About this file
E-DNA Auto Sampler Requirements and Attachments
This document is a requirements specification for autonomous environmental sampling equipment. The procurement calls for two (2) autonomous e-DNA auto sampler units capable of collecting water samples for environmental DNA analysis in freshwater environments. The units must feature modular systems supporting filtration, preservation, and water collection with autonomous and/or tethered operation modes. Each sampler shall have a user-friendly programming interface and integrated Radio Frequency Identification (RFID) or equivalent traceability system with Bluetooth and/or satellite communication capability. All components must be compatible with standard sterilization protocols (autoclave or bleach).
The samplers must meet specific performance parameters: maximum eight (8) samples per deployment with operable filter pore size ≥5.0µm; operating temperature range of 0–40°C; lift capability 0–2m above water surface; water-resistant enclosure rating (IP 43); flow rates between 0.1–1.4L/min; and volume measurement accuracy ≥95%. GPS compatibility must support GNSS standards (GPS, Galileo, GLONASS, BeiDou) with ~2.5M position accuracy. The physical design shall be compact (24–30 in length, 17–24 in width, 12–16 in height) and weigh ≤100 lb for remote deployments. Power specifications include 12.8V 22.5Ah Lithium Iron Phosphate rechargeable batteries with minimum 3+ hour pumping runtime and ~30-day maximum deployment duration on battery alone, plus optional solar panel charging capability. Deliverables include four (4) total batteries (two per unit), two (2) chargers, two (2) solar panel chargers (if available), intake hoses, connectors, strainers, sterilization filters, two (2) custom carrying cases, and shipping straps. A manufacturer's standard warranty with minimum one-year parts and labor coverage is required.
View the file
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
E-DNA AUTO SAMPLER REQUIREMENTS AND ATTACHMENTS
Functional Requirements
• Modular system capable of filtration, preservation, and/or water collection.
• Autonomous and/or tethered operation with user-friendly programming interface.
• Ability to collect multiple independent samples and replicates per deployment.
• Integrated Radio Frequency Identification (RFID) or equivalent system for sample traceability. Bluetooth and/or satellite capability.
• Compatible with sterilization protocols (autoclave or bleach).
Performance Specifications
• Max. number of samples ~8
• Operable filter pore size ≥5.0µm recommended
• Operable interior temperature 0–40°C
• Operable lift (unit elevation above water)0–2m
• Enclosure rating water-resistant (~IP 43)
• Freshwater environmental sampling sensors
• Minimum flow rate ~0.1L/min
• Max. flow rate ~1.4L/min
• Max. measurable volume per sample ~1,000L
• Volume accuracy ≥95%
• GPS system compatibility GNSS (GPS, Galileo, GLONASS, BeiDou)
• GPS position accuracy ~2.5M
• Pump
• Ports ~0.5 in OD Push-to-connect
• Tubing 0.5 in OD, 0.25 in ID Polyurethane tubing 95 A durometer
• Priming (dry) vacuum approx. 10–12 PSI
• Max. wetted vacuum approx. 10–12 PSI
Dimensions & Weight
• Compact design suitable for transport and remote deployments.
• Size and Weight:
o Length 24–30 in (~760 mm) o Width 17–24 in (~610 mm) o Height 12–16 in (304–406 mm) o Weight ≤100 lb (≤45.35 kg)
Power System
• Swappable battery packs for autonomous operation.
• Minimum runtime: 3+ hours pumping, weeks in sensing mode.
• Low-power indicator and quiescent mode for extended deployments.
• Solar panel for battery recharging if available.
• Power:
o Sampler battery ~12.8V 22.5Ah Lithium Iron Phosphate o Max. deployment duration on battery ~30 days o AC power 85–264V 48–63Hz
Sensors & Data
• Integrated sensors for contextual data (e.g., GPS, CTD, DO).
• Wireless or tethered command and data interface.
• Data logging capability paired with each sample.
Accessories
Each unit shall include:
• Two (2) rechargeable batteries per unit.
• Battery chargers.
• Intake hoses, connectors, and strainers.
• Filters for sterilization.
• Custom carrying cases.
• Shipping straps.
Operational Environment
• Designed for rugged field conditions, including rough handling and variable temperatures.
• Corrosion-resistant components for prolonged water exposure.
Warranty
• Manufacturer’s standard warranty.
• Minimum coverage: 1-year parts and labor.
Attachments
• Two (2) autonomous e-DNA auto sampler units.
• Four (4) batteries (two per unit).
• Two solar panel chargers (if available)
• Two (2) chargers.
• Intake hoses, connectors, strainers.
• Filters for sterilization.
• Two (2) custom carrying cases.
• Shipping straps.