Attachment 1 - Specs SBRU 06.24.2024.docx

DOCX document 64 KB Posted

Attached to
Electronic Nose Flash Gas Chromatographer (GC) Analyzer Federal contract opportunity
Solicitation number
12805B24Q0544
Issued by
Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Plains Area

About this file

This document is a Statement of Work (SOW) for procuring an electronic nose flash gas chromatographer (GC) analyzer to be used for evaluating aroma profiles of dry bean and pulse flour samples. The analyzer must have capabilities such as liquid and headspace injection, fast analysis time, and software for data processing and reporting.

The related federal contract opportunity is a combined synopsis/solicitation (Solicitation Number 12805B24Q0544) issued by the USDA Agricultural Research Service for the electronic nose flash GC analyzer. The government intends to award a firm fixed price purchase order to the responsible offeror whose quote conforms to the specifications and is the most advantageous considering technical acceptance, past performance, and price. Quotes are due by August 16, 10:00am Central Time, and must address all RFQ requirements and have an active SAM.gov registration.

View the file

Other files for this federal contract opportunity

Other files attached to Electronic Nose Flash Gas Chromatographer (GC) Analyzer, newest first.
File Type Posted
Solicitation 12805B24Q0544.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Statement of Work (SOW) Project: Electronic Nose Flash gas chromatographer (GC) analyzer

GENERAL INFORMATION

1.0 Scope of Work: Purpose is to procure an electronic nose flash gas chromatographer (GC) to be used to evaluate aroma profiles of dry beans and other pulses. The equipment should be able to quickly and accurately conduct sensory and chemical analysis of flour samples to differentiate sensory profiles.

2.0 Background: SBRU conducts plant genetics and food processing research on using dry beans and other pulses as flours. Flour aroma is an important consideration since pulses often have strong beany or vegetative aromas that limit consumer appeal. A method is needed to quickly assess aroma profiles of pulse flour and food product samples. The method should be high throughput with the ability to evaluate thousands of samples per year. The method should be amenable to use as a screening tool in pulse breeding to identify germplasm with acceptable aroma profiles.

CONTRACTOR REQUIREMENTS

3.0 Technical Requirements/Specs: Provide and deliver one Electronic Nose Flash GC analyzer with the following characteristics/capabilities:

*Capable of both Liquid and headspace injection modes via manual or automated injection.

*Fits on a lab bench and weighs less than 20 kg *120V system *Contains at least 2 metal capillary columns *Minimum of 2 Flame Ionization Detectors (FID) with operating temp up to 300 C *Max start-up time is 25 minutes *Detection limits at least < 50pg of n C12 > *Less than 10 min analysis time per sample *Includes Autosampler for gas and liquid samples *Includes Software to control and monitor the instrument.

*Includes a full chemometrics package for compounds and sensory descriptors.

Capable of: Data acquisition, Auto-sampler monitoring and set-up, Data processing, reporting, multivariate statistics.

4.0 Government Furnished: N/A

5.0Deliverables / Schedule: The system should be delivered within 6 months of when order is placed.
Components:
ITEM

QTY

Electronic Nose Flash GC analyzer

Delivery location: Attn Karen Cichy, 1066 Bogue St, rm 367, East Lansing, MI 48824

6.0 Travel: N/A

7.0 Contractor’s Key Personnel: N/A

8.0 Security Requirements: None

9.0 Data Rights: N/A

10.0 Section 508 – Electronic and Information Technology Standards: N/A

Attachment: