Attachment 1 - PWS.docx
DOCX document 21 KB Posted
- Solicitation number
- 36C25926Q0576
About this file
This is a Performance Work Statement (PWS) for comprehensive maintenance and repair services of sterilization, washing, and drying equipment at the Salt Lake City VA Medical Center.
The contractor must provide full-service maintenance ensuring all equipment operates in conformance with original manufacturer specifications and NFPA-99, UL, OSHA, and VA manuals, with a 95% operational uptime requirement during normal operating hours (8:00 AM-5:00 PM, Monday-Friday, excluding national holidays). Emergency repair services must be available 24 hours a day, 7 days a week, with a 30-minute response time for phone calls and a 4-hour maximum on-site arrival time if the issue cannot be resolved remotely. The contractor must supply all parts, tools, service manuals, diagnostic software, and equipment necessary to meet performance standards, using only new OEM-equivalent or superior parts. Scheduled preventive maintenance services must be performed semi-annually on all equipment, including cleaning, software diagnostics review, calibration, lubrication, parts replacement, electrical and mechanical component inspection, and annual chamber cleaning for steam sterilizers. Documentation requirements include detailed Equipment Service Reports submitted via email to the Contracting Officer's Representative, documenting technician name, service dates and times, equipment identification, problems reported, services performed, and associated costs. The PWS covers 41 equipment items across multiple categories including ultrasonic cleaners, sterilizers, disinfectors, warming cabinets, prep and pack tables, and transfer carriages, with inspection frequencies ranging from 1 to 4 annual inspections depending on equipment type.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment 1 - PWS Revised Final.docx | DOCX document | |
| 36C25926Q0576 0001.docx | DOCX document | |
| Attachment 2 - Price Schedule.docx | DOCX document | |
| 36C25926Q0576.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
B.2 PERFORMANCE WORK STATEMENT
The Contractor shall provide comprehensive full-service maintenance, routine and emergency repair service for the government owned sterilizers, washers, dryers and other related equipment located at the Salt Lake City VA Medical Center.
Contract service shall ensure that the equipment functions in conformance with the original manufacturer’s specifications and the latest published edition of NFPA-99, UL, OSHA and VA manuals.
All equipment shall be operable and available for use 95% of the normal operating hours of the equipment, which are 8:00 AM-5:00 PM, Monday through Friday, excluding national holidays. Downtime will be computed from notification of the problem during normal work hours. Scheduled maintenance will be excluded from downtime. Operational uptime will be computed on a monthly basis. Repeated failure to meet this requirement may subject the contractor to default actions.
All service/repairs shall be performed during normal working hours unless requested or approved by the COR or designee. Work performed outside of normal working hours at the request of the COR or designee will be billed at the contractors prevailing hourly service rate.
Unscheduled Maintenance (Emergency Repairs) Contractor will provide emergency repair service between regularly scheduled services and calibration. Priority technical support and remote monitoring will be provided 24 hours a day, 7 days a week for the duration of the contract.
The Field Service Engineer (FSE) must respond with a return phone call to the COR or designee within 30 minutes of receipt of telephone notification of an issue, 24 hours a day. If the problem cannot be resolved over the phone, the FSE will arrive on-site and commence repairs no more than 4 hours after receipt of notification that the problem remains unresolved.
The contractor will only use parts that are equal to or exceed OEM specs.
Scheduled Maintenance The contractor shall perform scheduled PM services semi-annually on all equipment unless otherwise noted, ensuring that all equipment performs in accordance with the original manufacturer’s specifications and the latest published edition of NFPA-99, UL, OSHA and VA manuals. An outline of the PM procedures and schedule will be provided to the COR within two weeks of award of contract. Contractor will utilize manufacturer recommended procedures and checklists, documenting work performed and actual values obtained (as applicable) during each PM and will provide this documentation to the COR at completion. PM services shall include, but need not be limited to, the following:
1) Cleaning of equipment
2) Reviewing operating system software diagnostics to ensure that the system is operating in accordance with manufactures specifications
3) Calibrating and lubricating the equipment
4) Performing remedial maintenance of non-emergent nature
5) Testing and replacing faulty and worn parts and/or parts likely to become faulty, fail or become worn.
6) Inspecting electrical wiring and cables for wear, replacing as needed.
