Attach III- Basis of Design Drawings.pdf

PDF 512 KB Posted

Attached to
Z1DA--Install SCI Patient Lifts 539-26-504
Solicitation number
36C25026R0070
Issued by

About this file

Installation Drawing Summary

This is an installation drawing document for ceiling lift systems prepared by Arjo for a facility in Cincinnati, Ohio (VA CINCINNATI). The document contains 10 pages of detailed technical specifications, room-by-room installation layouts, and assembly requirements for ECS (Electrically Controlled System) ceiling lift track installations across five rooms (B107, B108, B112, B115, B117).

The system specifications include a 600-pound safe working load (272 kg) with an initial load test requirement at 1.5 times the SWL. All installations require a minimum concrete slab structure capacity of 4000 PSI, with ceiling mounting details showing embedment requirements and lateral bracing specifications. Key components provided by Arjo include track, brackets, shims, ceiling plates, channel nuts, bolts, washers, and wiring kits (700-11228 ECS to ECS wiring kit). The facility is responsible for removing obstructions, patching and painting walls, uninstalling existing systems, modifying door headers as needed, and verifying structural capacity. Critical notes specify that cutting of rails occurs on site only, deviations from specifications must be captured and returned to Arjo for revision, and the facility must verify building structure adequacy. Room layouts vary in dimensions and track configurations, with lateral brace requirements calculated based on track bracket joint quantities. The document requires customer approval and signature acknowledging review of all conditions and specifications prior to installation.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

DISTRIBUTOR :

REPRESENTATIVE :

DATE (DD-MMM-YYYY) :

REFERENCE # : PAGE :

INSTALLATION DRAWING

1. The ceiling minimum structure capacity must be able to support the load specified on the installation drawing.

2. Electrical power source outlet already available by the facility at a maximum distance of 3 meters from charging station.

3. Any obstructions such as vents, pipes, lights and sprinklers that could possibly be in the path of the ceiling lifts to be removed or relocated by the facility prior to installation.

4. Modifications of door headers by the facility.

5. Arjo reserves the right to slightly modify the track layout if necessary.

6. See the "Agreed Conditions for the Installation of Ceiling Lifts" document included with this quote for important installation details.

7. Arjo will not be held responsible for any damages, inconvenience, resulting from hitting any non visible conduits during our installation process.

8. Signed drawings are required prior to ordering and installation of ceiling lifts. This signature of each drawings confirms agreement of all the conditions listed above.

IMPORTANT NOTES

MY SIGNATURE BELOW VERIFIES THAT I HAVE:

Reviewed all ____ pages, with a total quantity of ____ rooms, for the approbation of layout for each room and the approbation of structure for each room. I verify that the answers below are accurate to the best of my knowledge and I understand that the quote produced is dependent upon the conditions below as well as this full drawing set.

· Is a load dock available? YES NO

· Can facility accept shipment and store until install begins? YES NO

· Will a drop trailer be needed? YES NO

· Is parking available on site? Is there a fee involved? YES NO

· Are there any installation time frame limitations?

What are they? YES NO

· Is a level of infectious control required? YES NO

If YES, then what level ___________________________

CUSTOMER

APPROVAL: DATE:

REVISION(S) DATE(S) (D-M-Y)

VA CINCINNATI

IR-145577

25-FEB-2026

ARJO

JIM BARTRAM

1 / 10

Legend

IV

(*) Symbol may vary depending of equipment type

Track

ServZone

Track with gate

X-Y Trolley (Suspended)

X-Y Trolley (Embedded)

Turntable (Auto/Manual)

Track with gate reverse

Wall post

Wall bracket

Wall post side

Track with service zone

Track with wall fixation

Track with T-junction

Exchanger

Swivel pole

IV track

Curtain track

Sprinkler

Smoke detector

Electrical outlet

Ceiling light

Ceiling vent

Access trap required

Existing access trap

Additional items

Motor/charger location

SPRINKLER SHIM

ECS-CSTAT

X-Y alim. type + options

VA CINCINNATI 3200 VINE ST CINCINNATI, OH 45220

H

H

105"H140

ECS-ECS

10"

5"

13"

12" 90" 18"

120"

NOTE

UNINSTALL EXISTING LIKO LINEAR WALL

POST SYSTEM.

