Attachment 1 Statement of Work.docx

DOCX document 85 KB Posted

Attached to
HEPA Filter Testing Federal contract opportunity
Solicitation number
12639526Q0281
Issued by
Department of Agriculture Animal and Plant Health Inspection Service Marketing and Regulatory Programs Business Services

About this file

This is a Statement of Work (SOW) for annual in-place HEPA filter, Magnehelic gauge, and biosafety/chemical fume hood certification services at the National Center for Applied Plant Protection (NCAPP) in Beltsville, Maryland. The contract covers a one-year base period with four optional one-year periods. The contractor must perform comprehensive testing and certification of 53 in-place HEPA filters located in 44 pieces of equipment, 30+ Magnehelic pressure gauges, 13 biosafety cabinets (including four horizontal/vertical laminar airflow units), and nine chemical fume hoods. Testing must comply with NCAPP Standard Operating Procedures (SOP 90-016-01), NSF/ANSI Standard 49, and ASME N510-1989 standards for HEPA filter bank testing.

The contractor must provide all labor, materials, certified test equipment, and challenging agents required for certification. The contractor personnel must have a minimum of five years of experience in testing, repairing, and certifying HEPA filters for federal and state agencies, hospitals, universities, or military facilities and must be NSF and IAFCA accredited technicians. All work must be conducted during normal business hours (6:30 a.m. to 4:30 p.m. Eastern Time) and must maintain cascading negative pressure gradients within the containment building. Contractor personnel require Government escort while in the facility. The SOW includes detailed testing requirements for HEPA filter integrity (including upstream test concentration, pass/fail results, patch documentation, and repair up to 3% allowable patch area), biosafety cabinet certification (downflow/inflow air velocity, filter integrity, and alarm testing), and chemical fume hood certification (air velocity grids, smoke patterns, and cross-draft velocity). The Government will furnish replacement HEPA filters and pre-filters as needed.

View the file

Other files for this federal contract opportunity

Other files attached to HEPA Filter Testing, newest first.
File Type Posted
26Q0281.pdf PDF
Attachment 2 Wage Rates.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Scope of Work In-Place HEPA Filter, Magnehelic and Fume hood, BSC Certification National Center for Applied Plant Protection

Service Location: National Center for Applied Plant Protection (NCAPP)

BARC-East, Bldg. 580
Beltsville, MD 20705

Performance Period: The base period starts on the effective date of the contract through one year Optional periods that may be exercised at the discretion of the Government.

The first option period begins the day after the base period ends through one year The second option period begins the day after the first option period ends through one year The third option period begins the day after the second option ends through one year The fourth option period begins the day after the third option period ends through one year Background The BSL-3 Enhanced High Containment Building and Waste Sterilization Building located at the National Center for Applied Plant Protection (NCAPP) in Beltsville, Maryland is utilized for the testing and containment of Ag Select Agents and high consequence plant pathogens. There are Fifty-three (53) in-place HEPA filters located in forty-four (44) pieces of equipment these HEPA filters contribute to the containment of the plant pathogens; the HEPA filters require annual testing and certification. See attachment one to this Scope of Work (SOW).

The In-Place HEPA filter integrity testing/certification shall be in accordance with the NCAPP Standard Operating Procedures (SOP 90-016-01), NSF/ANSI Standard 49 as well as ASME N510-1989 standards for HEPA Filter Bank In-Place Testing. The standards will be provided to the participants of the site visit. All modifications to the HEPA Filter Testing protocol shall be submitted prior to the effective date of the contract; submit to the NCAPP Facility Manager for approval.

Magnehelic testing and certification of thirty (30) gauges located throughout the High Containment facility. Magnehelic gauges are located above room doors to display the differential pressure between two locations and on filter banks to show pressure drop on each side of filter to alarm of a dirty or clogged filter. Each magnehelic gauge shall be field calibrated and certified.

Certification of thirteen (13) Bio-Safety Cabinets located in laboratories throughout the building 580 facility. Certification of nine (9) Chemical Fume hoods located throughout the building 580 facility. The available information such as Location, Make, Model, Serial Number and Property Number is supplied as an attachment to this Statement of Work. Bio-Safety cabinet test reports shall include but not limited to, all pertinent information of the Bio-safety cabinet, down flow air velocity test, inflow air velocity test, filter integrity test, alarms tested and HEPA Filter test if integral to the unit. All Chemical Fume Hoods test reports shall include all pertinent information of the Chemical Fume Hood, including all but not limited to air velocity test with grid, smoke pattern test and cross-draft velocity test. A blank copy of each report shall be submitted with each proposal.

Scope of Work The contractor shall provide all labor, materials, certified test equipment, and challenging agent to perform certification testing. Annual certification is required. The Base Period in-Place HEPA filter integrity testing/certification shall be conducted during the month of performance under this contract and then within a year thereafter for each of the four option periods. All work schedules will need to be reviewed and approved by the Government for access to the facility when Ag Select Agent plant material is not being testing.

