Attachment B - Compliance_Checklist.xlsx
XLSX spreadsheet 19 KB Posted
- Attached to
- RFI- Aeromedical Certified Ventilator
- Solicitation number
- SPE2D1-26-RFI-0001
About this file
This is a Compliance Checklist for a Request for Information (RFI) solicitation for aeromedical certified portable and transport ventilators issued by the Defense Logistics Agency Troop Support Medical. Vendors must respond with compliance codes indicating whether their products currently comply (C), have planned compliance (P), do not comply (N), or the product is not offered (NA), along with supporting technical documentation and vendor comments.
The required transport ventilator must be FDA-cleared or approved (PMA, 510(k), or legally marketed under 21 CFR 807.97(a)(3)), comply with AAMI MCIP standards (ANSI/AAMI ID26), and hold Joint Airworthiness Certification including a USAF Safe-to-Fly letter and US Army Fleet Airworthiness Release and Aeromedical Certification Letter for use on all military aircraft platforms. The ventilator must serve all patient populations (infant, pediatric, adult) with specifications including tidal volume control of 50-2,000 mL, respiration rate of 0-60 breaths/min, FiO2 of 21-100%, PEEP/CPAP of 0-25 cm H₂O, and support for multiple operating modes including Assist/Control, SIMV, CPAP, and non-invasive ventilation. Physical requirements specify dimensions not exceeding 13"H x 9"W x 5"D and weight not exceeding 13 lbs (with 10 lbs or less preferred), operation on 110-220 VAC (50-60 Hz) with automatic switching and rechargeable internal battery providing 4-10 hours of operation, operating temperature range of 32-104°F, and a ruggedized, de-contaminable carrying case. The device must display comprehensive monitored parameters and alarms, include Night Vision Goggle-compatible display readable in low-light environments, possess an internal air compressor, and transition seamlessly between ambulances and fixed/rotary-wing aircraft without requiring ventilator changes. Vendors must provide initial and distance learning training, two sets of operating and service manuals (electronic and hard copy), one-year minimum manufacturer warranty (two years preferred), maintenance capability by military Biomedical Equipment Technicians using standard tools without voiding warranty, and repair parts availability through ECAT for direct customer ordering.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| RFI.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Sheet1
| Transport Ventilator Detailed Requirement Description | |||||
| Functional Area | Threshold (T) (Minimum Acceptable) | Objective | Please include in the vendor response the following compliance codes: C- Current offerred product fully complies; P- Not part of current product but planned in schedule release (Vendor shall indicate the time when the requirement will be available); N- Product does not comply: NA - Product not offered | Referenced Technical literature, and pages in technical literature validating compliance | Vendor Comments/feedback |
| Description: Mechanical ventilators provide temporary or permanent ventilatory support for patients who cannot breathe on their own or who require assistance maintaining adequate ventilation because of illness, trauma, congential defects, or the effects of drugs (e.g., anesthetics). A transport ventilator normally consists of a flexible breathing circuit, a control system, a gas supply, and some type of monitoring and alarm system. Portable transport ventilators are designed to take the place of manual ventilation or "bagging" during emergency or transport situations. Portable transport ventilators are well suited for both prehospital and in-hospital applications. Prehospital applications include use in ground and air ambulances and in emergency field hospitals. In healthcare facilities, portable ventilators are used for short term mechanical ventilation and to facilitate transport of ventilator-dependent patients to and from various areas of the facility (e.g., ICU, OR, Radiology). | |||||
| FDA Clearance | Device must be “approved for market” (PMA), “cleared to market” (FDA 510k approval); or “legally marketed” (as defined in the Code of Federal Regulations – 21 CFR807.97(a)(3) for Class I and II devices) and comply with current versions of the Association for the Advancement of Medical Instrumentation (AAMI) MCIP standards (ANSI/AAMI ID26) | T=O | |||
| Approved for Aircraft Use | Shall have a Joint Airworthiness Certification (JAC) consisting of a “Safe-to-Fly” letter from the USAF for fixed and rotary wing aircraft and a Fleet Airworthiness Release (AWR) and an Aeromedical Certification Letter (AWC) from the US Army.Proof of this Joint Aeromedical Certification shall be provided with proposal submittal. Shall be acceptable for use during all phases of flight on all military aircraft platforms (including small and large body, fixed and rotary wing) while operating on battery or 110-120VAC 60 Hz in the Aeromedical Evacuation environment. Airworthiness assessment will be conducted in accordance to the JECETS standard. | T=O | |||
| Shall be used with all ranges of patients (infant, pediatric, adult) and should include an option of closed-loop mechanical ventilation as a safety mechanism for the non-provider operator. | Adult, pediatric, infant | T=O | |||
| Controls | |||||
| Tidal volume, mL | 50 - 2,000 | T=O | |||
| Respiration rate, breaths/min | 0-60 | T=O | |||
