Attachment 1- Minimum Specifications.pdf
PDF 78 KB Posted
- Attached to
- Matrix Switch System
- Solicitation number
- N00173-25-Q-1301269372
About this file
This document details the minimum specifications for a Matrix Switch System for the Naval Research Laboratory (NRL). The system requires a Keysight Technologies E5250A mainframe with one year of manufacturer support, four E5252A-FG matrix switches, a Relay Function Test Adapter, and a rack mount kit. The system must be configured as a 10x48 switchable matrix, capable of connecting 10 high-speed coaxial or low-noise triaxial inputs to 48 high-speed, low-noise outputs.
Specific technical requirements include a maximum switch current rating of 1.0 A, a maximum voltage rating of 200 V channel to ground, and specialized input ports for precision measurements. The system must support low-leakage ports for I-V measurements, precision C-V measurement ports with minimal error, and high-bandwidth ports for transmission measurements. It must be compatible with NRL's existing Keysight B1500A semiconductor device analyzer, rack-mountable, capable of self-testing, and controlled through a GPIB interface.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Matrix Switch System CSS - Full and Open Competition.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Minimum specifications for Matrix Switch System
• a single Keysight Technologies part E5250A mainframe, with one year of support from the manufacturer and return-to-manufacturer repair warranty
• four each of part E5252A-FG matrix switch
• one each of part E5250A-301 Relay Function Test Adapter
• one each of part E5250A-1CM rack mount kit
This system satisfies several requirements that are specific to NRL’s application:
• can be configured to provide a 10x48 switchable matrix – in other words, can be controlled to connect one of 10 high speed coaxial or low noise triaxial inputs to one of
48 high speed, low noise outputs;
• maximum switch current rating 1.0 A;
• maximum voltage rating, channel to ground, 200 V;
• the 10 input ports described above to be distributed as at least two low-leakage (< 100 fA) ports for precision I-V measurements, at least two ports for precision C-V measurements with measurement errors less than 1% +/- 0.5 pF, and at least two high-bandwidth ports exhibiting less than 3 dB loss below 10 MHz for high frequency transmission measurements;
• can be controlled by commands specific to the Keysight E5250A mainframe unit in order to maintain compatibility and facile integration with an existing, NRL-owned Keysight
B1500A semiconductor device analyzer;
• Rack-mountable in order to optimize utilization of space in the laboratory;
• Capable of self-test as directed by software commands;
• Controlled through GPIB interface.
File details come from the government source that posted it. Updated .