This document contains the salient characteristics for an X-ray inspection system required by the Department of the Army Materiel Command Contracting Command at the Detroit Arsenal. The key requirements include a tunnel size of 59 inches x 70.9 inches, a conveyor belt capable of pulling at least 2,646 lbs and moving bidirectionally, and the ability to display at least 10 previously scanned images. The system must also meet technical specifications such as a minimum steel penetration of 28 mm and a radiation dose of less than 0.1 mR per hour. This requirement will be solicited as a small business set-aside using Simplified Acquisition Procedures with a NAICS code of 334519 (Other Measuring and Controlling Device Manufacturing) and a related business size of 600 employees.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics for X-Ray Inspection System
General
· Tunnel Size: Must be able to scan objects up to 59 inches x 70.9 inches (W x H)
· Conveyer: Pulling weight must be no less than 1200 kg (2,646 lbs) and belt movement must be bidirectional (forward and reverse)
· Operating System: Must be able to connect to external monitors via HDMI cable. Must contain an image recall function that can show at least 10 previously scanned objects. Must include controls for belt movement with options to stop, move forward and backward. Comment by Worrell, Tracy E CIV USARMY ACC (USA): Do we want to address the operator control panel and ability to at a minimum move objects forward, backward and stop? Comment by Worrell, Tracy E CIV USARMY ACC (USA): The Autoclear specs call out a specific quantity for Autoimage recall, maybe we can use that as a minimum standard Autoimage recall must be 10 previously scanned images or more or something like that
· System Power: 220 VAC +/- 10%
Technical
· Steel Penetration: Must be at least 28 mm
· Wire Resolution: Must be at least 38 AWG
X-Ray Generator
· Duty Cycle: 100%
· Dose per Exam: Must be less than 0.1 mR per hour
· Radiation Leakage: Must be less than 0.5 mR per hour