non-retroactive)?
22 Sight flow indicator (non-retroactive)
4.5.13 Clear, no cracks, near drain valve?
Minimization of vaporization or foaming
4.5.14 Description:
24 Fill pipe 4.5.15 Required if larger than 20 L (5 gal) used for top fill.
Appears to be within 10 cm � 15 cm (4 in � 6 in) of bottom cone?
25 Bottom loading 4.5.16 Placement, leak free, diverter plate with vent?
26 Air entrapment 4.5.17 Note any possible sources of air entrapment.
27 Bleed line (non-retroactive)
4.5.18 Equipped with a bleed line?
Leak free, unrestricted, protected?
28 Bleed valves not allowed (non-retroactive)
4.5.19 Equipped with a bleed valve? (Should be replaced with bleed line if major repair is required.)
29 Vapor recovery 4.5.20 Placement, pressure & vacuum relief valves?
30 Hardware 4.5.21 Handles, trunnions, & plumbing?
31 Leveling 4.5.22 Leveling jacks adequate, test measure hangs level?
32 Levels 4.5.23 Adequate, protected, agree with plumb neck?
Workmanship, finish, and appearance
4.6 Check individual requirements for overall approval.
33 Finished quality 4.6.1 Joints leak free and smooth, neck cover, pipe caps, pipe joints sealed?
34 Coatings 4.6.2 No interior rust?
Exterior paint in good condition?
35 Ladders and platforms 4.6.3 Sturdy, safety rails, non-slip surfaces?
Other requirements 4.7 No Approval Implied � Note any gross problems.
36 Electrical requirements 4.7.1-3 Adequate pump, ground lug, wiring intact.
37 Electronic overflow system 4.7.4 Note if present.
38 Braking system 4.7.5 Note if present.
39 Labeling and placarding 4.7.6 Note if present.
Test methods 7 Examine in preparation for calibration.
40 Cleanliness 7 No residue, clear sight glass(es), non-slippery.
Requirement HB 105-3 Ref. Comments