Capillary Electrophoresis (CE) Instrument
Solicitation number
20341426Q00033
Issued by

About this file

This document is a Salient Characteristics attachment for a federal procurement of a Capillary Electrophoresis (CE) instrument by the Beverage Alcohol Laboratory (BAL) of the Alcohol and Tobacco Tax and Trade Bureau (TTB), located in Beltsville, Maryland.

The BAL seeks a CE instrument capable of analyzing, separating, and quantitating eight organic acids in alcoholic beverages: tartaric acid, acetic acid, malic acid, lactic acid, citric acid, gluconic acid, succinic acid, and fumaric acid. The instrument will support determination of standard of identity for formula submissions, product integrity investigations, and the Alcohol Beverage Sampling Program. Future expansion capabilities should include analysis of sugars (erythritol, fructose, glucose, sucrose, maltose, and lactose) and cyanide detection. The instrument must operate under specified conditions: 100 mbar injection pressure, 20°C capillary temperature, -25 kV voltage, and 350/20 nm detector signal with 200/10 nm reference. Required specifications include a programmable pressure system (-100 to +100 mbar), safety features (current leak detection, liquid leak sensor, high voltage safety sensors), self-correcting injection system with programmable range up to 10,000 seconds, 50-position autosampler with temperature control (10-40°C), satellite station for buffer replenishment, and 100 μL, 1 mL, and 2 mL vial compatibility. The capillary cassette must include a high-speed forced-air cooler with Peltier element, temperature range 10°C below ambient to 60°C, and support 365 μm outer diameter capillaries with minimum 33 cm length. The detector must be a real-time UV-Visible diode-array detector (190-600 nm) with 1 nm wavelength accuracy, 0.025 to 10 second response time, 1 μM sensitivity, and baseline noise less than 50 μAU at 2 second response time. The system must operate under Windows XP (SP3) with time-programmable parameters, constant voltage/current/power modes, voltage range 0 to ±30 kV, current 0-300 μA, and power 0-6 W. Power requirements are 100-240 V line voltage, maximum 300W, 50-60 Hz frequency. The winning proposal must include five line items: CPU with monitor and printer, familiarization training at installation, on-site method consulting, an Organic Acids Kit, and the capability to add multiple detectors in the future.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Capillary Electrophoresis (CE) Instrument AƩachment 1: Salient CharacterisƟcs

The Beverage Alcohol Laboratory (BAL) located at 6000 Ammendale Road, Beltsville, Maryland, is part of the ScienƟfic Services Division of TTB. The laboratory provides technical experƟse and assistance to other divisions within TTB by analyzing approximately 1,200 alcoholic beverage samples (wine, malt beverages, and disƟlled spirits) in a year across a variety of programs.

TTB’s Beverage Alcohol Laboratory (BAL) is interested in purchasing a capillary electrophoresis

(CE) instrument that would be capable of performing BAL’s previously developed test method for determining organic acids in alcoholic beverages. This would be used for determinaƟon of standard of idenƟty for Formula submissions with lab analysis, Product Integrity invesƟgaƟons, and the Alcohol Beverage Sampling Program (ABSP). The instrument also allows BAL to possibly expand their capabiliƟes to determine sugars (erythritol, fructose, glucose, sucrose, maltose, and lactose) used in nutriƟonal labeling assessment, and cyanide in alcoholic beverages which is a limited ingredient.

Method Parameters

The instrument should be able to analyze, separate, and quanƟtate the following organic acids in alcoholic beverages such as wines, malt beverages, and disƟlled spirits:

Tartaric Acid AceƟc Acid

Malic Acid LacƟc Acid

Citric Acid Gluconic Acid

Succinic Acid Fumaric Acid

Capillary Electrophoresis System AƩachment 1: Salient CharacterisƟcs

Instrument CondiƟons include:

Parameter Value

Injection Pressure 100 mbar

Capillary Temperature 20°C

Voltage -25 kV

Detector Signal 350/20 nm, Reference 200/10nm

Instrument Specifications

Pressure system

Programmable with -100 to +100 mbar on inlet

Flushing with 1 bar or with high pressure 2–12 bar

Vial pressurization with high pressure 2–12 bar on inlet and/or outlet

Safety features

Current leak detection low current limit

Liquid leak sensor

Safety sensors at door and cover disabling high voltage

Vial sensor

Injection modes

Self correcting injection system with injection from inlet or outlet

Programmable range up to 10,000 seconds

Pressure –100 to +100 mbar

Electrokinetic –30 to +30 kV

Autosampler/ fraction collector

50-position carousel

All vials are randomly accessible from inlet or outlet end of capillary

Temperature control with external waterbath with vial temperature from 10–40 °C (non-condensing conditions, minimum waterbath temperature +1 °C)

Replenishment

Satellite station for refilling of inlet or outlet buffer vials with fresh buffer for automatic, continuous operation.

Selectable buffer levelling.

Vials

100 μL (polypropylene or glass) with resealing snap caps

1 mL (polypropylene) with resealing snap caps

2 mL (glass) with resealing snap caps

Capillary cassette

High-speed forced-air cooler with Peltier element

Temperature range 10 °C below ambient, up to 60 °C (min. 10 °C cassette temperature)

Minimum total capillary length 33 cm

Capillary compatibility 365 µm o.d.

Detector

Real time UV-Visible diode-array detector (190–600 nm)

Temperature controlled

Wavelength accuracy

1 nm

Response time 0.025 to 10 s

Light source prealigned deuterium or high brightness lamp

Signals up to eight signals simultaneously, full spectral acquisition with Agilent ChemStation

Sensitivity 1 μM 4-hydroxy-aceto-phenon injected at 50 mbar*5 sec, 3 x 50 µm bubble cell capillary, signal/noise >6* (50 mM borate buffer pH 9.3, 25 kV)

Baseline noise <50 μAU (at 2 sec response time)*

Linear dynamic range

1x104 (3x50 μm bubble cell capillary)*

Raw data channels

Detector signals and spectrum, voltage, current, leak current, power, cassette temperature, pressure, lamp voltage and detector temperature

Diagnostic functions

RFID tag for lamp, early maintenance feedback, supported by Agilent Lab Advisor software with integrated diagnosis test suite

System control

Operating with graphical user interface under WindowsXP (SP3)

Time programmable parameters

Time programmable parameters: voltage, current, power, polarity, pressure, inlet and outlet vial, capillary temperature, pre and post-run conditioning with pressure and/or voltage, Replenishment, fraction collection**

Electrophoresis power

Voltage range setable 0 to ± 30 kV supply

Current setable 0–300 μA

Power setable 0–6 W

Operation under constant voltage, current or power Programmable polarity switch

CE specific software

Mobility report, time corrected areas, pI calibration and bio polymer size calibration

Power requirements

Line voltage 100-240 V, max. 300W

Line frequency 50-60 Hz

Acquisition Requirements

The winning proposal will meet the conditions specified above in the Method Parameters and

Instrument Specifications sections. It will also include the following items in line item format:

1. CPU including monitor and printer

2. Familiarization at installation

3. Method consulting on site

4. Organic Acids Kit

5. The ability to add multiple detectors if needed in the future.