This document is an Instrumentation Test Plan for the Radio Test Set, Model TS-4549/T manufactured by Astronics. The test plan outlines the procedures and equipment required to verify the performance of the TS-4549/T to the requirements specified in the MIS-PRF-60487B Performance Specification. Key areas tested include radio frequency (RF) signal generation, audio frequency (AF) generation, frequency and level measurement, modulation, and various other instrumentation capabilities. The test plan indicates that the government intends to operationally test a prototype model of the TS-4549/T with four application program sets (APSs) after completion of an Operational Test and Readiness Review. The government will require the contractor to provide the intellectual property, including source code and radio-specific information, for the APSs developed for the TS-4549/T.
View the file
Other files for this federal contract opportunity
Show all 19
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
31 January 2022 Page 1 of 34
Instrumentation Test Plan
For
Test Set, Radio: TS-4549/T( )
Manufacturer: Astronics
Model: ATS3100
Serial Number: 21080005
Army Enterprise Support Directorate Electrical Measurements Division, U.S. Army TMDE Activity Building 5437 (ATTN: AMAM-TML-EM) Redstone Arsenal, Alabama 35898-5000
TML-EM, USATA
REDSTONE ARSENAL, ALABAMA 35898-5000
31 January 2022 Page 2 of 34
TABLE OF CONTENTS
1 INTRODUCTION 3
2 EQUIPMENT REQUIREMENTS 4
2.1 SYSTEM REQUIREMENTS. 4
3 EQUIPMENT USED IN TESTING 4
4 TEST PLAN/PROCEDURE 4
4.1 RADIO FREQUENCY (RF) SIGNAL GENERATOR (TRANSMIT/RECEIVE (T/R) AND DUPLEX CONNECTOR. 4
4.2 AUDIO FREQUENCY (AF) GENERATOR #1 AND #2 AND BOOST MODE. 11
4.3 AUDIO FREQUENCY COUNTER. 14
4.4 RADIO FREQUENCY COUNTER. 15
4.5 RADIO FREQUENCY ERROR METER. 16
4.6 RF POWER METER. 16
4.7 RF RECEIVER. 18
4.8 DEVIATION METER. 19
4.9 AM MODULATION METER. 20
4.10 SINGLE SIDE BAND. 21
4.11 PM DEVIATION METER. 22
4.12 DISTORTION METER. 23
4.13 SINAD METER. 24
4.14 AF LEVEL METER. 25
4.15 DIGITAL MULTIMETER. 26
4.16 OSCILLOSCOPE. 28
4.17 SPECTRUM ANALYZER. 29
4.18 BER METER. 32
4.19 INTERNAL REFERENCE OSCILLATOR. 33
4.20 DIGITAL DATA GENERATOR. 34
31 January 2022 Page 3 of 34
1 Introduction
The TS-4549/T shall be a single test set to test all current and future Army radios. Once a radio is connected to the TS-4549/T it shall automatically and manually, via the APSs, step through a comprehensive series of test parameters and compare the radio’s response to the known limits of the test parameters. All aberrations of limits shall be flagged and reported to the user. The user shall then be able to use the test set to perform operational screening tests, troubleshoot and repair the radio as recommended by the Radio OEM/Army maintenance philosophy.
At this point in time Product Director Test Measurement and Diagnostic Equipment (PD-TMDE) does not possess any of the Intellectual Property (IP), such as Application Source Code and radio specific information for APSs that runs on the AN/GRM-122. At the time of APSs (historically called TPS) creation, the Government only required the executable code. As such, PD-TMDE has no IP to use in generating new APSs for the development of the TS-4549/T Radio Test Set. The selected Contractor(s) shall be responsible for generating a new Prototype Radio Test Set with four (4) APSs. The Contractor(s) shall negotiate with the Radio Original Equipment Manufacturers (OEMs) to acquire the IP they need. PD-TMDE will also require the Contractor(s) to submit Government Purpose Rights, IP along with the APSs, development software and compilers for all radios they are required to develop for archival purposes to be used when the TS-4549/T goes obsolete and is replaced. For future radios, the program offices for those radios will be required to submit the same data for radios they want tested on the TS-4549/T.
Two APSs shall demonstrate the ability to support the SINCGARS family of radios and the remaining two APSs shall demonstrate the ability to support Software Defined Radios (SDRs). Vendors shall provide Executable Application Program Software with each Prototype TS-4549/T Radio Test Set. Other required deliverables are listed in Section 4.0. PD-TMDE will require three months of testing to conduct an assessment of the prototypes (to include the APSs) and Executable Application Program Software using representative fielded radios.
The Government intends on operationally testing this Prototype Representative Model as Version 1.0 with the four (4) APSs after the completion of the Operational Test and Readiness Review (OTRR). This assessment will utilize User Facility of Use (UFOU) evaluation in conjunction with CASCOM at Fort Gordon. The operational & calibration test will require the United States Army TMDE Activity (USATA) to conduct a validation and verification of the TS- 4549/T Radio Test Set’s instrumentation ability to meet the requirements outlined in accordance with Performance Specification, MIS-PRF-60487A during Operational and Calibration Testing.
31 January 2022 Page 4 of 34
2 Equipment Requirements
2.1 System Requirements.
The full system requirements may be found in the MIS-PRF-60487B Performance Specification/Item Specification for the Test Set, Radio: TS-4549/T( ) (dated 26 June 2020). The performance requirements for the TS-4549/T( ) are also shown throughout this document in association with each test.
