Attachment 1 - Salient Characteristics.docx
DOCX document 26 KB Posted
- Solicitation number
- 12805B25Q0032
About this file
This document is an Attachment 1 - Salient Characteristics for a federal contract opportunity. The solicitation number is 12805B25Q0032 for two (2) pressure and control pumps for automated Triaxial and Permeability Tests. The pumps must meet or exceed 500 psi with fully automated control and data acquisition, and have specific volume and pressure accuracy requirements. The pumps must be compatible with the existing Geocomp LoadTrac II soil testing system and software. The pumps must be delivered by February 1, 2025 to the USDA Agricultural Research Service facility in Stillwater, OK.
The related federal contract opportunity is a combined synopsis/solicitation for a commercial item. It is an RFQ (Request for Quote) with a NAICS code of 334516 and a small business size standard of 1,000 employees. The USDA Agricultural Research Service has a requirement for the two (2) pressure and control pumps. The government intends to award a firm fixed price purchase order. Quotes are due by 1700 ET on November 27, 2024 and should be sent via email.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 12805B25Q0032 Combined Synopsis Solicitation.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specifications / Salient Characteristics Pressure and Control Pumps for Automated Triaxial and Permeability Tests Solicitation No. 12805B25Q0032
Technical Requirement:
FlowTrac II Pressure and Control Pumps for Automated Triaxial and Permeability Tests (Brand Name or Equal)
· Part Number: F1-40677
· Quantity: Two (2) each
Salient Characteristics:
· Pressure and volume control pumps shall meet or exceed 500 psi with fully automated control and data acquisition compatible with existing facility’s Geocomp LoadTrac II soil testing system and software.
· Volume accuracy should meet or exceed 0.0005%.
· Pressure accuracy should meet or exceed 0.03%.
· Measurement and control accuracy should meet or exceed 0.35 kPa and 0.001 cc volume.
· Accurate volume rate control should meet or exceed 0.00001 to 36 fl oz/min (0.0003 to 1054 cc/min).
· Calibration of all equipment with NIST Traceable Calibration Certificate
Delivery Address Delivery shall be received by February 1, 2025, at:
| 9501 W. Lakeview Rd. |
| Stillwater, OK |
| 74075 |
Point of Contact Technical point of contact: Daniel Livsey, daniel.livsey@usda.gov, 405-269-1175