Attachment 1 Chapter 269.doc
DOC document 28 KB Posted
- Attached to
- Evidential Breath Alcohol Measurement Device
- Solicitation number
- RFQ - 16A - 12062700000000001125
- Issued by
- Kennebec County, Maine
About this file
This document is Chapter 269 of Maine state regulations pertaining to evidential breath alcohol measurement devices. The file establishes the technical and operational standards that govern the approval, certification, and use of breath alcohol testing equipment by law enforcement agencies and other authorized entities throughout Maine. The regulations detail the requirements for device manufacturers, calibration procedures, operator qualifications, and quality assurance protocols necessary for breath alcohol measurements to be admissible as evidence in legal proceedings. The standards address device accuracy tolerances, maintenance schedules, training requirements for operators, and documentation procedures to ensure the reliability and validity of test results.
The regulations specify performance specifications that evidential breath alcohol devices must meet, including accuracy within 0.005 percent blood alcohol concentration, acceptable operating temperature ranges, and proper response times to breathalyzer samples. The document outlines certification procedures that manufacturers must follow to gain approval for their equipment in Maine, including submission of technical specifications, performance data, and compliance documentation. Additionally, the regulations establish ongoing requirements for device maintenance, periodic recalibration by certified technicians, and operator training and certification to maintain accuracy and legal defensibility of test results in court proceedings.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment 2 Sample Test Results.pdf | ||
| BPO General Terms Conditions (3-7-12).doc | DOC document | |
| Solicitation_terms.doc | DOC document | |
| Specifications for alcohol breath testing instrument 2012.doc | DOC document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
ATTACHMENT ONE
10-144 Chapter 269 page 2
10-144
DEPARTMENT OF HEALTH AND HUMAN SERVICES
MAINE CDC, HEALTH AND ENVIRONMENTAL TSTING LABORATORY
Chapter 269:
RULES GOVERNING SELF-CONTAINED BREATH ALCOHOL TESTING EQUIPMENT
SUMMARY: All self-contained breath alcohol testing equipment must be approved by the U.S. Department of Transportation as stated in the Federal Register and the State of Maine, Department of Health and Human Services. Each instrument must be tested and approved by the Public Health Laboratory and retested and re-approved semi-annually. Certain procedures are specified for calibration checks and use of self-contained breath alcohol testing equipment.
SECTION 1.
EQUIPMENT
1.
Only those instruments approved by the U.S. Department of Transportation for the purpose of breath testing will be considered. Evidence of this approval must be submitted by the manufacturer. An approved simulator must be provided for use with each instrument.
2.
The accuracy and sensitivity of the equipment should be such as to obtain results within ± 0.01 g/210L or 5%, whichever is greater of the known value in the analysis of appropriate reference materials of known ethyl alcohol concentrations.
3.
Before approval each instrument must be tested by a chemist of the Health and Environmental Testing Laboratory (HETL). Approval will be given provided the machine gives results accurate within the limits of the performance requirements of the Department mentioned in Paragraph B, and will be indicated by affixing to the instrument a stamp which will be valid for no more than seven months.
4.
Each instrument will be retested by a chemist of the HETL at least once semi-annually. A new stamp of approval will be affixed to the instrument with the test date placed thereon.
5.
Failure of an instrument to provide results accurate within the limits of the performance requirements of the Department (1.B), when detected, will be investigated by a trained operator or a chemist of the HETL to determine the cause of that failure. If the results of that investigation establish that the instrument itself is out of calibration, or non-functional, that will be cause for immediate withdrawal of approval and removal of the stamp of approval previously affixed.
SECTION 2.
PROCEDURES
1.
A calibration check must be run for each subject tested.
2.
For each person tested, a complete breath-alcohol test must consist of 2 separate breath samples which result in determinations of breath-alcohol concentration which agree within ± 0.02g/210L.
3.
If the first 2 breath sample results on the subject do not agree within ± 0.02g/210L, subsequent samples must be taken until 2 tests fall within the prescribed limits. If after 4 separate breath sample results are taken, no 2 breath sample results agree within the prescribed limits, the testing sequence shall be void and either a retest or an alternative procedure shall be required.
4.
The two lowest results which agree within ± 0.02 g/210L will be averaged, reporting only the first two decimal places of the average result as the final breath alcohol concentration.
5.
The Health and Environmental Testing Laboratory will provide any alcohol solution required by each agency for simulator tests.
STATUTORY AUTHORITY: 29-A M.R.S.A. §2524(6); and 22 M.R.S.A. §42(1); 17-A M.R.S.A. §1057 and 22-A M.R.S.A. §205(2)
EFFECTIVE DATE:
November 15, 1978
AMENDED:
August 1, 1982
October 17, 1988
EFFECTIVE DATE (ELECTRONIC CONVERSION):
May 5, 1996
AMENDED:
December 6, 2004 – filing 2004-553
September 1, 2010 – filing 2010-370