3 Equipment Used in Testing
Nomenclature Mfr/Model S/N Cal Due
Audio Analyzer Boonton, 1121 76301 25 April 2022 Function
Generator Agilent, 33250A US40001959
7 June 2023 Calibrator Fluke 5720A CT5230005
23 January 2023 Measuring
Receiver Keysight, N5531S SG48250159
22 January 2023 Multimeter Fluke, 8846A 1985003
27 February 2023 RF Reference
Source Fluke 96040-LPN 928353742
29 May 2022 Signal
Generator
MG3694C 114113
17 July 2022
4 Test Plan/Procedure
This section describes the test plans/procedures along with the associated specifications that are needed to verify the requirements specified in MIS-PRF-60487B Performance Specification/Item Specification for the Test Set, Radio System: TS-4549/T( ) (dated 26 June 2020). Bid samples will be evaluated for conformance to the requirements specified in section 3.0, steps 3.1.1.1 to 3.1.20.3 of the TS-4549 Specification. All Bid Sample performance testing will be conducted at room temperature.
4.1 Radio Frequency (RF) Signal Generator (Transmit/Receive (T/R) and Duplex Connector.
Government Specification:
31 January 2022 Page 5 of 34
TS-4549 Specification Specification Requirement
3.1.1.1 Frequency The TS-4549 shall meet frequency generation in accordance with (IAW) requirements of paragraph 3.1.1.1.1 through 3.1.1.1.3 of
MIL-PRF-60487B
3.1.1.2 Level The TS-4549 shall meet level generation IAW requirements of paragraph 3.1.1.2.1 through 3.1.1.2.6 of MIL-PRF-60487B
3.1.1.3 Spectral Purity The TS-4549 shall meet spectral purity of radio frequency signals IAW requirements of paragraph 3.1.1.3.1 through 3.1.1.3.5 of
MIL-PRF-60487B
3.1.1.4 Internal
Modulation – Frequency Modulation (FM)
The TS-4549 shall meet the capability of internal frequency modulation - FM IAW requirements of paragraph 3.1.1.4.1 through 3.1.1.4.7 of MIL-PRF-60487B
3.1.1.5 Internal
Modulation – Amplitude Modulation (AM)
The TS-4549 shall meet the capability of internal amplitude modulation - AM IAW requirements of paragraph 3.1.1.5.1 through 3.1.1.5.7 of MIL-PRF-60487B
3.1.1.6 Internal Phase
Modulation (PM)
The TS-4549 shall meet the capability of internal phase modulation - PM IAW requirements of paragraph 3.1.1.6.1 through 3.1.1.6.5 of MIL-PRF-60487B
3.1.1.7 Internal Digital
Modulation (DM)
The TS-4549 shall meet the capability of internal digital modulation - DM IAW requirements of paragraph 3.1.1.7.1 through 3.1.1.7.5 of MIL-PRF-60487B
3.1.1.8 External
Modulation
The TS-4549 shall meet the capability of external modulation - IAW requirements of paragraph 3.1.1.8 of MIL-PRF-60487B
3.1.1.9 Frequency
Agility (For Both RF Generator and Receiver)
The TS-4549 shall meet the capability of frequency agility - IAW requirements of paragraph 3.1.1.9.1 of MIL-PRF-60487B
Equipment Required:
Bid Sample Unit
Measuring Receiver Audio Analyzer Power Splitter
Test Plan/Procedure Radio Frequency (RF) Signal Generator
Frequency Test Point (MHz) Pass/Fail
10.000000 Pass
50.000000 Pass
31 January 2022 Page 6 of 34
100.00000 Pass
500.00000 Pass
1000.0000 Pass
1250.0000 Pass
1500.0000 Pass
1750.0000 Pass
2000.0000 Pass
2250.0000 Pass
2500.0000 Pass
2800.0000 Pass
Test Plan/Procedure Radio Frequency (RF) Signal Generator
Level Test Point
Frequency (MHz) Level (dBm) Pass/Fail
1 0.0 Pass 1 -15.0 Pass 1 -16.0 Pass
3.5 0.0 Pass
3.5 -39.0 Pass
3.5 -40.0 Pass
3.5 -63.0 Pass
3.5 -79.0 Pass
3.5 -110.0 Pass
3.5 -127.0 Pass
30.65 0.0 Pass
30.65 -1.0 Pass
30.65 -2.0 Pass
30.65 -4.0 Pass
30.65 -8.0 Pass
30.65 -16.0 Pass
30.65 -32.0 Pass
