Attachment C - Bill of Materials-PR459123.pdf
PDF 81 KB Posted
- Attached to
- 550-101A UL1000 Fab Mobile helium leak detector
- Solicitation number
- PR459123
- Issued by
-
Department of Energy
About this file
This is a pricing template attachment (Attachment C) to RFP459123 for the procurement of a mobile helium leak detector. The solicitation requires one unit of the Inficon model 550-101A UL1000 Fab Mobile helium leak detector, part number 550-101A. Vendors are instructed to complete the template by providing their company name, quote number, date, unit price, total price, lead time, and country of origin for the specified equipment.
The detector must meet specific performance specifications including operation at 100/115 V (±10%) at 50/60 Hz with the capability to detect helium and hydrogen. Key technical requirements include a minimum detectable leak rate for helium of less than 5E-12 mbar·l/s in vacuum mode and less than 5E-8 mbar·l/s in sniffing mode, maximum inlet pressure of 15 mbar, pumping speed in pump-down mode of 30 m³/h at 60 Hz, helium pumping speed in ULTRA mode of 2.5 l/s, inclusion of a test leak TL7, and a dry backing pump. The template includes a specifications checklist requiring vendors to confirm compliance with each requirement using a Yes/No response format.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Attachment C- RFP459123
Vendor Name Quote Number Date
Line Item Item Description Part Number Manufacturer Quantity Unit Price Total Price Lead Time Country of Origin 1 550-101A UL1000 Fab Mobile helium leak detector 550-101A Inficon 1
Total
Specifications Yes/No:
Mobile helium leak detector 100/115 V (± 10%) 50/60 Hz Detects: Helium, Hydrogen Min. detectable leak rate for helium [mbar·l/s]:
- vacuum mode: < 5 E-12
- sniffing mode: < 5 E-8 Max. inlet pressure: 15mbar Pumping speed in pump-down mode: 30 m³/h @ 60 Hz Helium pumping speed in ULTRA mode: 2.5 l/s Test leak TL7 Dry backing pump
Sheet1