All contractor personnel shall always be escorted by Government personnel while entering, exiting, and working inside the High Containment facility and the Waste Sterilization Building. All work shall be conducted during normal working hours 6:30 a.m. through 4:30 p.m. Eastern Time unless approval by the facility manager has been provided prior to the visit. The contractor shall limit disturbance to the normal working operations within the facilities by using the bypass tandem HEPA bank during the certification procedure. A cascading negative pressure gradient must be maintained at all times within the containment building during the testing of the HEPA filters.

The contractor performing air testing shall have a minimum of five years of experience in testing, repairing, and certifying HEPA filters for Federal and State agencies, hospitals, universities, public and/or private institutions and military facilities by NSF and IAFCA accredited technicians.

The following are the tasks that shall be completed as part of this requirement:

· Inspect all pre-filters. Replacement pre-filters will be provided as Government Furnished Material and will be replaced by the Government if required).

· Conduct in-place HEPA filter integrity test and certification for a total of 53 HEPA filters.

· Report test results. The final test report shall include at a minimum the following information for each of the fifty-three (53) HEPA filters:

· Test-face velocity of the filter.

· Upstream test concentration of the challenging agent.

· Test results; pass or fail.

· All previous patch sizes and location.

· All new patch sizes and location.

· Total patch percentages of the filter face.

· Repair any leaks up to the 3% allowable percentage total patch of face area.

· Remove and replace any filter that does not meet the specified criteria; install replacement HEPA filter and conduct certification of the replacement filter. Replacement HEPA filters will be provided as Government Furnished Material.

· Conduct test and certification on a total of thirteen (13) Bio-Safety Cabinets including four (4) Horizontal Laminar Air Flow/Vertical Laminar Air Flow HLAF/VLAF Units

· Report test result on a total of thirteen (13) Bio-Safety Cabinets including four (4) Horizontal Laminar Air Flow/Vertical Laminar Air Flow HLAF/VLAF Units

· Conduct test and certification on nine (9) Chemical Fume Hoods

· Report test results on nine (9) Chemical Fume Hoods

· Test, calibration, and certification of 30+ Magnehelic Gauges.

· Report test results of 30+ Magnehelic Gauges.

Government Furnished Material

· Replacement HEPA filters

· Replacement pre-filters

Apendix 1 - Location, Make, Model, Serial Number and Property Number

Filter
Type
Housing #
Upstream Concentration

(ug/L) Leakage %

EF-2T
B2-212-11
H99123
EF-2B
B2-212-11
H99128
EF-3T
B2-212-11
H99121
EF-3B
B2-212-11
H99122
EF-4T
B2-212-11
H99133
EF-4B
B2-212-11
H99134
EF-5T
B2-212-11
H99126
EF-5T
B2-212-11
H99127
EF-6T
B1-1/21/2
H99151
EF-6B
B1-1/21/2
H99120
EF-7T
B2-212-11
H99125
EF-7B
B2-212-11
H99124
EF-8T
B2-212-11
H99130
EF-8B
B2-212-11
H99129
EF-9T
B2-212-21
H99143
EF-9B
B2-212-21
H99144
EF-10T
B2-212-21
H99148
EF-10B
B2-212-21
H99147
EF-31T
B2-212-12
H99142
EF-32B
B2-212-12
H99140
EF-33T
B1-1/21/2
H99119
EF-33B
B1-1/21/2
H99118
EF-37T
B1-1/21/2
H99111
EF-37B
B1-1/21/2
H99117
PF-12T
B2-212-12
H99136
PF-12B
B2-212-12
H99135
PF-13T
B2-212-12
H99139
PF-13B
B2-212-21
H99141
PF-14T
B2-212-11
H99131
PF-14B
B2-212-11
H99132
PV-104
B1-1/21/2
H99154
PV-110-1
B1-1/21/2
H99155
PV-110-2
B1-1/21/2
H99156
PV-112
B1-1/21/2
H99181
PV-124 East
B1-1/21/2
H99152
PV-124 West
B1-1/21/2
H99113
PVSF-149
B1-1/21/2
H99153
SF-17T
B2-212-12
H99137
SF-17B
B2-212-12
H99138
SF-18T
B1-1/21/2
H99112
SF-18B
B1-1/21/2
H99116

EF-32T

EF-33T
B1-1/21/2
H9919
EF-33B
B1-1/21/2
H9918
Gauges
Location
Make
Model
Serial Number
Property Number

Bio-Safety Cabinets and Chemical Fume hoods

PROJECT CONTACT PERSONS:

Latisha Hollman Latisha.a.hollman@usda.gov

Contracting Officer’s Representative (COR) Mr. Wesley Hutchinson NCAPP Facility operations Specialist

USDA, APHIS, PPQ, FO

BARC-EAST Bldg. 580 Powder Mill Road Beltsville, Maryland 20705 Tel. No.: 301-313-9256 image1.png

File details come from the government source that posted it. Updated .