This document is an Attachment containing the salient characteristics and technical specifications for a BLU-MED 16' x 32.5' XP® Medical Response Package offered on a brand name or equal basis. The Government requires one complete deployable medical response package meeting minimum physical, functional, and performance requirements across six primary subsystems.
The shelter component must be a 16 feet wide x 32.5 feet long x 8 feet high Quonset-style structure with a free-span aluminum frame suitable for saltwater or corrosive environments. It shall feature 15 oz. Mil-Spec vinyl cover and end panels certified to Mil-PRF-44103D, minimum two entries and four insulated windows, non-slip removable flooring, and fire-resistant plenum and insulating liner. The shelter must withstand 15 psf snow load, 80 mph sustained winds with 100 mph gusts, 4 inches of rain per hour, and 100 setup/strike cycles without critical failure. The ECU/HVAC system requires minimum 30,000 BTUs cooling capacity, 8 kW heat strip, 1 inch static pressure and 1100 CFM airflow, with operational temperature range of -25°F to +132°F and maximum 45 dB operating sound level. The electrical distribution system shall include an interior power distribution box with minimum two GFCI-protected outlet circuits and at least 10 individually controlled water-resistant lights. The generator must be 20 kW, 60 Hz, 3-phase diesel with approximately 45-gallon fuel capacity in a noise-attenuating enclosure. The storage/transport trailer shall be approximately 7 feet wide x 14 feet long with minimum 7,000 lbs GVWR, featuring rear ramp door, side door, electric brakes with breakaway kit, lighting, and tandem axles. The complete package includes carry bags, spares/repair kits, instruction manuals, parts lists, and top-level drawings. Offerors proposing equal products must provide sufficient technical documentation demonstrating compliance with all listed salient characteristics.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
ATTACHMENT 1
SALIENT CHARACTERISTICS
BLU-MED 16' x 32.5' XP® MEDICAL RESPONSE PACKAGE
BRAND NAME OR EQUAL
1. General
The Government requires one complete deployable medical response package, brand name or equal, meeting the following minimum physical, functional, and performance characteristics.
2. Shelter
The shelter shall meet the following minimum requirements:
1. Approximate dimensions of 16 feet wide x 32.5 feet long x 8 feet high in a Quonset-style configuration.
2. Free-span aluminum frame with no center poles.
3. Corrosion-resistant frame suitable for saltwater or corrosive environments.
4. Cover and end panels made of 15 oz. Mil-Spec vinyl certified to Mil-PRF-44103D or equivalent.
5. Minimum of two entries, four insulated windows, and two sets of HVAC/ECU ports.
6. Double-layer end walls with vestibule-connect capability.
7. Non-slip floor that is removable and replaceable while erected, with sealing features to reduce dust, insects, and water intrusion.
8. Fire-resistant plenum and insulating liner system.
9. Capability to connect end-to-end with same-width/height shelters.
10. Demonstrated performance equivalent to:
15 psf snow load
80 mph sustained wind with 100 mph gusts
4 inches rain per hour with 60 mph sustained wind and 75 mph gusts
100 setup/strike cycles without critical failure
3. ECU/HVAC
The ECU/HVAC system shall meet the following minimum requirements:
1. Minimum 30,000 BTUs cooling capacity.
2. Minimum 8 kW heat strip.
3. Minimum 1 inch static pressure and 1100 CFM airflow.
4. Operational temperature range approximately -25°F to +132°F.
5. Maximum A/C operating sound level of approximately 45 dB.
6. Includes insulated ducts, power cable, fresh air filter, and forklift-compatible handling features or equivalent.
4. Electrical Distribution and Lighting
The package shall include an electrical distribution and lighting system meeting the following minimum requirements:
1. Interior power distribution box for shelter lighting and convenience outlets.
2. Minimum of two GFCI-protected outlet circuits and sufficient outlet capacity for shelter operations.
3. Interior lighting system with at least 10 individually controlled lights or equivalent lighting capability.
4. Lighting shall be water resistant and suitable for field use.
5. Generator
The package shall include a generator meeting the following minimum requirements:
1. 20 kW, 60 Hz, 3-phase generator, or equivalent.
2. Prime and standby power output sufficient to support the complete shelter package.
3. Diesel engine in noise-attenuating enclosure.
4. Fuel capacity of approximately 45 gallons, or equivalent.
5. Controls and safety/protection features suitable for deployable field use.
6. Storage/Transport Trailer
The package shall include a storage/transport trailer meeting the following minimum requirements:
1. Approximate size of 7 feet wide x 14 feet long.
2. Minimum GVWR of 7,000 lbs.
3. Trailer shall support secure storage and transport of the complete shelter package.
4. Trailer shall include a rear ramp door, side door, electric brakes with breakaway kit, lighting, tandem axles, and spare tire, or equivalent transport features.
7. Included Components
The package shall include, at a minimum:
1. Shelter
2. ECU/HVAC system
3. Electrical distribution components
4. Interior lighting
5. Generator
6. Storage/transport trailer
7. Carry bags
8. Spares/repair kits
9. Instruction manuals
10. Parts list
11. Top-level drawings or equivalent manufacturer documentation
8. Equal Product Documentation
Offerors proposing an equal product shall provide sufficient technical documentation to demonstrate compliance with all listed salient characteristics.