Attachment 1 Specifications.pdf
PDF 550 KB Posted
- Solicitation number
- 12905B23Q0162
About this file
This document provides specifications for two quarantine walk-in plant growth chambers to be delivered to the USDA Agricultural Research Service in Hilo, Hawaii. The growth chambers must each have 144 square feet of growing area that can accommodate plant heights from 20 to 72 inches. The chambers must maintain temperatures between 10 to 35 degrees Celsius lights off and 15 to 35 degrees lights on, with 0.5 degree control point. Additional requirements include humidity control between 30-70% RH, carbon dioxide supplementation, 1000 micromoles/m2/s minimum lighting intensity, LED broad spectrum lighting with supplemental far-red, programmable control and monitoring systems, mechanical dehumidification of 6 pounds/hour, HEPA filtered exhaust, and seismic safety standards. The vendor will provide a three-year warranty and annual preventative maintenance for three years on all equipment. Installation and commissioning must be completed within eight weeks of mobilizing to the site.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| SF30_12905B23Q0162_0002.pdf | ||
| Questions and Answers_12905B23Q0162.pdf | ||
| SF30_12905B23Q0162_0001.pdf | ||
| RFQ_12905B23Q0162 Terms and Conditions.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
List of Specifications and Salient Characteristics for Quarantine Walk in Plant Growth Chambers (2)
Delivery Location
USDA ARS DKI PBARC
64 Nowelo Street Hilo, HI, 96720
Item Specification
General o Walk in growth chamber, made with material that is smooth and impervious to liquids and can be cleaned and disinfected.
o 144 sq ft of growing area that can accommodate different heights of plants from 20’ to 72”.
o Air flow in chamber should be able to be switched from controlled vertical airflow for single tier growth mode, to horizontal airflow in the multi-tier configuration.
o Air-cooled system refrigeration and specifically be designed using Ashrae 20-year extreme guidance for Hilo.
o Multiple growth chambers will be connected by vestibule/anteroom which is sealed and prevents contents from growth chamber being directly released into the building.
o External air exhaust will be passed through HEPA filter.
o Room should be secured for potential seismic activity.
Controller o Should be programmable on daily schedule o Data should be able to be viewed locally at chamber and downloaded for record keeping.
o Alarms should allow for local monitoring and through internet or other system to allow for alerts.
Lighting o Provide lighting intensity of 1000 micromoles/m²/s in single tier configuration and 600 micromoles/m²/s in two or three tier configuration.
o Intensity should be able to be controlled for different light requirements, automatic dimming capability should be combined with specific R/FR manipulation using separate 730nm sources.
o Lights should be LED broad spectrum with supplemental far-red light.
o Lights should be monitored and available on data display.
Temperature o Chamber shall operate between 10°C to 35°C lights OFF, 15°C to 35°C ON with ±0.5°C at the control point.
o Provide a programmable min and max temperature or a limit tracking alarm that automatically follows the programmed set point o Provide an independent factory-set high and low temperature limit.
o Sensors are located in a self-contained portable aspirator. This allows accurate measuring and recording by sensing at plant location.
o Adjustable fan speed.
Humidity o Provide additive humidity function achieving up to 70% RH lights OFF/ON, limited by a 25°C maximum dew point.
o Ensure humidity control of ±3% RH and incorporate a dry humidity sensor to directly measure humidity in %RH.
o Provide dehumidification using mechanical means to achieve a desiccation rate of 6 pounds/hour of moisture at minimum.
Supplement Carbon Dioxide o Provide additive C02 control including all necessary sensors, control valves, and injection system for additive CO2 Installation o Vendor shall include full installation of equipment and components at project site.
o Full installation and commissioning shall be completed by vendor within eight weeks of mobilizing to site.
Warranty o Vendor shall provide a full three-year warranty on all equipment including parts and labor.
o Preventative Maintenance: Vendor shall provide a full three-year on-site preventative maintenance program on a minimum once-annual basis.
o Chamber and components should provide a U.S. electrical inspection marking by an authority accredited by OSHA. This third-party certification shall apply to the chamber as an entire unit, fit for purpose. Incorporation of individual UL listed components, not certified as an entire unit fit for purpose, shall not be acceptable.