INSTALL NEW ARJO XY ECS

NEW OUTLET

VA RESPONSIBLE

PATCH AND PAINT WALLS

LISTING OF ROOM NUMBER(S):

- ALL HARDWARE INCLUDED.

(ACCORDING TO STRUCTURE)

QUANTITY OF ROOMS:

DRAWN BY :

REFERENCE # : PAGE :

SCALE :

DATE CREATED (DD-MMM-YYYY) :

SWL WORKING LOAD BELOW LIFT:

INITIAL LOAD TEST TO BE DONE @ SWL x 1.5

VA CINCINNATI

IR-145577

25-FEB-2026

un001629 1/4" = 1'

2 / 10

600LB [272KG]

THIS LAYOUT HAS BEEN PREPARED WITH SUPPLIED DATA. NOTE: CUTTING OF RAILS ON SITE ONLY

B D

27.6 MPA [4000 PSI] MIN.

A

TILE

B107

A=96"

B=43"

D=6" ECS SYSTEM

NOTE: ATTENTION INSTALLERS

*****Any deviation/changes from the Arjo specified installation details i.e. hardware, bracing, track layout

MUST be captured, marked on the drawings and returned to Arjo for an as built revision.*****

NOTE

UNINSTALL EXISTING LIKO LINEAR WALL

POST SYSTEM.

INSTALL NEW ARJO XY ECS

NEW OUTLET

VA RESPONSIBLE

PATCH AND PAINT WALLS

118"H90

118"H90

H

E C

S E

C S

79"H90

79"H90

25"

6"

228"

- ALL HARDWARE INCLUDED.

(ACCORDING TO STRUCTURE)

QUANTITY OF ROOMS:

DRAWN BY :

REFERENCE # : PAGE :

SCALE :

DATE CREATED (DD-MMM-YYYY) :

SWL WORKING LOAD BELOW LIFT:

INITIAL LOAD TEST TO BE DONE @ SWL x 1.5

VA CINCINNATI

IR-145577

25-FEB-2026

un001629 1/4" = 1'

3 / 10

600LB [272KG]

LISTING OF ROOM NUMBER(S):

THIS LAYOUT HAS BEEN PREPARED WITH SUPPLIED DATA. NOTE: CUTTING OF RAILS ON SITE ONLY

B D

27.6 MPA [4000 PSI] MIN.

A

TILE

B112

A=96"

B=43"

D=6" ECS SYSTEM

NOTE: ATTENTION INSTALLERS

*****Any deviation/changes from the Arjo specified installation details i.e. hardware, bracing, track layout

CURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAIN

H

H

105"H140

ECS-ECS

5"

13"

10"

12"90"18"

NOTE

INSTALL NEW ARJO XY ECS

INSTALL NEW OUTLET

VA RESPONSIBLE

NONE

120"

- ALL HARDWARE INCLUDED.

(ACCORDING TO STRUCTURE)

QUANTITY OF ROOMS:

DRAWN BY :

REFERENCE # : PAGE :

SCALE :

DATE CREATED (DD-MMM-YYYY) :

SWL WORKING LOAD BELOW LIFT:

INITIAL LOAD TEST TO BE DONE @ SWL x 1.5

VA CINCINNATI

IR-145577

25-FEB-2026

un001629 1/4" = 1'

4 / 10

600LB [272KG]

LISTING OF ROOM NUMBER(S):

THIS LAYOUT HAS BEEN PREPARED WITH SUPPLIED DATA. NOTE: CUTTING OF RAILS ON SITE ONLY

B D

27.6 MPA [4000 PSI] MIN.

A

TILE

B108

A=96"

B=43"

D=6" ECS SYSTEM

NOTE: ATTENTION INSTALLERS

*****Any deviation/changes from the Arjo specified installation details i.e. hardware, bracing, track layout

UNINSTALL EXISTING LIKO XY WALL POST

SYSTEM.