30.65 -39.0 Pass
30.65 -63.0 Pass
30.65 -64.0 Pass
30.65 -79.0 Pass
30.65 -110.0 Pass
30.65 -127.0 Pass
65.020 0.0 Pass
65.020 -39.0 Pass
65.020 -40.0 Pass
65.020 -63.0 Pass
65.020 -79.0 Pass
31 January 2022 Page 7 of 34
65.020 -110.0 Pass
65.020 -127.0 Pass
75.020 0.0 Pass
75.020 -39.0 Pass
75.020 -40.0 Pass
75.020 -63.0 Pass
75.020 -79.0 Pass
75.020 -110..0 Pass
75.020 -127.0 Pass
150.020 0.0 Pass
150.020 -39.0 Pass
150.020 -40.0 Pass
150.020 -63.0 Pass
150.020 -79.0 Pass
150.020 -110.0 Pass
150.020 -127.0 Pass
300.020 0.0 Pass
300.020 -39.0 Pass
300.020 -40.0 Pass
300.020 -63.0 Pass
300.020 -79.0 Pass
300.020 -110.0 Pass
300.020 -127.0 Pass
500.020 0.0 Pass
500.020 -39.0 Pass
500.020 -40.0 Pass
500.020 -63.0 Pass
500.020 -79.0 Pass
500.020 -110.0 Pass
500.020 -127.0 Pass
750.020 0.0 Pass
750.020 -39.0 Pass
750.020 -40.0 Pass
750.020 -63.0 Pass
750.020 -79.0 Pass
750.020 -110.0 Pass
750.020 -127.0 Pass
999.020 0.0 Pass
999.020 -39.0 Pass
999.020 -40.0 Pass
999.020 -63.0 Pass
999.020 -79.0 Pass
999.020 -110.0 Pass
999.020 -127.0 Pass
31 January 2022 Page 8 of 34
Radio Frequency (RF) Signal Generator
Spectral Purity RF Generator Harmonics Frequency (MHz)
RF Generator Non- Harmonics Frequency (MHz)
Pass/Fail
1 Pass 10 Pass 100 Pass 200 Pass 300 Pass 400 Pass 500 Pass 600 Pass 700 Pass 800 Pass 900 Pass
1000 Pass 1200 Pass 1400 Pass 1600 Pass 1800 Pass 2000 Pass 2200 Pass 2400 Pass 2600 Pass 2800 Pass
1 Pass 10 Pass 100 Pass 200 Pass 300 Pass 400 Pass 500 Pass 600 Pass 700 Pass 800 Pass 900 Pass 1000 Pass 1200 Pass 1400 Pass 1600 Pass 1800 Pass 2000 Pass 2200 Pass
31 January 2022 Page 9 of 34
2400 Pass 2600 Pass 2800 Pass
Test Plan/Procedure Radio Frequency (RF) Signal Generator
Internal Modulation - FM Test Point (MHz)
Frequency (Hz) Modulation Rate (Hz) Deviation (Hz) Pass/Fail
750 k 30 100 Pass 750 k 30 10k Pass 750 k 30 25k Pass 750 k 1k 100 Pass 750 k 1k 10k Pass 750 k 1k 25k Pass 750 k 10k 10k Pass 750 k 10k 25k Pass 900 M 30 100 Pass 900 M 30 10k Pass 900 M 30 25k Pass 900 M 1k 100 Pass 900 M 1k 10k Pass 900 M 1k 25k Pass 900 M 20k 10k Pass 900 M 20k 25k Pass
Test Plan/Procedure Radio Frequency (RF) Signal Generator
Internal Modulation - AM Test Point
RF Frequency (Hz) Modulation Rate (Hz) Modulation (%) Pass/Fail
1M 100 30 Pass 1M 100 60 Pass 1M 100 90 Pass 1M 1k 30 Pass 1M 1k 60 Pass 1M 1k 90 Pass 1M 10k 30 Pass 1M 10k 60 Pass 1M 10k 90 Pass 10M 100 30 Pass 10M 100 60 Pass
31 January 2022 Page 10 of 34
10M 100 90 Pass 10M 1k 30 Pass 10M 1k 60 Pass 10M 1k 90 Pass 10M 10k 30 Pass 10M 10k 60 Pass 10M 10k 90 Pass 900M 100 30 Pass 900M 100 60 Pass 900M 100 90 Pass 900M 1k 30 Pass 900M 1k 60 Pass 900M 1k 90 Pass 900M 10k 30 Pass 900M 10k 60 Pass 900M 10k 90 Pass
Test Plan/Procedure Radio Frequency (RF) Signal Generator
Phase Modulation Test Point
Carrier Frequency (MHz) PM Deviation Pass/Fail
10 3 Pass 100 5 Pass 1000 7 Pass 2000 10 Pass
Test Plan/Procedure Radio Frequency (RF) Signal Generator
Internal Digital Modulation Test Point Pass/Fail Not Tested Not Tested
Test Plan/Procedure Radio Frequency (RF) Signal Generator
External Modulation Test Point Pass/Fail Same as Internal Modulation Pass
31 January 2022 Page 11 of 34
The US Army does not possess the capability to calibrate/test “Digital modulation”.
Addition of this requirement will have to be addressed regardless of outcome of other tests in the instrumentation test plan. This parameter is not critical to down-select, and it shall be addressed at contract award.
