This is a Product Description and Minimum Essential Characteristics attachment for a federal contract opportunity. CLIN 0001 specifies a complete NGS STAR system for automated library preparation in next-generation sequencing applications. The system includes the NGS STAR instrument configured with an 8/96 AL ISWAP pipetting head, extended housing, waste management components, thermal cycling capability, magnetic separation plates, and a Windows 11 system controller. The package encompasses all necessary accessories, including Ergotron monitor and keyboard mounting hardware, INHECO 96-well thermal cycler, Hamilton liquid handling components with conductive filter tips in three volumes (50 μL, 300 μL, and 1000 μL), and HHS2 cooling modules for temperature control from 4°C to 70°C. Consumable kits for FVK II independent channels and MPH-96 applications are included, along with shipping, inside delivery by four-person team, installation, method programming, and operator training services.
The minimum essential characteristics require the system to perform automated library preparation with validated pre-programmed protocols for both Illumina Viral Surveillance Panel version 2 and Qiagen QIAseq xHYB Viral and Bacterial Library Kit. The system must process up to 96 samples simultaneously using eight independent 1000 μL pipetting channels with asymmetric span-8 spacing capability for varied tube geometries. Technical requirements mandate locked-in-place tip coupling equivalent to Hamilton CO-RE II technology to prevent misalignment, conductive liquid level detection (cLLD) for low-volume samples, accurate pipetting of 10 μL to 1000 μL volumes, an integrated 96-well thermal cycler equivalent to the Inheco 96 ODTC Left, high-performance active cooling modules, and a universal magnetic plate optimized for bead-based extraction and cleanup protocols.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Synopsis Solicitation - HT942726QE016.pdf | ||
| A02 JA for Other than Full and Open Competition_Redacted.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Attachment 1 – Product Description and Minimum Essential Characteristics
CLIN 0001 shall include:
Description Quantity NGS STAR (New), 8/96 AL ISWAP 1
MPH EXT HOUSING, LEFT, STAR-2 1
Waste Chute, MPH, Left-Side Front 1 CO-RE Grip 1mL Waste Block, NEW 1 STAR Table w/ Shelf and Doors 1 System Controller, WIN11 1 Ergotron Monitor Keyboard Arm 1 Ergotron Arm Mounting Bracket, Standard Tables 1 Cable Extension Kit, Ergotron Monitor Arm 1 NGS STAR MOA ARW Hardware Kit (STAR-2) 1 NGS ARW Hardware Kit 1
INHECO 96 ODTC LEFT 1
Driver, ODTC Venus 1 ABI Plate Adapter 1 Magnum FLX Enhanced Universal Mag Plate - no springs 1 Driver, ODTC Venus 1
HHS2 3.0 FLAT BOTTOM 1
Control Box - HHS/HHC 1 DeckWatch Monitoring Kit 1 DeckWatch Connect Kit 1 50 μL CO-RE II Conductive Filter Tips 1 300 μL CO-RE II Conductive Filter Tips 1 1000 μL CO-RE II Conductive Filter Tips 1 Hamilton PCR ComfortLid 1 60 mL Reagent Reservoir Self-Standing with Barcode 1 Clear waste bags that fit the white plastic waste 1 Inside Delivery 4 Person 1 Installation, STAR 1 Method Programming and Operator Training 1 Method Programming and Operator Training 1 FVK II Independent Channels Consumable Kit 1 1 FVK II MPH-96 Consumable Kit 4 1 FVK II Solutions 1 Shipping and Handling 1
MEC Package Title: NGS STAR
NOMENCLATURE: Automated Library Preparation for Next Generation Sequencing
Minimum Essential Characteristics
- Automated library preparation for next generation sequencing
- Must include validated pre-programed protocols for Illumina Viral Surveillance
Panel version 2 and Qiagen QIAseq xHYB Viral and Bacterial Library Kit
- Be able to process up to 96 samples at a time.
- Must include eight (8) independent 1000 𝜇𝜇L pipetting channels capable of asymmetric spacing (span-8) to perform liquid transfers from varied tube geometries to microplates Ability to minimize hazardous waste generation through segregated sample and solvent waste streams.
- Must utilize locked-in-place tip coupling (equivalent to Hamilton CO-RE II technology) to prevent tip misalignment, drop-offs, or spatial distortion during high-force lateral movements. The system must support conductive liquid level detection (cLLD) to detect liquid boundaries in low-volume wastewater samples dynamically.
- Must accurately and reproducibly pipet volumes ranging from 10 𝜇𝜇L to 1000 𝜇𝜇L using aerosol-barrier, conductive filter tips (50 𝜇𝜇L, 300 𝜇𝜇L, and 1000 𝜇𝜇L).
- Must feature an integrated, automated 96-well thermal cycler (equivalent to the Inheco 96 ODTC Left) to perform automated on-deck PCR amplification.
- Must feature high-performance active cooling modules capable of chilling master mixes, reagents, and samples at a temperature range of 4∘C to 70∘C.
- Must feature a high-performance, universal magnetic plate optimized for bead-based clean-up and extraction protocols.