INSTALL NEW ARJO XY ECS

EXISTING NEW OUTLET

EXTENSION DC CORD

VA RESPONSIBLE

PATCH AND PAINT WALLS

CURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAINCURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

CURTAIN

118"H90

118"H90

H

E C

S E

C S

68"H90

68"H90

6"

216"

- ALL HARDWARE INCLUDED.

(ACCORDING TO STRUCTURE)

QUANTITY OF ROOMS:

DRAWN BY :

REFERENCE # : PAGE :

SCALE :

DATE CREATED (DD-MMM-YYYY) :

SWL WORKING LOAD BELOW LIFT:

INITIAL LOAD TEST TO BE DONE @ SWL x 1.5

VA CINCINNATI

IR-145577

25-FEB-2026

un001629 1/4" = 1'

5 / 10

600LB [272KG]

LISTING OF ROOM NUMBER(S):

THIS LAYOUT HAS BEEN PREPARED WITH SUPPLIED DATA. NOTE: CUTTING OF RAILS ON SITE ONLY

B D

27.6 MPA [4000 PSI] MIN.

A

TILE

B117

A=96"

B=43"

D=6" ECS SYSTEM

NOTE: ATTENTION INSTALLERS

*****Any deviation/changes from the Arjo specified installation details i.e. hardware, bracing, track layout

UNINSTALL EXISTING LIKO XY WALL POST

SYSTEM.

INSTALL NEW ARJO XY ECS

NEW OUTLET

VA RESPONSIBLE

PATCH AND PAINT WALLS

118"H90

118"H90

H

E C

S E

C S

68"H90

68"H90

6"

24"

216"

- ALL HARDWARE INCLUDED.

(ACCORDING TO STRUCTURE)

QUANTITY OF ROOMS:

DRAWN BY :

REFERENCE # : PAGE :

SCALE :

DATE CREATED (DD-MMM-YYYY) :

SWL WORKING LOAD BELOW LIFT:

INITIAL LOAD TEST TO BE DONE @ SWL x 1.5

VA CINCINNATI

IR-145577

25-FEB-2026

un001629 1/4" = 1'

6 / 10

600LB [272KG]

LISTING OF ROOM NUMBER(S):

THIS LAYOUT HAS BEEN PREPARED WITH SUPPLIED DATA. NOTE: CUTTING OF RAILS ON SITE ONLY

B D

27.6 MPA [4000 PSI] MIN.

A

TILE

B115

A=96"

B=43"

D=6" ECS SYSTEM

NOTE: ATTENTION INSTALLERS

*****Any deviation/changes from the Arjo specified installation details i.e. hardware, bracing, track layout

LIFT CAPACITY= 600 LBS

LIFT TRACK BY

ARJO

" MAX. DISTANCE

BETWEEN ANGLE

FITTING AND CEILING

14" MAX

10" MIN

EXISTING CONCRETE

SLAB MIN 3 1/8"

EQUAL +2" OF

CENTERLINE

OF STRUT

2 7 9 11

BRACE

"NOTE: ARJO GUARANTEES THE STRUCTURAL INTEGRITY OF THE TOTAL LIFT SYSTEM DESIGN, INCLUDING TRACK, BRACING, HANGERS, AND ALL CONNECTIONS TO THE BUILDING STRUCTURE. OUR DESIGN IS BASED ON THE REQUIREMENTS OF THE

CURRENT INTERNATIONAL BUILDING CODE (IBC) AND THE SESMIC LOADS OF THE REGION. THE FACILITY MUST VERIFY THE

ADEQUACY OF THE BUILDING STRUCTURE TO SUPPORT THE LOADS IMPOSED BY THE LIFT SYSTEM."