4.2 Audio Frequency (AF) Generator #1 and #2 and Boost Mode.
TS-4549 Specification Specification Requirement
3.1.2.1 Frequency The TS-4549 shall meet the capability of AF generated signal frequency - IAW requirements of paragraph 3.1.2.1.1 through
3.1.2.1.3 of MIL-PRF-60487B
3.1.2.2 Alternating
Current (AC) Level (Sine Wave)
The TS-4549 shall meet the capability of AC levels to a AF generated signal - IAW requirements of paragraph 3.1.2.2.1 through 3.1.2.2.3 of MIL-PRF-60487B
3.1.2.3 Direct Current
(DC) Level
The TS-4549 shall meet the capability of DC levels to an AF generated signal - IAW requirements of paragraph 3.1.2.3.1 through 3.1.2.3.2 of MIL-PRF-60487B
3.1.2.4 Boost The TS-4549 shall meet the capability of level boost to an AF signal - IAW requirements of paragraph 3.1.2.4.1 through
3.1.2.4.3 of MIL-PRF-60487B
3.1.2.5 Spectral Purity The TS-4549 shall meet the capability of AF generator spectral purity IAW requirements of paragraph 3.1.2.5.1 through
3.1.2.5.2 of MIL-PRF-60487B
3.1.2.6 Digital Data
Generator and Pattern Generator for BER
The TS-4549 shall meet the capability of AF digital data generator and pattern generation IAW requirements of paragraph
3.1.2.6.1 through 3.1.2.6.7 of MIL-PRF-60487B
Radio Frequency (RF) Signal Generator
Frequency Agility Test Point (MHz) Pass/Fail
10.000000 Pass
10.025000 Pass
10.050000 Pass
31 January 2022 Page 12 of 34
Audio Frequency Generator
Frequency Test Point (Hz) Pass/Fail
210.0 Pass
210.1 Pass
500.0 Pass
1.000k Pass 5.000k Pass
10.000 k Pass
15.000 k Pass
20.000 k Pass
23.999 k Pass
24.000 k Pass
Test Plan/Procedure Audio Frequency Generator and Boost Mode
AC Level Test Point
Level Audio Frequency Pass/Fail
0.7 210 Pass
1.0 210 Pass
1.5 210 Pass
2.5 210 Pass
2.5 1 k Pass
1.5 1 k Pass
1.0 1 k Pass
0.7 1 k Pass
0.7 10 k Pass
1.0 10 k Pass
1.5 10 k Pass
2.5 10 k Pass
2.5 24 k Pass
1.5 24 k Pass
1.0 24 k Pass
0.7 24 k Pass
31 January 2022 Page 13 of 34
Boost mode was not tested due to concurrence by the government; it possesses no impact on the capability of the device. Boost mode is an amplification of a signal that is already tested, and proven in the instrumentation test plan.
The US Army does not possess the capability to calibrate/test “Digital modulation”.
Addition of this requirement will have to be addressed regardless of outcome of
Audio Frequency Generator and Boost Mode DC Level
Test Point Level Audio Frequency DC Offset
Pass/Fail
.7 210 10m Pass
1.5 1 k 10m Pass
2.5 24 k 10m Pass
Test Plan/Procedure Audio Frequency Generator and Boost Mode
Boost Mode Test Point Pass/Fail Not Tested Not Tested
Test Plan/Procedure Audio Frequency Generator and Boost Mode
Spectral Purity Test Point (Hz) Pass/Fail 210 Pass 1k Pass 10k Pass 20k Pass 24k Pass
Test Plan/Procedure Audio Frequency Generator and Boost Mode
Digital Data Generator and Pattern Generator for BER
Not Tested Not Tested
31 January 2022 Page 14 of 34
4.3 Audio Frequency Counter.
Government Specification:
TS-4549 Specification Specification Requirement
3.1.3.1 Frequency The TS-4549 shall meet the capability of AF signal frequency counter - IAW requirements of paragraph 3.1.3.1.1 through
3.1.3.1.4 of MIL-PRF-60487B
3.1.3.2 Signal
Characteristics
The TS-4549 shall meet the capability of measuring signal characteristics of an AF signal IAW requirements of paragraph
3.1.3.2.1 through 3.1.3.2.2 of MIL-PRF-60487B
RF Reference Source
Audio Frequency Counter
Frequency Frequency (kHz) Pass/Fail .10000 Pass
1.0000 Pass
10.000 Pass
20.000 Pass
30.000 Pass
40.000 Pass
Test Plan/Procedure Audio Frequency Counter
Level Test Point
Frequency (kHz) Level Vrms Pass/Fail
10 1 Pass 10 2 Pass 10 3 Pass
31 January 2022 Page 15 of 34
4.4 Radio Frequency Counter.
TS-4549 Specification Specification Requirement
3.1.4.1 Frequency The TS-4549 shall meet the capability of RF signal frequency counter - IAW requirements of paragraph 3.1.4.1.1 through
3.1.4.1.3 of MIL-PRF-60487B
3.1.4.2 Level The TS-4549 shall meet the capability of RF signal levels IAW requirements of paragraph 3.1.4.2.1 of MIL-PRF-60487B
Measuring Receiver Function Generator
Radio Frequency Counter
Frequency Test Point (MHz) Pass/Fail
1.000000 Pass
5.000000 Pass
10.00000 Pass
50.00000 Pass
100.0000 Pass
500.0000 Pass
1000.000 Pass
1500.000 Pass
2000.000 Pass
2500.000 Pass
2800.000 Pass
Test Plan/Procedure Radio Frequency Counter
Level Test Point
Frequency (MHz) Level (dBm) Pass/Fail
10.00000 -10 Pass
31 January 2022 Page 16 of 34
4.5 Radio Frequency Error Meter.
TS-4549 Specification Specification Requirement
3.1.5.1 Frequency The TS-4549 shall meet the capability of determining RF frequency error - IAW requirements of paragraph 3.1.5.1.1 through 3.1.5.1.5 of MIL-PRF-60487B
