Attachment C- Technical Specifications.pdf
PDF 188 KB Posted
- Attached to
- Isotope Ratio Mass Spectrometer
- Solicitation number
- PR388179
- Issued by
-
Department of Energy
About this file
This document is a specification sheet for an Isotope Ratio Mass Spectrometer (IRMS) detailing its technical capabilities and physical characteristics. The instrument features a gas feed with dual inlet and supports sample options including SiF4, Xe, CO2, O2, N2O, Ar, and SF6, with 10^13 Ohm amplification and a dynamic range of 10^7. The mass spectrometer utilizes an isotope ratio multi-collector measurement type and requires a minimum acceleration voltage of 10 kV.
The machine has a compact footprint of 1.5 meters in depth and height, and 1.1 meters in width, with a delivery timeline of 3 to 6 months after purchase. Potential additional features and upgrades include analysis of exotic gases like GeF4, capability to pump sticky gases (SiF4, GeF4), compatibility with fluorinated gases, and power conditioning. The specification sheet notes that the footprint measurement does not include clearances for work or maintenance zones, which may be important for laboratory or research facility planning.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Isotope Ratio Mass Spectrometer
IRMS Specifications
Capabilities Feed (sample) form Gas feed, dual inlet Sample options SiF4 (SiF3+), Xe, CO2, O2, N2O, Ar, SF6 (SF5+)
Amplification 10^13 Ohm amplification
Dynamic range 10^7
Measurement type Isotope ratio multi-collector
Acceleration Voltage Minimum of 10 kV
Machine Footprint Depth 1.5 meter
Width 1.1 meter
Height 1.5 meter
Does not include any clearances for work or maintenance zones.
Delivery in 3 to 6 months of purchase
Additional Features Upgrades
• Analysis of additional, exotic gases, ex. GeF4 may be required
• Pumping of sticky gases (SiF4, GeF4)
• Compatibility with fluorinated gases.
• Power conditioning
| Isotope Ratio Mass Spectrometer IRMS Specifications |
| Capabilities |
| Machine Footprint |
| Additional Features Upgrades |
File details come from the government source that posted it. Updated .