TEST & CERTIFICATION OF CHEMICAL FUME HOODS AND BIOLOGICAL SAFETY CABINETS
Solicitation number
B2727637
Issued by
Pennsylvania

About this file

This is a scope of work attachment for a Pennsylvania state and local contract opportunity for testing and certification of chemical fume hoods and biological safety cabinets. The contract period runs from July 1, 2026 through June 30, 2027. The procurement will be awarded as a whole or by line item, whichever serves the City's best interests, to the lowest responsive and responsible bidder. All bidders must quote all items to be considered eligible for award, and no services may be delivered or provided until official notification from the department is received.

The vendor must hold ISO 17025 accreditation for controlled environment testing and calibration and must perform specified minimum testing protocols for each equipment category: ducted, ductless, and fuming chemical fume hoods; biological safety cabinets; drying cabinets; and laminar flow hoods. Testing requirements include airflow smoke pattern tests, airflow velocity tests, HEPA filter leak tests, down-flow velocity profile tests, inflow velocity tests, and worker comfort and safety tests, depending on equipment type. All testing must comply with manufacturer specifications and applicable state and federal regulations including NSF/ANSI Z9.5 and 49-2002. The vendor is responsible for adjusting equipment to meet specifications at no additional charge where applicable, and must provide repair cost estimates for equipment that cannot be recertified to meet specified standards.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

The following requirements must be met:

a. The vendor must be ISO 17025 Accredited for controlled environment testing and calibration

b. The following minimum test must be performed on Chemical Fume Hoods (Ducted)

i. Airflow Smoke Pattern Test

II. Airflow Velocity Test

c. The following minimum test must be performed on Chemical Fume Hoods (Ductless)

i. Airflow Smoke Pattern Test

ii. Airflow Velocity Test

iii. Filter Check

d. The following minimum test must be performed on Chemical Fume Hoods (Fuming Chambers)

i. Airflow Smoke Pattern Test

ii. Airflow Velocity Test

iii. Filter Check

e. The following minimum test must be performed on Biological Safety Cabinets:

i. HEPA Filter Leak Test

ii. Airflow Smoke pattern Test

iii. Down-flow Velocity Profile Test

iv. Inflow Velocity Test

v. Workers Comfort and Safety Tests

f. The following minimum test must be performed on Drying Cabinets:

i. Airflow Smoke Pattern Test

ii. Airflow Velocity Test with velocity profile recorded

g. The following minimum test must be performed on Laminar Flow Hoods:

i. HEPA Filter Leak Test

ii. Airflow Smoke pattern Test

iii. Airflow Velocity Test

iv. Workers Comfort and Safety Tests

Testing must meet manufacturer’s specifications, as well as any state or federal specifications such as those outlined in NSF/ANSI Z9.5 and 49-2002 or other regulations which may apply.

Supplier will adjust equipment to meet specifications at no additional charge where applicable. If hood cannot be recertified the supplier will supply an estimate on the repair cost for bring the hood within the set specifications.