Measuring Receiver Function Generator
4.6 RF Power Meter.
TS-4549 Specification Specification Requirement
50.00000 -20 Pass
100.0000 -30 Pass
250.0000 -40 Pass
500.0000 -50 Pass
1000.000 -10 Pass
1500.000 -20 Pass
2000.000 -30 Pass
2500.000 -40 Pass
Test Plan/Procedure Radio Frequency Error Meter
Frequency Test Point (MHz) Pass/Fail
10.000 Pass
100.000 Pass
1000.000 Pass
1500.000 Pass
2000.000 Pass
2500.000 Pass
31 January 2022 Page 17 of 34
3.1.6.1 Frequency The TS-4549 shall meet the capability of RF power measurements frequency requirements of paragraph 3.1.6.1.1 of
MIL-PRF-60487B
3.1.6.2 Level The TS-4549 shall meet the capability of RF power measurements level IAW requirements of paragraph 3.1.6.2.1 through 3.1.6.2.4 of MIL-PRF-60487B
3.1.6.4 Display
Measurement Type
The TS-4549 shall meet the capability of displaying multiple RF measurement types IAW requirements of paragraph 3.1.6.4 of
MIL-PRF-60487B
RF Reference Source RF Power Meter Low Pass Filter RF Power Amplifier
RF Power Meter Frequency/Level
Test Point Range(mW) Amplitude (W) Frequency(MHz)
Pass/Fail
20 400.00 u 82.000000 Pass 20 400.00 u 960.00000 Pass 20 5.0000 m 82.000000 Pass 20 15.000 m 82.000000 Pass 200 40.000 m 100.0000 Pass 200 150.00 m 250.0000 Pass 200 150.00 m 750.0000 Pass 200 150.00 m 1500.000 Pass 200 150.00 m 2000.000 Pass 200 150.00 m 2500.000 Pass 200 100.00 m 2500.000 Pass 200 100.00 m 1500.000 Pass 200 100.00 m 500.0000 Pass 2,000 1.5 1000.000 Pass 20,000 15.0 1500.000 Pass 200,000 45.0 2000.000 Pass
Test Plan/Procedure RF Power Meter
Display Measurement Type
W Pass
31 January 2022 Page 18 of 34
4.7 RF Receiver.
TS-4549 Specification Specification Requirement
3.1.7.1 Frequency
Range
The TS-4549 shall meet the capability of receiving RF frequencies over a given range IAW requirements of paragraph
3.1.7.1 MIL-PRF-60487B
3.1.7.2 Sensitivity The TS-4549 shall meet the capability of demodulating an RF signal with the approved sensitivity IAW requirements of paragraph 3.1.7.2 of MIL-PRF-60487B
3.1.7.3 Selectivity The TS-4549 shall meet the capability of demodulating an RF signal with the approved selectivity IAW requirements of paragraph 3.1.7.3 of MIL-PRF-60487B
3.1.7.4 Demodulation
Output
The TS-4549 shall meet the capability of demodulating an RF signal IAW requirements of paragraph 3.1.7.4 of MIL-PRF- 60487B
3.1.7.5 Quieted
Frequencies
The TS-4549 shall meet the capability of reduction of a signal IAW requirements of paragraph 3.1.7.5 of MIL-PRF-60487B
N/A
Section 4.7 RF Receiver will be tested by the tests performed in test 4.17 Spectrum Analyzer.
V Pass dBm Pass
31 January 2022 Page 19 of 34
4.8 Deviation Meter.
TS-4549 Specification Specification Requirement
3.1.8.1 Deviation Range The TS-4549 shall meet the capability of measuring the deviation of a signal IAW requirements of paragraph 3.1.8.1.1 through 3.1.8.1.3 of MIL-PRF-60487B
3.1.8.2 Modulation
Rate
The TS-4549 shall meet the capability of measuring the modulation rate IAW requirements of paragraph 3.1.8.2 of MIL-
PRF-60487B
3.1.8.3 Carrier Range The TS-4549 shall meet the capability of measuring a signals frequency across its full range IAW requirements of paragraph
3.1.8.3 of MIL-PRF-60487B
3.1.8.4 Carrier Level The TS-4549 shall meet the capability of measuring a signals level across its range IAW requirements of paragraph 3.1.8.4 of
MIL-PRF-60487B
RF Reference Source
Measuring Receiver
Deviation Meter
Test Point Carrier (MHz) Carrier Level(dBm) Mod Rate (Hz) Dev (kHz)
Pass/Fail
10 -10 100 1 Pass 10 -20 100 1 Pass 10 -30 100 100 Pass 10 -40 100 100 Pass 10 -50 2k 100 Pass 10 -40 2k 100 Pass 10 -30 2k 45 Pass 10 -20 2k 45 Pass 10 -10 5k 10 Pass
31 January 2022 Page 20 of 34
4.9 AM Modulation Meter.
TS-4549 Specification Specification Requirement
3.1.9.1 Modulation The TS-4549 shall meet the capability of modulation a signal with amplitude modulation IAW requirements of paragraph
3.1.9.1.1 through 3.1.9.1.4 of MIL-PRF-60487B
3.1.9.2 Carrier Range The TS-4549 shall meet the capability of modulating a signal with amplitude modulation over the full carrier range IAW requirements of paragraph 3.1.9.2 of MIL-PRF-60487B
3.1.9.3 Carrier Level The TS-4549 shall meet the capability of modulating a signal with amplitude modulation with its full level IAW requirements of paragraph 3.1.9.3 of MIL-PRF-60487B
50 -10 5k 10 Pass 50 -20 5k 1 Pass 50 -30 5k 1 Pass 50 -40 14k 100 Pass 100 -50 14k 100 Pass 100 -40 14k 45 Pass 100 -30 14k 45 Pass 100 -20 14k 10 Pass 100 -10 14k 10 Pass 300 -10 25k 10 Pass 300 -20 25k 10 Pass 300 -30 25k 45 Pass 300 -40 25k 45 Pass 500 -50 25k 100 Pass 500 -40 30k 100 Pass 500 -30 30k 45 Pass 500 -20 30k 10 Pass 500 -10 30k 10 Pass 1000 -10 30k 10 Pass 1000 -20 30k 10 Pass 1000 -30 40k 45 Pass 1000 -40 40k 45 Pass 1000 -50 40k 100 Pass 1000 -40 40k 100 Pass 1000 -30 40k 100 Pass
