Attachment A - Technical Specifications for Emittance Scanner - Dated March 4 2026.pdf

PDF 92 KB Posted

Attached to
Request for Proposal RFP-PR437816 Emittance Scanner
Solicitation number
PR437816
Issued by

About this file

This is a technical specifications document for an Emittance Scanner system dated March 4, 2026. The document details the performance and operational requirements for both the probe component and the control system.

The Emittance Scanner Probe specifications include a maximum travel distance of 250mm with a slit-to-flange distance ranging from 95mm to 345mm, customizable flange design, and a maximum beam diameter of 100mm. Key performance metrics include Y resolution of 100μm with 1μm minimum step, Y' maximum of ±126 mRad for kinetic energy less than 65 keV, Y' resolution of 1.2mRad, and sweep voltage of ±1250 V. The slit is factory-configurable from 25 to 250μm width with a fixed length of 104μm. The system supports a maximum beam power of 10kW with maximum beam intensity of 100 W/cm² at full power. The probe requires cooling water at 23 liters per minute at 450 kPa and incorporates a molybdenum/copper cooling plate and slits (with alternative material options available for discussion). The Control System operates on a 16U 19-inch industrial PC rack with Windows and LabView platforms, capable of completing a 30-by-30 step scan in less than one minute. The current meter ranges from 2nanoamperes to 20milliamperes full scale with a typical noise floor of 5 picoamperes. System power requirements are 115 or 220 VAC. The system generates 2D and 3D phase-scale intensity distributions, computer emittance calculations, RMS ellipses, and Twiss parameters, with output available in CSV and PDF formats.

View the file

Other files

Show all 13

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Technical Specifications

Emittance Scanner

Dated March 4, 2026

Emittance Scanner Probe Specifications:

• Max Travel: 250mm

• Distance – Slit to Flange: 95mm to 345mm

• Flange: Ability to customize to customer specifications

• Y Resolution: 100μm

• Y Step Min: 1μm

• Y’ Max: ±126 mRad(for KE <65 keV)

• Y’ Resolution: 1.2mRad

• Y’ Step Min: 16 bit resolution over +/-Y’ Max

• Sweep Voltage: ±1250 V

• Slit Width: Factory configurable 25 to 250μm

• Slit Length: 104μm

• Beam Ømax: 100mm

• Bias Voltage: -100 V

• Electrode Gap: 6.0mm

• Electrode Length: 76mm

• Cooling Water: 23 LPM @ 450 kPa (65 PSI)

• Max Beam Power3: 10kW

• Max Beam Intensity3: 100 W/cm2@ 10 kW Beam Power

• Cooling Plate: Molybdenum/Copper ideally, may discuss options

• Slits: Molybdenum/Copper ideally, may discuss options

Emittance Scanner Control System Specifications:

• Instrument Rack: 16U 19” rack, w/ keyboard, power bar

• Platform: Industrial PC / Windows™ / LabView™

• Typical Scan Time: <1 minute for 30 step Y by 30 step Y’ scan

• Current Meter:

o 2nA to 20mA (full scale) o 5 pA (typical noise floor)

• System Power: 115 or 220 VAC

• Data Plots & Analysis: 2D & 3D phase-scale intensity distributions, computer emittance, RMS ellipses, and Twiss Parameters

• File Report: CSV, PDF

File details come from the government source that posted it. Updated .