7) Inspecting all mechanical components such as door assemblies, gaskets, steam and water valves, solenoid valves, pressure regulator valves, steam, water strainer, jacket and chamber strainer, air filter cartridge, chamber and jacket traps, printer assembly, battery-backed RAM, buzzer, water level sensor, cooling fan, fan filter, air connection, steam, water and drain connections, door lock piston, bearings and tracks, interlocks, motors, keypad, air pressure, door latch switches, distilled water reservoir, EO gas vent line and moisture trap. All listed items shall be replaced as needed.
8) Annual chamber cleaning of all steam sterilizers
9) Returning equipment to operating condition in full compliance with specs at completion of PM.
10) Providing documentation of services performed.
11) PM services shall be provided in accordance with the PM schedule provided herein. Any exceptions must be arranged and approved in advance with the COR or designee.
12) All parts, services, manuals, tools or software required to successfully complete the PM schedule are included within this contract unless specifically stated in writing otherwise.
Parts The contractor shall furnish and replace parts as needed to meet uptime requirements. The contractor will have ready access to all parts, including unique and/or high mortality replacement parts. All parts shall be compatible with existing equipment and meet or exceed all manufacturer specs. Rebuilt, used, or parts removed from another piece of equipment shall not be installed. This contract will include all parts required to maintain required uptimes unless specifically excluded in writing.
Service Manuals/Tools/Equipment The contractor will supply all tools, test equipment, service manuals and diagnostic software to meet all performance standards of this contract.
Documentation/Reports Documentation shall include detailed description of all parts and services provided for scheduled and unscheduled (i.e. emergency repairs) services performed. Each report must document the following data in detail:
1) Name of the contractor and contract number.
2) Name of Field Service Engineer performing the service
3) Date, times (start and finish), equipment downtime, and hours on site for service call.
4) VA Purchase Order number if service is provided outside of normal working hours.
5) Detailed description of reported problem.
6) Identification of equipment serviced: Inventory number, Mfr. Name, device name, model number, serial number or any other manufacturer identification numbers.
7) Itemized description of service performed, including any costs associated with parts and/or services performed outside of normal working hours.
Reporting Requirements When arriving to perform both scheduled and unscheduled services, the FSE must report and sign in at the Boiler Plant, located in Building 6, upon arrival to the VA Medical Center. After signing in he/she shall proceed directly to the area being serviced and locate the equipment operator who reported the problem, or an area supervisor. When services/repairs are completed the FSE will document services rendered on a legible Equipment Service Report (ESR), then submit the ESR to the equipment user or area supervisor for an “acceptance signature”. A copy of the accepted ESR will be sent via e-mail to the COR as verification of services rendered.
Equipment list, inspections, and # of inspections CAVIWAVE ULTRASONIC CLEANER 20 GAL 208V/1PH/132KHZ 0217TC0006 BLDG 14, 2ND FLR SPD No. of Annual Inspections 2
CAVIWAVE ULTRASONIC CLEANER 20 GAL 208V/3PH/132KHZ 0417TC0011 BLDG 14, 2ND FLR SPD No. of Annual Inspections 2
CAVIWAVE PRO ULTRSC CNSL L2RWF 17GAL208V/3PH/132HZ 0512ST0120057 DECONT
No. of Annual Inspections 2
AMSCO DRYING CABINET, CABINET, 38IN, PURGE, 120V 2749 SPD Clean side.
No. of Annual Inspections 2
24 IN. DIGITAL WARM CABINET 0425608027 Cardiology 2nd floor No. of Annual Inspections 2
24 IN. DIGITAL WARM CABINET 0402709062 OR Recovery No. of Annual Inspections 2
24 IN. DIGITAL WARM CABINET 0422408061 Surgery Outpatient No. of Annual Inspections 2
VISION 1327 CW DISINFECTOR,200-208V, 60HZ, STEAM, DPD (WIN 7) 3625816002 SPD
No. of Annual Inspections 4
48 IN. PLATFORM PREVAC STERILIZER, SGL HINGE, LH, REC,208/240V 032160921 Building 14; SPD #2 No. of Annual Inspections 2
48 IN. PLATFORM PREVAC STERILIZER, SGL HINGE, LH, REC,208/240V 031840906 Building 14, SPD #1, No. of Annual Inspections 2
60 IN. PLATFORM PREVAC STERILIZER, SGL HINGE, STM, REC, LH,208/240V 031940932 Building 14, SPD No. of Annual Inspections 2
V-PRO MAX, SD, REC, 208/230V, 60HZ ENGLISH 030361311 2B 42B Bldg. 14 S5 In Wall No. of Annual Inspections 2 V-PRO MAX, SD, CAB, 208/230V, 60HZ ENGLISH 030321317 BLDG. 14, RM2B42B S6 Stand alone.