31 January 2022 Page 21 of 34
4.10 Single Side Band.
TS-4549 Specification Specification Requirement
3.1.10 Phase Noise The TS-4549 shall meet the capability of measuring single side band phase noise IAW requirements of paragraph 3.1.10 of MIL-
PRF-60487B
Equipment Required:
Bid Sample Unit
RF Reference Source
AM Modulation Meter
Test Point Mod Rate (Hz) Modulation (%)
Pass/Fail
1k 30 Pass 1k 40 Pass 1k 50 Pass 1k 60 Pass 1k 70 Pass 1k 80 Pass 1k 90 Pass 100 30 Pass 100 90 Pass 10k 90 Pass 10k 30 Pass
Test Plan/Procedure Single Side Band
Modulation Carrier Frequency Offset dBc Pass/Fail 1 GHz 1 kHz -90 Pass
31 January 2022 Page 22 of 34
4.11 PM Deviation Meter.
TS-4549 Specification Specification Requirement
3.1.11.1 Deviation The TS-4549 shall meet the capability of measuring deviation of phase modulation IAW requirements of paragraph 3.1.11.1.1 through 3.1.11.1.4 of MIL-PRF-60487B
3.1.11.2 Modulation
Rate
The TS-4549 shall meet the capability of measuring modulation rate of a phase modulated signal IAW requirements of paragraph
3.1.11.2 of MIL-PRF-60487B
3.1.11.3 Carrier The TS-4549 shall meet the capability of measuring modulation of a phase modulated signal over its full range IAW requirements of paragraph 3.1.11.3.1 through 3.1.11.3.2 of MIL-PRF-60487B
Equipment Required:
Bid Sample Unit
RF Reference Source
PM Deviation Meter
Modulation Test Point
Modulation Rate (kHz) Carrier (MHz) Deviation (rad) Pass/Fail
3 10.000 1 Pass 3 100.00 3 Pass 3 1000.00 7 Pass 3 1000.00 9 Pass
31 January 2022 Page 23 of 34
4.12 Distortion Meter.
TS-4549 Specification Specification Requirement
3.1.12.1 Distortion The TS-4549 shall meet the capability of measuring distortion IAW requirements of paragraph 3.1.12.1.1 through 3.1.2.1.4 of
MIL-PRF-60487B
3.1.12.2 Signal
Characteristics
The TS-4549 shall meet the capability of measuring distortion with all applicable signal characteristics IAW requirements of paragraph 3.1.12.2.1 through 3.1.12.2.2 of MIL-PRF-60487B
3.1.12.3 Input
Impedance
The TS-4549 shall meet the capability of measuring distortion with the applicable input impedance IAW requirements of paragraph 3.1.12.3 of MIL-PRF-60487B
RF Reference Source
Distortion Meter
Test Point Frequency (Hz) Target Level (%)
Pass/Fail
600 .1 Pass 700 .5 Pass 800 1 Pass 900 5 Pass 1000 10 Pass 1100 20 Pass 1200 30 Pass 1300 40 Pass 1400 50 Pass
31 January 2022 Page 24 of 34
4.13 SINAD Meter.
Government Specification:
TS-4549 Specification Specification Requirement
3.1.13.1 SINAD The TS-4549 shall meet the capability of measuring SINAD IAW requirements of paragraph 3.1.13.1.1 through 3.1.13.1.3 of
MIL-PRF-60487B
3.1.13.2 Signal
Characteristics
The TS-4549 shall meet the capability of measuring SINAD with applicable signal characteristics IAW requirements of paragraph
3.1.13.2.1 through 3.1.13.2.2 of MIL-PRF-60487B
RF Reference Source
SINAD Meter
Test Point Frequency (Hz) Target Level (dB)
Pass/Fail
100 1 Pass 500 5 Pass 1000 10 Pass 2000 15 Pass 3000 20 Pass 4000 25 Pass 5000 30 Pass 6000 35 Pass 7000 40 Pass 8000 35 Pass 9000 30 Pass 10000 25 Pass
31 January 2022 Page 25 of 34
4.14 AF Level Meter.
Government Specification:
TS-4549 Specification Specification Requirement
3.1.14.1 Input
Connector
The TS-4549 shall meet the requirement of an applicable input connector IAW requirements of paragraph 3.1.14.1 of MIL-PRF- 60487B
3.1.14.2 Input Level
Range
The TS-4549 shall meet the capability of AF input level range IAW requirements of paragraph 3.1.14.2 of MIL-PRF-60487B
3.1.14.3 Resolution The TS-4549 shall meet the capability of AF input level resolution IAW requirements of paragraph 3.1.14.3 of MIL-PRF- 60487B
3.1.14.4 Frequency
Response
The TS-4549 shall meet the capability of AF frequency response IAW requirements of paragraph 3.1.14.4 of MIL-PRF-60487B
3.1.14.5 Accuracy The TS-4549 shall meet the capability of AF accuracy IAW requirements of paragraph 3.1.14.5 of MIL-PRF-60487B
RF Reference Source
AF Level Meter
Test Point (Vrms) Frequency )kHz) Pass/Fail .5 1 Pass 1 1 Pass 2 1 Pass 3 1 Pass 4 1 Pass 5 1 Pass 6 1 Pass 7 1 Pass 8 1 Pass 9 1 Pass
