Attachment A is a Salient Characteristics document detailing specifications for Transport Patient Monitors for the Department of Veterans Affairs (VA). The document specifies a Philips IntelliVue X3 (part number 867030) monitor that must be constructed of durable hard plastic with antimicrobial Corning Gorilla Glass, capable of withstanding hospital disinfectants. The device requires a 5-hour battery life, wireless connectivity to existing 802.11 networks or IntelliVue Smart-Hopping systems, and must interface with Philips PIIC iX Central Station.
The monitor must be compact (no more than 15.5 cms) with a high-resolution color LCD capacitive multi-touch screen, and designed for critical care patient monitoring during transport and bedside care. Specific monitoring capabilities include ECG, multiple SpO2 technologies, respiration, blood pressure, invasive pressures (CO2 and additional SpO2), cardiac output, and various CO2 measurement methods. The device must be ergonomic, survive drops up to 1 meter, have customizable screens, and be intended for hospital operations monitoring critical care patients.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| ATTACHMENT B - Offeror Reps and Certs - Return with Offer.docx | DOCX document | |
| RFQ Description - Transport Patient Monitors.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
ATTACHMENT‐A
SALIENT CHARACTERISTICS
a. Common Nomenclature (commercial description): TRANSPORT PATIENT MONITORS b.
Kind of material (i.e., type, grade, alternatives, etc.):
Must be made out of durable hard plastic and antimicrobial Corning Gorilla Glass. Built from chemical resistant materials to withstand disinfectants.
c. Electrical Data (i.e., 110 v, 60 Hz, 3 Ph, etc.):
Battery life must be 5 hours or more (5 hours of run time on a single charge). Must be able to connect wirelessly (via existing 802.11 network or IntelliVue Smart‐Hopping system, to IntelliVue central stations as well as host‐system devices and remote display screens.
d. Dimensions, size, capacity:
Must be no more than 15.5 cms (6.1 inches in size) with an active matrix color LCD display with capacitive multi‐touch screen (resolution of at least 1024 x 480).
e. Principles of operation:
Must be able to monitor ECG, SpO2 (Philips FAST‐SpO2, Nellcor OxiMax SpO2, Masimo Rainbow SET SpO2), respiration, NBP, 2 invasive pressures – CO2 and second SpO2, cardiac output (right‐heart and PiCCO), Covidien Microstream CO2, Respironics mainstream Co2 and/or LoFlo Sidestream CO2.
f. Restrictive environmental conditions: For hospital settings only (must survive a drop of up to 1 meter) and must be ergonomic to handle.
g. Intended use:
Must be able to monitor critical care patients during transport and at the bedside. Reduces number of steps to prepare patients for transport. Each screen must be customizable to user needs h.
Equipment with which the item is used with (N/A if not applicable):
Must interface with existing Philips PIIC iX Central Station
i. End item application: Hospital operations to monitor critical care patients
j. Original equipment manufacturer part number: 867030 (IntelliVue X3)