Attachment 5 - Clinitek.pdf
PDF 154 KB Posted
- Attached to
- Preventive Maintenance & Calibration for Various Laboratory Analyzers at U. S. Naval Hospital Okinawa
- Solicitation number
- HT941025Q2035
- Issued by
-
Defense Health Agency
About this file
This is a maintenance checklist for the Clinitek Advantus urinalysis analyzer, which appears to be associated with a federal contract opportunity for preventive maintenance and calibration services at U.S. Naval Hospital Okinawa. The checklist outlines detailed inspection and testing procedures that must be performed on the analyzer.
The checklist includes specific testing criteria and acceptable ranges for various components, including: strip centering (between 4-40), strip walking (must be A or B), touch screen accuracy (within 4mm radius), strip detector sensitivity (10-254 range), display pixel verification, lead head testing (CAL values 600-1820, DK values 1-400), and engagement testing (500 cycles of continuous operation). The document also covers sterilization procedures, accessory verification, initial condition checks, parts replacement confirmation, and final preparation steps for shipping the equipment after service is completed.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
I i
Clinitek Advantus Maintenance Checklist-
Product Name: Uririalysis Analyzerj "Clinitek Advantus" Serial Numb.er: j Requested Content; Regular lnspet:tion l '
Inspection content i
1. Sterilization
2. Confirmation .of Accessories, eti:.
3. Confirmation of initial Crinditio�
3.1 Jnstrumen� Setting/Conteht
3.2 Confirmatibn of i:he initial!condition of Lead Head Test
l ;
4: Confirmation of replaced parts :
5. Confirmation of 1:he operation after repair
5.1 Confirmati�n of Strip Centiering
* Judgment criteria: Betw�en 4 a:nd 40
5.2 Confirmation of)Strip Walking ;
* Judgment criteria: The jlldgment result shall be A or 8.
5.3 Confirmation ofjTouch Screen i
* Judgment crit�ria:·"+" mark shall beldisplayed in a 4mm radius from Touch· point.
i . ! .
5.4 Confirmati6n of the sensitivity of Strip Detector
* JudgmJnt criteria: Thelsensitfvity of each detectOr shall be the range of 10 to 254.' ' ' . .
* Confirm to sound;the answ_er to?e (ble�p sound) at any posi tion when moving step by step from the left side to the right side Of Reagent loading area.
5.5 Confirmation of Display
* VisuallVi coryfirm that e�ch pixel of display does i-iot have abnormality.
5.6 ConfirmatiOn of Lead Head
* Judgr:n�nt criteria: CAii and.CAl values shall be the range of 600 to 1820.
i
DKl and 0:K2 values shall be the range of 1 to 400.
RH2 t;:AL Edge Corr. Shall be-1, 0 or +l.
5.6 Confirmatitjn of Engaging
Result:
Result:·
Result:
Result:
Result:
Result:
Result:
Result:
Result:
Result:
Result:
Result:
Result:
Result:
* Judgmeint criteria: No �bnorn:iality shall be occurred after running the c�ntinuous operating of 500 cydes with Regent feeding motion. (Approx. for·l hour) Result:
6. Preparation of.Sh)pment
6.1 ConfirmatiOn of lnstrumerit settilig
6.2 ConfirmatiJn of Appearanbe
6.3 Confirmaticin of Accessori�s
Result:
Result:
Result:
Line Item # 7 and # 8