Attachment 1 Minimum Specifications.pdf
PDF 85 KB Posted
- Attached to
- Refrigerated Shaking Incubators
- Solicitation number
- FA700026Q0065
About this file
This document specifies minimum technical requirements for refrigerated shaking incubators and related laboratory equipment across five distinct contract line items (CLINs).
CLIN 0001 requires a refrigerated shaking incubator with active cooling capable of maintaining temperatures from 16.0°C to at least 60.0°C, with cooling capability to 15°C below ambient room temperature. The unit must feature a digital microprocessor controller with LED, LCD, or touchscreen display; temperature accuracy of ±0.1°C to ±0.2°C at 37°C; continuously adjustable shaking speed from 30 to 300 RPM; and an orbit diameter between 0.75 and 1.0 inches. The interior chamber must be constructed of corrosion-resistant stainless steel. The incubator must accommodate a universal platform (CLIN 0002) capable of holding either six 1-liter Erlenmeyer flasks or sixteen 250 mL Erlenmeyer flasks simultaneously, be stackable to at least 2 units high, operate on 120V ± 10% at 60Hz with cUL/UL certification, and include a minimum one-year commercial warranty covering parts and labor. CLIN 0002 specifies two universal platforms with an 18" x 18" minimum footprint constructed of corrosion-resistant materials compatible with CLINs 0001, 0003, and 0004. CLIN 0003 requires four 1-liter spring-action Erlenmeyer flask clamps with corrosion-resistant mounting hardware for attachment to CLIN 0002. CLIN 0004 specifies twenty 250 mL spring-action Erlenmeyer flask clamps with compatible corrosion-resistant mounting hardware. CLIN 0005 requires a manufacturer-approved stacking kit for safely stacking two refrigerated shaking incubators, with mounting hardware that must not obstruct rear exhaust vents, cooling intake grilles, power plugs, or service access panels.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Solicitation - FA700026Q0065.pdf | ||
| Attachment 2 Buy American Certificate.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Minimum Specifications (Salient Characteristics)
CLIN 0001: Refrigerated Shaking Incubator meeting or exceeding the following minimum specifications:
• Temperature Range (Refrigerated): Must feature active refrigeration cooling. The unit must be capable of maintaining setpoint temperatures from a minimum of 16.0°C up to a maximum of at least 60.0°C. To guarantee stable operation at the low end of the range, the unit's cooling system must have the capability to cool the chamber to at least 15°C below the ambient room temperature.
• Control Interface: Must feature a digital microprocessor controller with an LED, LCD, or touchscreen display for precise monitoring of temperature and speed.
• Temperature Accuracy: Must maintain high temperature accuracy of at least ±0.1°C to ±0.2°C at 37°C.
• Agitation Speed: Shaking speed must be continuously adjustable across a minimum full range of 30 RPM to 300 RPM.
• Orbit Diameter: Must have an orbit diameter between 0.75 inches (19 mm) and 1.0 inch (25 mm).
• Platform & Capacity: Must be compatible with a universal platform (CLIN 0002) capable of accommodating and securing up to six (6) 1-Liter Erlenmeyer flasks simultaneously, or at least sixteen (16) 250 mL Erlenmeyer flasks simultaneously.
• Chamber Material & Maintenance: The interior chamber must be constructed of corrosion-resistant stainless steel to allow for standard laboratory wipe-down and easy cleaning in the event of a flask spill.
• Stacking: Must be stackable to at least 2 units high (CLIN 0005).
• Power: Must operate on standard US voltage (120V ± 10%, 60Hz) with standard cUL/UL certifications.
• Warranty: Must include a standard commercial warranty of at least one (1) year covering parts and labor.
CLIN 0002: Universal Platform meeting or exceeding the following minimum specifications:
Description: Two (2) each, Universal platforms with a minimum physical footprint of 18" x 18" capable of accommodating and securing up to six (6) 1-liter Erlenmeyer flasks simultaneously, or at least sixteen (16) 250 mL Erlenmeyer flasks simultaneously. Must be constructed from corrosion-resistant materials to prevent rusting in high-humidity incubation environments. Must be fully compatible with CLINs 0001, 0003, and 0004.
CLIN 0003: 1L Flask Clamps/Holders
Description: Four (4) each, 1-liter spring-action Erlenmeyer flask clamps constructed of corrosion-resistant materials. Must include compatible corrosion-resistant mounting hardware to attach each individual clamp securely to CLIN 0002.
CLIN 0004: 250 mL Flask Clamps/Holders
Description: Twenty (20) each, 250 mL spring-action Erlenmeyer flask clamps constructed of corrosion-resistant materials. Must include compatible corrosion-resistant mounting hardware to attach each individual clamp securely to CLIN 0002.
CLIN 0005: Stacking Kit
Description: Manufacturer approved stacking kit intended to safely stack 2 refrigerated shaking incubators. Must feature mounting hardware that secures the upper unit's frame directly into the lower unit's chassis frame. The installation of the stacking kit must not block, cover, or impede access to critical rear exhaust vents, refrigeration cooling intake grilles, power plugs, or technical service access panels of either the top or bottom incubator units (CLIN 0001). It must be fully compatible with CLIN 0001.
File details come from the government source that posted it. Updated .