" NOMINAL

EMBEDMENT TYP.

121212

(X2) UPPER AND LOWER

CONNECTION FOR LATERAL

STANDARD TRACK INSTALLATION

L

U

S

P

T

M

A

X

U

N

I

S

T

R

U

P

M

X

***NOTE

EVERYTHING HIGHLIGHTED IN

BLUE IS PROVIDED BY ARJO AND

INSTALLED BY ARJO.****

FINISHED TILE CEILING

BY OTHERS

MIN 4000PSI CONCRETE SLAB STRUCTURE BY OTHERS

Item # Part # Description Qty

000.00394 THREADED ROD 3/8-16 X 10' ZINC 13

000.00402 LOCKNUT STOVER 3/8-16 ZINC 14

000.00405 NUT 3/8-16 ZINC 45

LUS-P3008 3

000.04430 FLAT WASHER M10 ZINC 5

200.11140 CEILING PLATE ø100MM 2

200.11170 BRACKET SHIM 12mm 1

000-01441 HILTI KB-TZ2 3/8 X 3 3/4 3

700.11100 TRACK BRACKET KWIKTRAK 1

LUS-P1068 UNISTRUT ANGLE 2

000.04435 TAB WASHER M10 1

1-5/8 x 10' SLOTTED CHANNEL

000.00425 LOCKWASHER 3/8" 4

000.04875 FLAT WASHER M12 4

16 LUS-P1000T

20"

200.11160

200.11150

BRACKET SHIM 3mm

BRACKET SHIM 1.5mm

000.00375 BOLT 3/8" - 16 X 1 1/4 "

LUS-P1063 1

UNISTRUT SQUARE PLATE

CHANNEL NUT W/O SPRING 3/8"-16

- ALL HARDWARE INCLUDED.

(ACCORDING TO STRUCTURE)

QUANTITY OF ROOMS:

DRAWN BY :

REFERENCE # : PAGE :

SCALE :

DATE CREATED (DD-MMM-YYYY) :

SWL WORKING LOAD BELOW LIFT:

INITIAL LOAD TEST TO BE DONE @ SWL x 1.5

VA CINCINNATI

IR-145577

25-FEB-2026

un001629 N/A

7 / 10

600LB [272KG]

900LB [408KG]

- ALL HARDWARE INCLUDED.

(ACCORDING TO STRUCTURE)

QUANTITY OF ROOMS:

DRAWN BY :

REFERENCE # : PAGE :

SCALE :

DATE CREATED (DD-MMM-YYYY) :

SWL WORKING LOAD BELOW LIFT:

INITIAL LOAD TEST TO BE DONE @ SWL x 1.5

VA CINCINNATI

IR-145577

25-FEB-2026

un001629 N/A

8 / 10

600LB [272KG]

900LB [408KG]

LATERAL BRACE REQUIREMENTS REV-3

LEGEND

LATERAL

BRACE

TRACK

BRACKET

TRACK

BRACKET JOINT

4 TRACKS SECTION

3 TRACK BRACKETS JOINT

9 LATERAL BRACES

1 TRACK SECTION

0 TRACK BRACKET JOINT

3 LATERAL BRACES

2 TRACKS SECTION

1 TRACK BRACKET JOINT

4 LATERAL BRACES

3 TRACKS SECTION

2 TRACK BRACKETS JOINT

6 LATERAL BRACES

- MINIMUM 3 LATERAL BRACES.

- NUMBER OF LATERAL BRACE = (NUMBER OF TRACK BRACKET JOINT X 2) + 2.

