This document is a specifications table for two bench hoods for a laboratory environment, detailing precise technical requirements. The two bench hoods (Hood 1 and Hood 2) are both located in R235, with identical interior dimensions of 48"x30" and an exterior length of 72". Both hoods require 2 circuits with 4 outlets each at 120V 20A, and are specified for Radio Chem processes. The material is radio-stainless steel with a 64 RMS finish, featuring a combination sash type. Each hood has a 3" diameter pass-through port, with Hood 1 having a left-side pass-through and Hood 2 a right-side pass-through, designed for handling acid and flame-related laboratory work. The delivery deadline for these hoods is set for 6/1/26, indicating a specific procurement timeline for specialized laboratory equipment.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment B_Reps-and-Certs-Abbreviated - 2025.12.22.pdf | ||
| RFP PR438202.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Hood location and name Process Material requirement Sash Type End Panel
Ports/notes Cabinets H W L (exterior) Gas Vacuum Electrical
Bench Hood 1 R235 48 30 72 3 1x 2 circuits 4 outlets each 120V 20 A Radio Chem
Radio-stainless steel 64 rms finish Combination
Left pass through 3" dia Acid,Flame
Bench Hood 2 R235 48 30 72 3 1x 2 circuits 4 outlets each 120V 20 A Radio Chem
Radio-stainless steel 64 rms finish Combination
Right pass through 3" dia Acid,Flame
Hoods are needed by 6/1/26.