9.99 1 Pass
31 January 2022 Page 26 of 34
4.15 Digital Multimeter.
TS-4549 Specification Specification Requirement
3.1.15.1 Voltmeter
(DC/AC)
The TS-4549 shall meet the capability of measuring voltage IAW requirements of paragraph 3.1.15.1.1 through 3.1.15.1.6 of
MIL-PRF-60487B
3.1.15.2 Ohmmeter The TS-4549 shall meet the capability of measuring ohms IAW requirements of paragraph 3.1.15.2.1 through 3.1.15.2.4 of MIL-
PRF-60487B
3.1.15.3 Current Meter
(AC/DC)
The TS-4549 shall meet the capability of measuring current both AC/DC IAW requirements of paragraph 3.1.15.3.1 through
3.1.15.3.6 of MIL-PRF-60487B
Calibrator
Voltmeter
(AC/DC)
Test Point Amplitude (V) Frequency (Hz)
Pass/Fail
100u 1k Pass 500u 1k Pass 100m 50 Pass 100m 100 Pass 100m 1k Pass 100m 10k Pass 100m 20k Pass
1.8 1k Pass 18 1k Pass 180 100 Pass 300 100 Pass 100u N/A Pass 500u N/A Pass 1m N/A Pass 1 N/A Pass 10 N/A Pass 100 N/A Pass
31 January 2022 Page 27 of 34
200 N/A Pass 300 N/A Pass
Test Plan/Procedure Ohmmeter
Test Point (Ω) Pass/Fail 1 Pass 10 Pass 100 Pass 1k Pass 10k Pass 100k Pass 1M Pass 10M Pass 20M Pass 30M Pass
Test Plan/Procedure Current
(AC/DC)
Test Point Current (A) Frequency (Hz)
Pass/Fail
.01m N/A Pass
.1m N/A Pass 1m N/A Pass 10m N/A Pass 100m N/A Pass 1A N/A Pass 3A N/A Pass 5A N/A Pass 10A N/A Pass .01m 1k Pass .1m 1k Pass 1m 1k Pass 10m 1k Pass 100m 1k Pass 1A 1k Pass 3A 1k Pass 5A 1k Pass 10A 1k Pass
31 January 2022 Page 28 of 34
4.16 Oscilloscope.
Government Specification:
TS-4549 Specification Specification Requirement
3.1.16.1 Vertical Input The TS-4549 shall meet the capability of vertical input IAW requirements of paragraph 3.1.16.1.1 through 3.1.16.1.7 of MIL-
PRF-60487B
3.1.16.2 Horizontal
Sweep
The TS-4549 shall meet the capability of horizontal sweep IAW requirements of paragraph 3.1.16.2.1 through 3.1.16.2.3 of MIL-
PRF-60487B
3.1.16.3 External Input
Impedance
The TS-4549 shall meet the capability of internal external input impedance IAW requirements of paragraph 3.1.16.3 of MIL-
PRF-60487B
Calibrator
Oscilloscope
Vertical Deflection Test Point (V) Pass/Fail 2m Pass 5m Pass 10m Pass 20m Pass 50m Pass 100m Pass 200m Pass 500m Pass 1 Pass 2 Pass 5 Pass 10 Pass 20 Pass
Test Plan/Procedure Oscilloscope
Horizontal Sweep Test Point (ms) Pass/Fail 1 Pass 2 Pass
31 January 2022 Page 29 of 34
External Input Impedance is only a nominal value; it is not a testable point.
4.17 Spectrum Analyzer.
Government Specification:
TS-4549 Specification Specification Requirement
3.1.17.1 Range The TS-4549 shall meet the capability of measuring a signal’s full range in spectrum analysis mode IAW requirements of paragraph 3.1.17.1of MIL-PRF-60487B
3.1.17.2 Frequency
Span
The TS-4549 shall meet the capability of measuring a signal’s frequency span in spectrum analysis mode IAW requirements of paragraph 3.1.17.2.1 through 3.1.17.2.3 of MIL-PRF-60487B
3.1.17.3 Level The TS-4549 shall meet the capability of measuring a signal’s level in spectrum analysis mode IAW requirements of paragraph
3.1.17.3.1 through 3.1.17.3.8 of MIL-PRF-60487B
3.1.17.4 Attenuator The TS-4549 shall meet the capability of measuring attenuation of a signal in spectrum analysis mode IAW requirements of paragraph 3.1.17.4. of MIL-PRF-60487B
3.1.17.5 Minimum
input signal level
The TS-4549 shall meet the capability of measuring a signal with minimum input signal level IAW requirements of paragraph 3.1.71.5 of MIL-PRF-60487B
RF Reference Source
5 Pass 10 Pass 20 Pass 50 Pass 100 Pass
Test Plan/Procedure Oscilloscope
External Input Impedance
Not tested Not tested
31 January 2022 Page 30 of 34
Spectrum Analyzer
Frequency Range Test Point (MHz) Pass/Fail 1 Pass 5 Pass 10 Pass 50 Pass 100 Pass 500 Pass 1000 Pass 1500 Pass 2000 Pass 2500 Pass 2800 Pass
Test Plan/Procedure Spectrum Analyzer
Frequency Span Test Point
RF Reference (MHz) Settings Scan (Hz) Pass/Fail
10 1k Pass 10 1k Pass 20 2k Pass 20 2k Pass 50 5k Pass 50 5k Pass 20 10k Pass 20 10k Pass 20 20k Pass 20 20k Pass 50 50k Pass 50 50k Pass 100 100k Pass 100 100k Pass 200 200k Pass 200 200k Pass 200 500k Pass 200 500k Pass 500 1M Pass 500 1M Pass 500 2M Pass
31 January 2022 Page 31 of 34
500 2M Pass 1000 5M Pass 1000 5M Pass 1000 10M Pass 1000 10M Pass 1500 20M Pass 1500 20M Pass 2000 50M Pass 2000 50M Pass 2500 100M Pass 2500 100M Pass
Test Plan/Procedure Spectrum Analyzer Level Vertical Range
Test Point (dB) Pass/Fail 10 Pass 0 Pass -10 Pass -20 Pass -30 Pass -40 Pass -50 Pass -60 Pass -70 Pass
Test Plan/Procedure Spectrum Analyzer
Attenuator Test Point (dB) Pass/Fail 10 Pass 0 Pass -10 Pass -20 Pass -30 Pass -40 Pass -50 Pass -60 Pass -70 Pass
