Attachment 1 - FA700025Q0072 Gas Chromatography-Mass Spectrometer System Salient Characteristics.pdf
PDF 88 KB Posted
- Attached to
- USAFA Gas Chromatography-Mass Spectrometer System Federal contract opportunity
- Solicitation number
- FA700025Q0072
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| USAFA Gas Chromatography-Mass Spectrometer QandA.pdf | ||
| USAFA Gas Chromatography-Mass Spectrometer Combined Synopsis Solicitation.pdf | ||
| Attachment 3 - FA700025Q0072 252.225-7035, Buy American, Balance of Payments Program Certificate.pdf | ||
| Attachment 2-FA700025Q0072 Provisions and Clauses.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
USAFA DFCH Gas Chromatograph Mass Spectrometer (GC-MS) System Salient Characteristics
The system’s salient characteristics at minimum must include:
• Gas Chromatograph Mass Spectrometer (GC-MS) must have a gas chromatograph oven capable of reaching 450C.
• Must have a GC with an injector port temperature able to reach 450C.
• Mass Spectrometer must have at a minimum a mass range of 1.5 to 1000 m/z
• Gas chromatograph must have either a split/splitless or an MMI injection capability.
• Must include a vacuum pump.
• Must include an autosampler with a minimum SO 1.5 ml vial sample tray.
• Mass Spectrometer must include a high-performance quadrupole mass filter.
• GCMS must be set up for hydrogen as the carrier gas.
• Must have a mass resolution of 2M at full width at half maximum.
• Must have a full scan and selective ion monitoring (SIM) modes.
• Must include GC-MS software that works with Agilent Technologies software.
• 1-year warranty.
• Must provide on-site training and installation.
• The ion source must be removable without removing the filaments or the lens.
• When running a batch or multiple samples, different methods can run within the batch.
• An Auto Sampler with slots between 15-24.
• Must have a Desktop PC, printer included, and instrument software.