No. of Annual Inspections 2
48 IN. PLATFORM PREVAC STERILIZER, SGL HINGE, LH, REC,208/240V 032160921 Building 14; SPD #2 NA-Chamber Cleaning No. of Annual Inspections 1
48 IN. PLATFORM PREVAC STERILIZER, SGL HINGE, LH, REC,208/240V 031840906 Building 14, SPD #1, NA-Chamber Cleaning No. of Annual Inspections 1
60 IN. PLATFORM PREVAC STERILIZER, SGL HINGE, STM, REC, LH,208/240V 031940932 Building 14, SPD NA-Chamber Cleaning No. of Annual Inspections 1
AMSCO PREP AND PACK TABLE DELUXE ELECTRIC 226170304 bldg. 14 FL 2 No. of Annual Inspections 1
AMSCO PREP AND PACK TABLE DELUXE ELECTRIC 226170306 bldg. 14 FL 2 No. of Annual Inspections 1
AMSCO PREP AND PACK TABLE DELUXE ELECTRIC 226170308 bldg. 14 FL 2 No. of Annual Inspections 1
AMSCO PREP AND PACK TABLE DELUXE ELECTRIC 230170306 bldg. 14 FL 2 No. of Annual Inspections 1
AMSCO PREP AND PACK TABLE DELUXE ELECTRIC 244170302 bldg. 14 FL 2 No. of Annual Inspections 1
WAR CAB STD SOLID SINGLE 24 IN 120V 0404219095
No. of Annual Inspections 2
WAR CAB STD SOLID DUAL 18 IN 120V
No. of Annual Inspections 2
WAR CAB STD SOLID DUAL 18 IN 120V 0409411010 ROOM 2A17-1
No. of Annual Inspections 2
WAR CAB STD SOLID DUAL 24" 120V 0410210147
No. of Annual Inspections 2
WAR CAB STD SOLID DUAL 24 IN. 120V MOBILE 0411024015
No. of Annual Inspections 2
WAR CAB STD SOLID DUAL 24" 120V 0412410096
No. of Annual Inspections 2
24 IN. DIGITAL WARM CABINET 0419808098
No. of Annual Inspections 2
WAR CAB STD SOLID DUAL 24" 120V 0420710166 2 WEST
No. of Annual Inspections 2
WAR CAB STD SOLID DUAL 24" 120V 0420710167
No. of Annual Inspections 2
IW ECO 5.2 GAL UC 100-127VAC 1PH 50/60HZ 05214680201 CP
No. of Annual Inspections 2
IW ECO 5.2 GAL UC 100-127VAC 1PH 50/60HZ05214680202 CP
No. of Annual Inspections 2 IW ECO 5.2 GAL UC 100-127VAC 1PH 50/60HZ 09217213604 Second Floor SPD No. of Annual Inspections 2
WAR CAB STD SOLID SINGLE 18 IN. 120V 0415014118
No. of Annual Inspections 2 WAR Cab serial 0403607013 No. Annual Inspections 2
Reliance/Hamo Vision Single Chamber Washer 3617017011 No. Annual Inspections 4
Reliance/Hamo Vision Single Chamber Washer 3617017010 No. Annual Inspections 4
Atlas Transfer Carriage 031122146 No. Annual Inspections 4
Atlas Transfer Carriage 031112109 No. Annual Inspections 4
Atlas Transfer Carriage 031032111 No. Annual Inspections 4
M!LINK SOFTWARE SUBSCRIPTION ML4196401 CONNECTCARE SERVICE PORTAL
The C&A requirements do not apply, and that a Security Accreditation Package is not required # Dept Agreement Product Coverage Type Annual