Attachment 2 - Specifications - IMAGING TABLES.docx

DOCX document 19 KB Posted

Attached to
6525--Cheyenne VAMC Pain Program Imaging Table Federal contract opportunity
Solicitation number
36C25926Q0798
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 19

About this file

This is a specifications document for an Integrated Headrest Imaging Table required by the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 19 for the Cheyenne VAMC Pain Management Program (Solicitation Number 36C25926Q0798).

The table must be a powered 4-motion imaging table designed to support fluoroscopy-guided pain management and interventional procedures, featuring a 22" × 84" radiolucent carbon fiber top with 56" metal-free imaging space for C-arm accessibility. Required powered movements include electric height adjustment (26"-43" without pad), Trendelenburg tilt (±15°/±11°), 10-inch longitudinal travel, and ±15° lateral tilt, controlled via multifunction hand control and optional foot control for height adjustment. The table must accommodate a 500-pound patient weight capacity, weigh approximately 475 pounds, include 4" or 5" locking casters suitable for mobile use, and come with a 2" radiolucent comfort foam pad, crescent face rest pad, and two patient safety straps. All equipment must be FDA-listed as a Class II medical device, CE-marked, cETLus-approved with CB Certificate compliance, and manufactured in an ISO 13485-certified facility meeting IEC/EN 60601-1, IEC/EN 60601-1-2, IEC/EN 60601-2-32, and ANSI/AAMI ES60601-1 electrical safety standards. The vendor must provide a four-year warranty on frame, understructure, upholstery, and padding, plus three-year parts and one-year labor warranty for electrical and moving components.

View the file

Other files for this federal contract opportunity

Other files attached to 6525--Cheyenne VAMC Pain Program Imaging Table, newest first.
File Type Posted
36C25926Q0798.docx DOCX document
Attachment 1 Combined Synopsis Solicitation - 36C25926Q0798 - IMAGING TABLES.docx DOCX document
Attachment 3 Buy American Certificate 52.225-2 36C25926Q0798 - IMAGING TABLES.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

SPECIFICATIONS –

Integrated Headrest Imaging Table that supports fluoroscopy‑guided pain management and interventional procedures, providing stable, radiolucent, ergonomic positioning compatible with mobile C‑arm imaging systems. An Integrated Headrest model is required due to its expanded motion capabilities, radiolucent top, and clinical suitability.

Carbon Fiber Top & Imaging Area

•22" × 84" radiolucent carbon fiber top
•56" metal‑free imaging space for C‑arm accessibility
•Integrated headrest for prone positioning and facial support

Powered Movements (4‑Motion Table)

•Electric height adjustment: 26"–43" (without pad)
•Trendelenburg tilt: +15° / −11°
•Longitudinal travel: 10 inches
•Lateral tilt: ±15° 13

Controls

•Multi‑function hand control for tilt, travel, and lateral movements
•Foot control available for height adjustment
•Weight Capacity & Construction
•500 lb (227 kg) patient weight capacity
•Table weight approx. 475 lbs
•Includes 4" or 5" locking casters (recommended for mobile use)

Padding & Patient Support

•2" Radiolucent Comfort Foam® pad included
•Crescent face rest pad included with integrated headrest top
•Two patient safety straps included

Regulatory Compliance

•FDA listed medical device (Class II)
•CE marked; cETLus approved; CB Certificate compliant
•Meets IEC/EN 60601‑1, IEC/EN 60601‑1‑2, IEC/EN 60601‑2‑32, ANSI/AAMI ES60601‑1, and related electrical safety standards
•Manufactured in ISO 13485‑certified facility

Warranty 4‑year warranty for frame, understructure, upholstery, padding 3‑year parts and 1‑year labor warranty for electrical/moving parts

END SPECIFICATIONS DOCUMENT

File details come from the government source that posted it. Updated .