| Trigger mechanism | Pressure | T=O | |||
| FiO2, % | 21-100 | T=O | |||
| Inspiratory flow rate, L/min | >100 | T=O | |||
| Inspiratory time, sec | 0 - 2% | T=O | |||
| Inspiratory pressure, cm H2O | 10-60, 80 with confirmation | T=O | |||
| IE ratio | 1:1 to 1:99 | T=O | |||
| PEEP/CPAP, cm H2O | 0-25 | T=O | |||
| Pressure support, cm H2O (measured in ref to PEEP) | 0-60 | T=O | |||
| Panel lock | No - Not necessary / YES is acceptable | T=O | |||
| Pressure ramp | YES | T=O | |||
| Accuracy shall be within +/- 10% of settings for all measured values and controls if not further specified | YES | T=O | |||
| Operating Modes | |||||
| Assist/control mode | YES | T=O | |||
| A/C volume breaths | YES | T=O | |||
| A/C pressure breaths | YES | T=O | |||
| SIMV mode | YES | T=O | |||
| SIMV volume breaths | YES | T=O | |||
| SIMV pressure breaths | YES | T=O | |||
| SIMV pressure support | YES | T=O | |||
| Spontaneous/CPAP mode | YES | T=O | |||
| CPAP pressure support | YES | T=O | |||
| Non-invasive ventilation | YES | T=O | |||
| Apnea-backup vent mode | YES | T=O | |||
| Monitored/ Displayed Parameters | |||||
| Peak inspiratory pressure (PIP) | YES | T=O | |||
| Mean airway pressure (MAP) | YES | T=O | |||
| PEEP pressure | YES | T=O | |||
| Tidal volume | YES | T=O | |||
| Minute volume | YES | T=O | |||
| FiO2 | YES | T=O | |||
| Respiratory rate | YES | T=O | |||
| Inspiratory time | YES | T=O | |||
| Expiratory Time | YES | T=O | |||
| IE ratio | YES | T=O | |||
| O2 in use | YES | T=O | |||
| SpO2 | YES | T=O | |||
| Battery % | YES | T=O | |||
| Patient Alarms | |||||
| Low/high FiO2 | YES | T=O | |||
| Low/high minute volume | YES | T=O | |||
| Low/ high pressure | YES | T=O | |||
| Loss of PEEP | YES | T=O | |||
| High PEEP | YES | T=O | |||
| Apnea | YES | T=O | |||
| Continuous high pressure/occlusion | YES | T=O | |||
| Inverse IE | YES | T=O | |||
| High respiratory rate | YES | T=O | |||
| Breathing circuit disconnect | YES | T=O | |||
| SpO2 | YES | T=O | |||
| Equipment Alarms | |||||
| Gas-supply failure | YES | T=O | |||
| Power failure | YES | T=O | |||
| Vent inoperative | YES | T=O | |||
| Low battery | YES | T=O | |||
| Self-diagnostic | YES | T=O | |||
| On-Board Air Compressor | |||||
| Internal compressor | YES | T=O | |||
| Power Requirements, VAC | |||||
| Shall be a portable device that operates on 110VAC, 60 Hz and 220VAC, 50 Hz and incorporated rechargeable battery. Automatic switching between these power sources required | 110v - 220 VAC; 50-60 Hz | T=O | |||
| Rechargeable internal battery | 4 hr | 10 hr | |||
| Physical Features | |||||
| Shall have Shipping and Storage temperature ranges of not less than -4°F to 140°F (-20°C to 60°C) and an Operating range of not less than 32°F to 104°F (0°C to 40°C) | YES | T=O | |||
| Shall measure no greater than 13 inches (H), 9 inches (W), and 5 inches (D) (33x23x13 cm) | YES | T=O | |||
| Shall weigh no greater than 13lbs (5.8 Kg) | YES | 10lbs (4.5 Kg) or less | |||
| Portable Ventilator display shall be easily readable in low-light environments and Night Vision Goggle compatible and provide Audible and visual alarms in multiple colors according to level of severity allowing continuous visual alarm monitoring by crew during all phases of flight. Consider limitations due to aircraft ambient noise degrading effectiveness of audio alarms. | YES | T=O | |||
| Shall be capable of blending oxygen concentration | YES | T=O | |||
| Shall include a ruggedized, carrying case that is de-contaminable and designed to prevent damage to the device during operation and transport. | YES | T=O | |||
| Patients are transported by ambulance, Am Bus, rotary wing aircraft and fixed wing aircraft. Ventilator should be applied/ attached to patient at hospital and transition through levels of care without requirement of movement to another ventilator | YES | T=O | |||
| Training | |||||
| Vendor must provide initial training | YES | T=O | |||
| Vendor must provide alternative training methods utilizing the latest distance learning technology | YES | T=O | |||
| Vendor will provide training materials for users on new functionality and enhancements | YES | T=O | |||
| Maintenance | |||||
| Shall contain two (2) sets of operating and service manuals | Electronic and hard copy | T=O | |||
| Must have manufacture warranty | 1 year | 2 years | |||
| Device must be able to be cleaned and disinfected with a solution in accordance with infection control guidance Per AFI 44-108 Infection Control Program | Yes | T=O | |||
| Maintainability at military location with Vendor training to military personnel. | The Device shall be maintainable by military Biomedical Equipment Technicians (BMET) using standard tools (as provided in standard BMET kit; no special tools required) without voiding warranty and shall be capable of being cleaned by commonly accepted methods. Repair parts and/or telephone consultation before and after warranty period will be provided to DoD BMETs without need to attend manufacture specific training or incurring additional telephone consultation costs. | T=O | |||
| The vendor shall include in their proposal all repair parts for the ventilator, and agree to allow the Government to add these parts to the resultant contract(s) and make the parts available on ECAT for direct customer ordering. | Yes | T=O | |||
| Vendor point of contact: | |||||
| Name: | |||||
| Job title: | |||||
| Phone: | |||||
| Email address: |