Attachment_A_-_P24-0882_Scope_of_Work.pdf

PDF 86 KB Posted

Attached to
P24-0882 Informal RFQ - Automated Solid Phase Extraction System
Solicitation number
P24-0882
Issued by
Sacramento County, San Francisco City, California

About this file

This is a Scope of Work document (P24-0882) from the Center for Analytical Chemistry (CAC) for automated SPE (Solid Phase Extraction) systems for processing water samples by the Environmental Analysis section. The contractor must deliver the automated SPE systems between 8:00 AM and 4:30 PM, include proper documentation on packing slips, and provide training for system operation. The technical requirements specify systems that weigh no more than 25 kg with 60cm³ dimensions, capable of running 8 samples simultaneously with parallel extraction steps, using positive pressure and continuous flow pumps. The systems must include a built-in touchscreen, perform automatic bottle rinsing, be compatible with multiple SPE cartridge sizes (1-20mL), 30mm Mini-disks, and 47mm SPE disks, feature adjustable bottle adapters, and collect 2-6 fractions per channel.

The document outlines CDFA's responsibilities including receiving and installing the instruments and providing appropriate laboratory space with proper benchtop and electrical requirements. No specific pricing terms, set-asides, or funding information is included in this scope of work. The contract manager is identified as Michael S. Gross, located at 3292 Meadowview Road, Sacramento, CA 95832.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Scope of Work

P24-0882 Automated SPE Systems

CENTER FOR ANALYTICAL CHEMISTRY (CAC)

1. INTRODUCTION

The Contractor shall provide automated SPE systems for the processing of water samples by the Environmental Analysis section of CAC.

2 . CONTRACTOR REQUIREMENTS

• Delivery of systems.

• Deliver items related to this purchase order (PO) between 8:00 AM and 4:30 PM

• Include on the Packing Slip: CAC / PO number to ensure proper delivery

• Ensure proper training for the operation of the system.

3 . INSTRUMENT REQUIREMENTS (Individual System) An automated Solid Phase Extraction System that must include the following:

• Weighs no more than 25 kg and has dimensions of 60cm^3.

• Runs up to 8 samples simultaneously, with all extraction steps performed in parallel.

• Perform all extraction steps, including sample loading, using positive pressure.

• Sample loading is performed with continuous flow pumps.

• Include a built-in touchscreen and does not require an external PC/laptop.

• Performs automatic sample bottle rinsing.

• Be compatible with Solid Phase Extraction cartridges (in sizes 1, 3, 6, 12, and

20mL), 30 mm Mini-disks, and 47 mm Solid Phase Extraction disks.

• Sample bottle adapters are adjustable to accommodate a wide range of bottle dimensions.

• Collect 2 fractions per channel and it is customizable up to 6 fractions per channel.

4. CDFA RESPONSIBILITIES

• CDFA will receive instruments and provide space to house instruments in the laboratory with proper benchtop and electrical requirements met

• CDFA will install instruments