POSSIBLE ANCHOR LATERAL TO THE CEILING OR TO THE WALL

ACCEPTABLE ANGLE

LATERAL BRACE

LUS-P1000T

FLAT WASHER M12

000.04875

CONCRETE

HILTI KBTZ2 3/8 X 3 3/4

000-01441

BELOW CEILING TRACK

QUANTITY OF ROOMS:

DRAWN BY :

REFERENCE # : PAGE :

SCALE :

DATE CREATED (DD-MMM-YYYY) :

SWL WORKING LOAD BELOW LIFT:

INITIAL LOAD TEST TO BE DONE @ SWL x 1.5

VA CINCINNATI

IR-145577

25-FEB-2026

un001629 N/A

9 / 10

600LB [272KG]

900LB [408KG]

MAX. STRUT BRIDGE "SPAN"

STRUT

LIFT CAPACITY

P1000T

P1000

P5500T

P5500

P5501

P5001

P1001

P5000

600#

1'-8"

2'-0"

3'-3"

4'-0"

5'-4"

4'-10"

8'-6"

10'-4"

STRUT BRIDGE PER TABLE

BELOW (SLOT- FACE DOWN)

"SPAN"

REF. TABLE BELOW

DUCT, PIPE

OR OTHER

OBSTRUCTION

AS OCCURS

STRUT TO BUILDING STRUCTURE

REF. PLANS AND DETAILS

B

R

A

C

E

O

S

R

U

C

T

U

R

E

R

E

F

P

L

A

N

S

N

D

D

E

T

A

I

L

S

BRIDGE DETAIL

Item # Part # Description Qty

000.00394 THREADED ROD 3/8-16 X 10' ZINC 21

000.00405 NUT 3/8-16 ZINC 6

TYP. ROD

HANGER PER

PLANS AND

DETAILS

SQUARE PLATE 3/8"

"NOTE: ARJO GUARANTEES THE STRUCTURAL INTEGRITY OF THE TOTAL LIFT SYSTEM DESIGN, INCLUDING TRACK, BRACING, HANGERS, AND ALL CONNECTIONS TO THE BUILDING STRUCTURE. OUR DESIGN IS BASED ON THE REQUIREMENTS OF THE

CURRENT INTERNATIONAL BUILDING CODE (IBC) AND THE SESMIC LOADS OF THE REGION. THE FACILITY MUST VERIFY THE

ADEQUACY OF THE BUILDING STRUCTURE TO SUPPORT THE LOADS IMPOSED BY THE LIFT SYSTEM."

LUS-P3008 CHANNEL NUT W/O SPRING 3/8 23

VERIFY IN FIELD

LUS-P1063

QUANTITY OF ROOMS:

DRAWN BY :

REFERENCE # : PAGE :

SCALE :

DATE CREATED (DD-MMM-YYYY) :

SWL WORKING LOAD BELOW LIFT:

INITIAL LOAD TEST TO BE DONE @ SWL x 1.5

VA CINCINNATI

IR-145577

25-FEB-2026

un001629 N/A

10 / 10

Z

600LB [272KG]

900LB [408KG]

EMBEDDED X-Y CONFIG.

700-11228 - ECS to ECS wiring kit

FLUSH X-Y CONFIG.

STANDARD X-Y CONFIG.

WALL MOUNTED X-Y CONFIG.

ECS - ECS CHARGING

CONTACT BOX (3X)

ECS ON FIXED RAIL

ECS ON MOVING RAIL

V5 Duo

MS600 ecs

MS2 ecs

MS440

V5 Duo

MS600 ecs

MS2 ecs

MS440

V5 Duo

MS600 ecs

MS2 ecs

MS440

V5 Duo

MS600 ecs

MS2 ecs

MS440

Sheets and Views
IR-145577 Va Cincinnati-Title
IR-145577 Va Cincinnati-02
IR-145577 Va Cincinnati-03
IR-145577 Va Cincinnati-04
IR-145577 Va Cincinnati-05
IR-145577 Va Cincinnati-06
IR-145577 Va Cincinnati-X-31210.01-S1
IR-145577 Va Cincinnati-X-31210.01-LateralUS
IR-145577 Va Cincinnati-X-31210.01-LateralUS (2)
IR-145577 Va Cincinnati-X-BB-BD-ECS.ECS-X-ECS-ECS

File details come from the government source that posted it. Updated .