31 January 2022 Page 32 of 34
4.18 BER Meter.
TS-4549 Specification Specification Requirement
3.1.18.1 Range The TS-4549 shall meet the capability of measuring BER across its full range IAW requirements of paragraph 3.1.18.1 of MIL-
PRF-60487B
3.1.18.2 Data The TS-4549 shall meet the capability of measuring BER data points IAW requirements of paragraph 3.1.18.2.1 through
3.1.18.2.6 of MIL-PRF-60487B
3.1.18.3 External Input
Level
The TS-4549 shall meet the capability of measuring BER over the full external input level IAW requirements of paragraph
3.1.18.3.1 of MIL-PRF-60487B
TS-4544
Spectrum Analyzer
Minimum input signal level Test Point (dB) Pass/Fail 30 Pass 20 Pass 10 Pass 0 Pass -10 Pass -20 Pass -30 Pass -40 Pass -50 Pass -60 Pass -70 Pass -80 Pass -90 Pass -100 Pass
31 January 2022 Page 33 of 34
The US Army does not possess the capability to calibrate/test “Bit Error Rate”.
other tests in the instrumentation test plan. This parameter is not critical to down-select, and it shall be addressed at contract award.
4.19 Internal Reference Oscillator.
Government Specification:
TS-4549 Specification Specification Requirement
3.1.19.1 Accuracy The TS-4549 shall meet the capability of an accurate internal reference oscillator generation of 10MHz IAW requirements of paragraph 3.1.19.1 of MIL-PRF-60487B
3.1.19.2 Stability Level The TS-4549 shall meet the capability of a stable internal reference oscillator generation of 10MHz IAW requirements of paragraph 3.1.19.2 of MIL-PRF-60487B
Range
Test Point Pass/Fail N/A N/A
Test Plan/Procedure Data
Test Point (bps) Pass/Fail 16k
Internal Reference Oscillator
31 January 2022 Page 34 of 34
4.20 Digital Data Generator.
Government Specification:
TS-4549 Specification Specification Requirement
3.1.20.1 Accuracy The TS-4549 shall meet the capability of digital data generation IAW requirements of paragraph 3.1.20.1 of MIL-PRF-60487B
3.1.20.2 Data
Production Rates
The TS-4549 shall meet the capability of digital data generation of data production rates IAW requirements of paragraph 3.1.20.2 of MIL-PRF-60487B
3.1.20.3 Range The TS-4549 shall meet the capability of digital data generation of a full digital data range IAW requirements of paragraph
3.1.20.3 of MIL-PRF-60487B
The US Army does not possess the capability to calibrate/test “Digital modulation”.
Tests completed by
Scott Hill
Steve Aguado
Accuracy/Stability Test Point (MHz) Pass/Fail 10 Pass
Test Plan/Procedure Digital Data Generator
Not Tested Not Tested
| 1 Introduction |
| 2 Equipment Requirements |
| 2.1 System Requirements. |
| 3 Equipment Used in Testing |
| 4 Test Plan/Procedure |
| 4.1 Radio Frequency (RF) Signal Generator (Transmit/Receive (T/R) and Duplex Connector. |
| 4.2 Audio Frequency (AF) Generator #1 and #2 and Boost Mode. |
| 4.3 Audio Frequency Counter. |
| 4.4 Radio Frequency Counter. |
| 4.5 Radio Frequency Error Meter. |
| 4.6 RF Power Meter. |
| 4.7 RF Receiver. |
| 4.8 Deviation Meter. |
| 4.9 AM Modulation Meter. |
| 4.10 Single Side Band. |
| 4.11 PM Deviation Meter. |
| 4.12 Distortion Meter. |
| 4.13 SINAD Meter. |
| 4.14 AF Level Meter. |
| 4.15 Digital Multimeter. |
| 4.16 Oscilloscope. |
| 4.17 Spectrum Analyzer. |
| 4.18 BER Meter. |
| 4.19 Internal Reference Oscillator. |
| 4.20 Digital Data Generator. |
| 2022-05-11T09:36:51-0500 | |
| AGUADO.STEVEN.EDWARD.1236607131 |
| 2022-05-12T21:04:08-0500 | |
| HILL.SCOTT.ALEXANDER.1270801583 |
File details come from the government source that posted it. Updated .